Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
• 200 N spring scale e.g. Part No. 44 15 113 RH090 • Torque wrench 20 Nm to 100 Nm e.g. Part No. 80 86 159 RE999 Siemens AG SPR2-130.061.02 Page 1 of 10 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Position labels The labels used in these instructions (left, right, front and back) are explained in Fig. 1. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 10 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 9
Avoid all contact with blood or other body secretions! Comply strictly with the preventive measures against infectious diseases specified in ARTD-002.731.37...! Siemens AG SPR2-130.061.02 Page 3 of 10 SIREMOBIL Compact Medical Solutions Rev.
Page 10
5 minutes before resuming work. CAUTION Electrical voltage! Noncompliance can lead to damage to property. ESD protection guidelines must be observed when servicing. SIREMOBIL Compact SPR2-130.061.02 Page 4 of 10 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 11
Perform the corresponding work with special care and attentive- ness. Wear working gloves if necessary. Siemens AG SPR2-130.061.02 Page 5 of 10 SIREMOBIL Compact Medical Solutions Rev.
Compact systems which exceed this threshold must be serviced. The values measured are to be recorded in the equivalent device leakage current / protocol (refer to log book, Register 3). SIREMOBIL Compact SPR2-130.061.02 Page 6 of 10 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Remount the protective conductor for the lower cover. • Install the cover and tighten the screws. • Remount the protective conductor for the Logbook cover. • Install the cover and tighten the screws. Siemens AG SPR2-130.061.02 Page 7 of 10 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
• Use a paper towel to wipe off excess sealing compound. • Refasten the SIREPHOS cover with the cover screws. SIREMOBIL Compact SPR2-130.061.02 Page 8 of 10 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Remount the back stop plate (3/Fig.1) and fasten it with the screws. • Reinstall the rubber cover (1/Fig. 1) and close it (adhesive closure). • Check the function of the horizontal brake. Siemens AG SPR2-130.061.02 Page 9 of 10 SIREMOBIL Compact Medical Solutions Rev.
1 - 10 Prerequisites Concluding steps • After completing all work, perform the protective conductor test (see ARTD-002.731.17). SIREMOBIL Compact SPR2-130.061.02 Page 10 of 10 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Optical isolation line Optical isolation line L 230 V ~ to cannot be adjusted voltage cable, N voltage cable 246 V ~ Tab. 1 Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact Medical Solutions Rev. 11 09.04 CS PS 24...
Page 18
2 - 2 Line voltage adjustment This page intentionally left blank. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
(0V) format for zoom ratio format UI 15 1:10000 30 kV, Anode UI 30 1:10000 Tab. 2 Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact Medical Solutions Rev. 11 09.04 CS PS 24 System Manual...
Page 20
3 - 2 Checking the operating voltages This page intentionally left blank. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
22-03-01_37 76 494_01234.txt • Store the file on the hard disk drive of the service PC to a path of your choice. • Exit the service program. Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Page 22
4 - 2 Loading the operating software • Copy the previously stored file to a formatted diskette. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 23
Release the radiation release button as soon as the message "Stop monitoring" appears in the ACTIONS line. • Click on "Update Stop". • Perform the generator adjustment (filament circuit learning program). Siemens AG SPR2-130.061.02 Page 1 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
• Activate the radiation release button. • Click on "Update Stop". • Switch the SIREMOBIL Compact OFF and back ON again. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Select "Cassette switch off" on the service PC in "Available Functions". Direct exposure mode is cancelled. • If the unit exceeds the tolerance range, the generator adjustment must be repeated. Siemens AG SPR2-130.061.02 Page 3 of 30 SIREMOBIL Compact Medical Solutions Rev.
Page 26
Tube current exceeds the lower limit 6 ms - preset filament current too low 14.5 mA 13 mA 11.5 mA 3 ms Fig. 1b SIREMOBIL Compact SPR2-130.061.02 Page 4 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
TV iris value. • Insert 25 mm Al and 2.1 mm Cu at the radiation exit port of the SIREPHOS. • The grid remains on the I.I. Siemens AG SPR2-130.061.02 Page 5 of 30 SIREMOBIL Compact Medical Solutions Rev.
