Maximum Process Temperature Chart

WARNING: Internal temperature must not exceed 80
The chart below is for guidance only. Information to be updated to SITRANS LR
Note:
460 information.
.
Maximum Flange and Process Temperatures versus Allowable Ambient for
Polypropylene Rod or Flange Adaptor versions of SITRANS LR 200
• The chart does not represent every
possible process connection
arrangement. For example, it will NOT
apply if you are mounting the LR 200 on
a nozzle greater than 8" nominal, or
directly on a metallic vessel surface.
• The chart does not take into
consideration heating from direct
sunshine exposure.
Where the chart does not apply, please use your own judgement regarding the use of
SITRANS LR 200. Open the menu View – Peak Values, and click on the temperature tab to see
the maximum and minimum temperatures recorded. These can be used to determine whether
the installation needs to be redesigned.
For example, if the internal temperature exceeds the maximum allowable limit, a sun shield or a
longer nozzle may be required.The peak temperatures indicate the extent of change required to
the installation in order to provide a reliable thermal-operating zone for SITRANS LR 200.
7ML19985JM01
polypropylene
rod
temperature
(no flange)
Process Flange Surface Temperature (
SITRANS LR 460 – INSTRUCTION MANUAL
o
C (176
flange
process
temperature
temperature
o
C)
internal
enclosure
temperature
o
F).
ambient
temperature
process
temperature