DIAGNOSTIC TESTS
DIAGNOSTIC TESTS
DIAGNOSTIC TESTS
DIAGNOSTIC TESTS
DIAGNOSTIC TESTS
The HI 8931 controllers are designed with built-in diagnostic functions to
enable the user to check and troubleshoot the instrument. The checks
performed are through the front panel keys and can be used to isolate the
cause of malfunction.
Press the "MEASURE" key before proceed-
ing the following tests.
Test Offset
A)
Press the "TEST OFFSET" key and the display should indicate the
following values:
HI 8931
0.0 mS
HI 8931
0.00 mS
HI 8931
000 µS
HI 8931
0.0 µS
39
±1.0 mS
±0.10 mS
±10 µS
±1.0 µS