LOW - Zoom format pos. >= LOW - Full format pos. Unacceptable parameter values are: LOW - Zoom format pos. < LOW - Full format pos. SIREMOBIL Compact SPR2-130.061.02 Page 6 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
• The "BRIGHTNESS DEVIATION" value is shown in the "BRIGHTNESS DEVIATION" line and should be 0 + 3, refer also to notice on previous page • Switch off fluoro. Siemens AG SPR2-130.061.02 Page 7 of 30 SIREMOBIL Compact Medical Solutions Rev.
If necessary, temporarily change the dose rate levels and kV-mA curves. Tolerance margin for the dose rates: + 15%. SIREMOBIL Compact SPR2-130.061.02 Page 8 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
If the maximum tube current was already reduced, the mA value displayed at 110 kV should be recorded and used for the calcula- tion. The maximum skin dose rate to be set should equal 9 R/min. Siemens AG SPR2-130.061.02 Page 9 of 30...
Only then will the modified parameters be stored on the floppy disk as a new file. Checking • After programming, check the maximum skin dose rate again. SIREMOBIL Compact SPR2-130.061.02 Page 10 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Close the status window by clicking on the OK button. • For the query "Do you want to reset the unit," click "No". • The X-Iris can now be completely opened and closed with the X-Iris keys. Siemens AG SPR2-130.061.02 Page 11 of 30 SIREMOBIL Compact Medical Solutions Rev.
Select the Data, Backup... menu. • Select Parameter in the backup window. • Click on "Backup". At "Remarks for Parameter" enter "X-Iris set". SIREMOBIL Compact SPR2-130.061.02 Page 12 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
The blades should be in the vertical on the monitor screen. • Switch off fluoro. • In the window "Available Functions:" select "0 DEGREE POSITION" and click "Execute". Siemens AG SPR2-130.061.02 Page 13 of 30 SIREMOBIL Compact Medical Solutions Rev.
Then click "OK" . The new position values are stored to disk. • You can now exit from the service program. SIREMOBIL Compact SPR2-130.061.02 Page 14 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
In the "Available Functions:" window select "Close Pos - II. Full Format" and click on "Ex- ecute". • Close the status window by clicking on the OK button. • For the query "Do you want to reset the unit", click on "No". Siemens AG SPR2-130.061.02 Page 15 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
For the query "Do you want to reset the unit", click on "Yes". Concluding tasks Before ending the service program, save the new parameter values to disk (Service PC, Data menu, Backup). SIREMOBIL Compact SPR2-130.061.02 Page 16 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
• In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Full Format" and click "Execute". • Close the status window by clicking on the OK button. Siemens AG SPR2-130.061.02 Page 17 of 30 SIREMOBIL Compact Medical Solutions Rev.
For the query "Do you want to reset the Unit", click on "Yes". Concluding tasks • Before exiting the service program, save the new parameter values to disk (Service PC, Data menu, Backup). SIREMOBIL Compact SPR2-130.061.02 Page 18 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
If necessary, repeat the programming. • Click "Save to file" . • Leave the service program with "Logoff" and "Quit" to save the parameters as a new file to disk. Siemens AG SPR2-130.061.02 Page 19 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
With the left camera rotation key (the arrow on the key is pointing counterclockwise), the marker superimposed on the monitor will rotate counterclockwise. The image contents will rotate clock- wise. SIREMOBIL Compact SPR2-130.061.02 Page 20 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 43
Click on "Backup". • In "Remarks for Parameter", enter "camera rotation set". • Click "OK". The new position values for camera rotation will be stored to the disk. Siemens AG SPR2-130.061.02 Page 21 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
The Qt * Kg value was checked and is programmed. Refer to Programming. • Connect the service PC to the serial interface on board D1. SIREMOBIL Compact SPR2-130.061.02 Page 22 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Preparations • Attach and center the center cross on the I.I. (Fig. 6). • Select survey format. • Select KV mA curve plateau HC1. Siemens AG SPR2-130.061.02 Page 23 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04 CS PS 24...
To calculate the measured area dose product: (measured area dose product FDPg) = (measured dose K ) * (diameter) * 0.829. SIREMOBIL Compact SPR2-130.061.02 Page 24 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ] / (measured area dose product FDPg) For deviations > 0.3 , the area dose product measurement device must be replaced. Siemens AG SPR2-130.061.02 Page 25 of 30...
