Summary of Contents for Siemens SIREMOBIL Compact L
Page 1
The reproduction, transmission or use of this document or its contents SIREMOBIL Compact L, part number 37 80 629 is not permitted without express and ARCADIS Varic, part number 80 80 017 written authority. Offenders will be liable for damages.
Page 2
0 - 2 Revision Chapter Page Revision SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 4 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Engineering...
Protective conductor test ....... 3 - 10 4 _______Changes to previous version _____________________________________ 4 - 1 Siemens AG n.a. SPR2-130.841.02 Page 3 of 4 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 4
0 - 4 Contents SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 4 of 4 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Solutions...
Compact L, part number 37 80 629 and ARCADIS Varic, part number 80 80 017 systems with the aid of the part set 71 41 000. Siemens AG n.a. SPR2-130.841.02 Page 1 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
TD00-000.860.01... must be complied with. CAUTION In Electrical voltage! Non-compliance can lead to material damage. When working on the system, ESD regulations must be observed. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 6 Siemens AG System Manual Rev.
Perform the particular work steps with special care and attention to detail. If needed, wear working gloves. Siemens AG n.a. SPR2-130.841.02 Page 3 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
(e.g. monitor cable between basic unit and monitor trolley) must be moved section by section to detect broken conductors. The protective conductor resistance may not exceed 0.2 ohms. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 4 of 6...
Page 9
DIN VDE 0751-1:2001-10, Fig. C2. 1 = System 2 = Application part (not present) 3 = Measuring arrangement (integrated in the measuring instrument) Siemens AG n.a. SPR2-130.841.02 Page 5 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Required material Viscogen KL300, 40 g 73 95 353 Heat conduction paste e.g. 20 48 650 Lifting column part set 71 41 000 SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 6 of 6 Siemens AG System Manual Rev. 01 07.05 CS PS 24...
Remove the spacer bushes from the 6 screws. • Open the strain relief of the ribbon cable X 4 (2/Fig.2). Siemens AG n.a. SPR2-130.841.02 Page 1 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 12
2 - 2 Removing the lifting column Fig. 3 Fig. 4 Fig. 5 Fig. 6 SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 6 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Solutions...
Stabilization bracket • Unscrew the stabilization bracket of the monitor cable plug from the lifting column (1/Fig. 6). Siemens AG n.a. SPR2-130.841.02 Page 3 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 14
2 - 4 Removing the lifting column Fig. 7 Fig. 8 Fig. 9 Fig. 10 Fig. 11 Fig. 12 SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 4 of 6 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Solutions...
Two persons are required to lift out the lifting column, pay attention to the connection cable of the lifting column. Siemens AG n.a. SPR2-130.841.02 Page 5 of 6 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 16
2 - 6 Removing the lifting column This page is intentionally left blank. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 6 of 6 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Solutions...
Set the lifting column in the guide (1/Fig. 1) and screw it tight from the bottom of the unit (torque 40 Nm). Siemens AG n.a. SPR2-130.841.02 Page 1 of 10 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 18
Phase to X7 A/2 for raising Phase to X7 A/3 for lowering NOTE Connect the emergency stop switch for moving the lifting column. The emergency stop switch must not be pressed. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 10 Siemens AG System Manual Rev.
Page 19
Three persons are again required to lift the C-arm out from its bearing (Fig. 5) and place it on a table. • Unscrew the bearing of the C-arm again from the lifting column (1/Fig. 4). Siemens AG n.a. SPR2-130.841.02 Page 3 of 10 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Screw the protective conductor (3/Fig. 7) back onto the chassis (on the operating side from the left below). • Refit the small rectangular cover panel underneath the steering lever (2 paneling screws). SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 4 of 10 Siemens AG System Manual Rev.
• Screw tight the cover plates under the swivel brake. • Fasten the railing handle back on the C-arm. Siemens AG n.a. SPR2-130.841.02 Page 5 of 10 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
• Fasten the line filter with two screws to the base plate. • Fasten the BNC plug on the base pate. • Plug in the BNC cable. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 6 of 10 Siemens AG System Manual Rev.
Fasten the D2 board with the spacers (Fig. 13). • Plug the intermediate plate back on. • Fit the spacer tubes on the 6 screws. Siemens AG n.a. SPR2-130.841.02 Page 7 of 10 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 24
Fasten the C-arm cable (1/Fig. 14) and plug in the connectors. • Draw the Diamentor cable through to the left behind the lifting column. • Fasten the bracket (2/Fig. 15) back of the lifting column. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 8 of 10 Siemens AG System Manual Rev.
• Install all covers; connect the protective conductors to the covers and reconnect the emergency stop button. Siemens AG n.a. SPR2-130.841.02 Page 9 of 10 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
• Perform the protective conductor test according to ARTD-002-731.17..Observe the instructions for the protective conductor test in the first chapter. The protective conductor resistance must not exceed 0.2 ohms. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 10 of 10 Siemens AG System Manual Rev.
Page 27
Changes to previous version 4 - 1 New publication Siemens AG n.a. SPR2-130.841.02 Page 1 of 2 SIREMOBIL Compact L Medical Solutions Rev. 01 07.05 CS PS 24 System Manual...
Page 28
4 - 2 Changes to previous version This page intentionally left blank. SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 2 Siemens AG System Manual Rev. 01 07.05 CS PS 24 Medical Solutions...