hit counter script

Replacing The Drain Filters - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

the Startup tab. Select
0.1 or less (1000 cells/µL).
are filling up with wash solution to their normal wash levels.
the St
is 5 or less (1000 cells/µL).
If required, run additional saline samples until acceptable background counts
are obtained for two consecutive samples. If more than five saline samples
are required, verify that the pump was installed properly.
7. Run controls to verify analyzer performance.

Replacing the Drain Filters

If the filter is not clogged but you want to clean it because of heavy buildup or
discoloration, use the backflush procedure.
Time: 10 minutes
Analyzer mode: Off
The analyzer must be off; otherwise, personal injury from the needle may occur.
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
1. Turn the analyzer off.
2. Unscrew the old filter from both fittings. Discard the old filter.
3. Connect the new filter to the fittings and hand tighten them to ensure they are
sealed.
Troubleshooting the Analyzer
If DP1 is replaced, run two saline primers. At the
Refresh
If DP4 is replaced, do a system wash and verify that the reaction chambers
If DP5 is replaced, run two saline primers. At the Operations menu, select
rtup tab. Select
a
Refresh
WARNING
BIOHAZARD
Operations
and verify that the WBC background count is
and verify that the platelet background count
menu, select
5-27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents

Save PDF