Threshold 120 is programmed, if not, program these val- ues. Circle mask setup Allows you to configure the diameter of the circle mask. SIREMOBIL Compact SPR2-130.061.02 Page 26 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 49
MEMOSKOP. The MEMOSKOP can be reset via "Load factory default," as required. Subsequently you have to program accord- ingly. The same applies if you replace the MEMOSKOP. Siemens AG SPR2-130.061.02 Page 27 of 30 SIREMOBIL Compact Medical Solutions Rev.
Page 50
Roadmap K-factor, phase A Noise suppression, phase A, is programmed in the road- map mode. Phase A: Mask acquisition. (1) in menu Diagnostics SIREMOBIL Compact SPR2-130.061.02 Page 28 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
"->" pressed 3 times: 120 Hz video frequency "->" pressed 4 times: 50 Hz video frequency. "->" = Press the right cursor key "<-" = Press the left cursor key. Siemens AG SPR2-130.061.02 Page 29 of 30 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
MEMOSKOP, configuration and country-specific language (User Setup) Start the program and transfer the country-specific language to SIREMOBIL Compact via the service PC. SIREMOBIL Compact SPR2-130.061.02 Page 30 of 30 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
• Ensure that the laser targeting device (where present) is functioning properly. • Test the FL / PFL / DR and direct exposure functions of the SIREMOBIL Compact. Siemens AG SPR2-130.061.02 Page 1 of 28 SIREMOBIL Compact Medical Solutions Rev.
Replace the D40 board. • Using the operating instructions for the basic system, check both the movements of the C-arm and the signal messages. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
• Replace the power supply. • Check the power supply voltage and adjust it if necessary. Refer to Chapter 3 of these in- structions and check the operating voltages. Siemens AG SPR2-130.061.02 Page 3 of 28 SIREMOBIL Compact Medical Solutions Rev.
Remove the old set of collimator labels near the SIREPHOS and attach the new labels included in the same position. Check that the original collimator labels are the same as the new ones. SIREMOBIL Compact SPR2-130.061.02 Page 4 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Ensure that the area/dose product measurement device is functioning properly (if present). • Check the dose rate settings and adjust them if necessary. • Check the settings of the maximum skin dose rate. Siemens AG SPR2-130.061.02 Page 5 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
If the indicator is discolored (inner field is black), proceed according to the IQ document RXD-000.038.01 Fig. 1 Fig. 2 Fig. 3 Fig. 4 SIREMOBIL Compact SPR2-130.061.02 Page 6 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Remove the retaining clips on the cables (4/Fig. 1). • Remove the O-ring and the cap nut from the anode cable. • Disconnect plug M30.X1 (4/Fig. 3). Siemens AG SPR2-130.061.02 Page 7 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
- When doing this, make sure that all the cables are correctly positioned. • Secure the I.I. on the mounting (5/Fig. 2). SIREMOBIL Compact SPR2-130.061.02 Page 8 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
- When doing this, check the resolution first and, if necessary, readjust the optical reso- lution of the Compact optics. • Complete the country-specific acceptance (§16 Partial acceptance... /DHHS...). Siemens AG SPR2-130.061.02 Page 9 of 28 SIREMOBIL Compact Medical Solutions Rev.
Turn the C-arm so that the I.I. is at the top. • Loosen and remove the I.I. cover. • Rotate the camera in the "0" position (6/Fig. 6). SIREMOBIL Compact SPR2-130.061.02 Page 10 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
To do this, connect an ohm meter to the potentiometer for camera rotation, location 2 and 3 (5/Fig. 7). The resistance must measure 5000 ohms + 40 ohms in the middle posi- tion. Siemens AG SPR2-130.061.02 Page 11 of 28...
Mark the center of the washer or lead cross on the monitor (see Fig. 9). • Rotate the camera in the -220° position. • Mark the center again on the monitor (see Fig. 9). SIREMOBIL Compact SPR2-130.061.02 Page 12 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
I.I. input screen diameter to the I.I. output screen diameter, a shift of the camera op- tics by approximately 0.1 mm corresponds to a position shift of the washer displayed on the monitor of more than 1.3 mm (17 cm I.I.) or 1.7 mm (23 cm I.I.). Siemens AG SPR2-130.061.02 Page 13 of 28...
