Technical data
11.5 Cable requirements
11.5
Cable requirements
Basic data
Max. cable loop resistance
Cable glands on sensor
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used
for the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.
72
No. of conductors
Min. sqr. area
Screen
Max. capacitance
Media temperature:
< 100 °C
> 200 °C
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
WARNING
Coil
cable
2
0.5 mm
2
Yes
N/A
40 Ω
6 Ω
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002
Electrode
cable
3
0.2 mm
2
Yes
350 pF/m
N/A
N/A