Check the display of the blades on the monitor and adjust if necessary. • Check the dose rate again and adjust it if necessary. • Perform the IQ quick test. SIREMOBIL Compact SPR2-130.061.02 Page 14 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 67
Display of rotated image Configure Yes. Monitor zoom Configure No. Default LUT Configure 2. Program a different LUT curve only upon customer request. Siemens AG SPR2-130.061.02 Page 15 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04 CS PS 24...
Page 68
Dose on hard copy output If the area dose product meter is present, (Diamentor) configure Yes. Otherwise configure No. • Perform the IQ quick test. SIREMOBIL Compact SPR2-130.061.02 Page 16 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Date / Time format Set the country-specific format. Keyboard Set to the keyboard layout being used. Language Set for language of the country. Siemens AG SPR2-130.061.02 Page 17 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Page 70
Configure horizontally. Configure vertically only upon customer request. Display of time Configure Yes. Set Date Time Set the current date and time. SIREMOBIL Compact SPR2-130.061.02 Page 18 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 71
Yes. Otherwise configure No. Dose on hardcopy output If area dose product meter is present, (Diamentor) con- figure Yes. Otherwise configure No. • Perform the IQ quick test. Siemens AG SPR2-130.061.02 Page 19 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Configure Yes. Select No only upon customer request. Display of rotated image Configure Yes. Monitor zoom Configure No. Monitor split Configure Yes. SIREMOBIL Compact SPR2-130.061.02 Page 20 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 73
Split screen Configure horizontally. Configure vertically only upon customer request. Display of time Configure Yes. Set Date Time Set the current date and time. Siemens AG SPR2-130.061.02 Page 21 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04 CS PS 24...
Page 74
Dose on hardcopy output If area dose product meter is present, (Diamentor) con- figure Yes. Otherwise configure No. • Perform the IQ quick test. SIREMOBIL Compact SPR2-130.061.02 Page 22 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Set for language of the country. Display of realtime clock Configure Yes. Select No only upon customer request. Display of rotated image Configure Yes. Monitor zoom Configure No. Siemens AG SPR2-130.061.02 Page 23 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Page 76
Program a different value only upon customer request. Roadmap Time of phase A Program "4.5". Program a different value only upon customer request. SIREMOBIL Compact SPR2-130.061.02 Page 24 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 77
Yes. Otherwise configure No. Dose on hardcopy output If area dose product meter is present, (Diamentor) con- figure Yes. Otherwise configure No. • Perform the IQ quick test. Siemens AG SPR2-130.061.02 Page 25 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
"Testing the accuracy of the area dose product measurement device". • Re-attach the SIREPHOS cover. • Seal the SIREPHOS cover using sealing compound (see Chapter 1, "Covers"). SIREMOBIL Compact SPR2-130.061.02 Page 26 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Adjust the laser diodes using the adjustment device supplied with the laser targeting de- vice. • Re-attach the SIREPHOS cover. • Seal the SIREPHOS cover using sealing compound (see Chapter 1, "Covers"). Siemens AG SPR2-130.061.02 Page 27 of 28 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Page 80
6 - 28 Replacing boards / replacing components This page intentionally left blank. SIREMOBIL Compact SPR2-130.061.02 Page 28 of 28 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
20 on the scale. • All measurements should be made with a spring measure on the railing near the image intensifier with the DHHS spacer, if present, installed on the X-ray tube. SIREMOBIL Compact L Movement Nominal values Comments Brake released:...
Oil to be used: Special purpose oil (Optimol GmbH; Viscogen KL300; 40 g), Item number 73 95 353 RH090. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 83
At 110kV, an overall current of 5.275 mA is measured. 110 [kV] tube current [mA] = 5.275 [mA] - = 5.275 [mA] - 0.275 [mA] = 5.0 [mA] 400 [MOhm] Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact Medical Solutions Rev.
Page 84
8 - 2 Supplement, Measuring the tube current This page intentionally left blank. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...
Page 85
Cover page, revision level, table of contents revised Chapter 1 to chapter 9 Revision level revised. Chapter 1 Safety information added. Chapter 9 Changes to previous version revised Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact Medical Solutions Rev. 11 09.04...
Page 86
9 - 2 Changes to previous version This page intentionally left blank. SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG System Manual Rev. 11 09.04 CS PS 24 Medical Solutions...