UROSKOP D
Maintenance Instructions
The maintenance protocol
RLL5-310.105.02.02.02
is required for this instructions
Print No.: RLL5-310.101.02.02.02
Replaces: RLL5-310.101.02.01.02
SP
© Siem ens A G 1 993
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 01.97

Advertisement

Table of Contents
loading

Summary of Contents for Siemens UROSKOP D

  • Page 1 UROSKOP D Maintenance Instructions The maintenance protocol © Siem ens A G 1 993 The reproduction, transmission or RLL5-310.105.02.02.02 use of this document or its contents is not permitted without express written authority. Offenders will be is required for this instructions liable for damages.
  • Page 2 0 - 2 Revision Chapter Page Revision 1 - 4 1 - 4 1 - 16 1 - 2 UROSKOP D RLL5-310.101.02 Page 2 of 4 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...
  • Page 3: Table Of Contents

    Concluding steps ........2 - 16 3 _______Changes to previous version _____________________________________ 3 - 1 Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 4: Contents

    0 - 4 Contents This page intentionally left blank. UROSKOP D RLL5-310.101.02 Page 4 of 4 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...
  • Page 5: Required Documentation

    1 kg can 73 95 445 • Optimol Optiflex ELM paste 1 kg can 15 90 293 • Silicon paste 100 g tube 20 49 914 Siemens AG RLL5-310.101.02 Page 1 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 6: Text Emphasis

    When performing service work and tests adhere to: CAUTION - the product-specific safety information in the document, - the safety instructions in RA0-000.012.40... in the UROSKOP D - Logbook, as well as - the general safety information contained in the TI folder / ARTD Part 2 (CD-ROM) .
  • Page 7: Explanation Of Abbreviations

    System Quality Check Software Maintenance The items identified by these abbreviations are contained in the maintenance certificate and should be checked off upon completion. Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 8 1 - 4 General information This page intentionally left blank. UROSKOP D RLL5-310.101.02 Page 4 of 4 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...
  • Page 9: Visual Inspection

    Remove the covers from the lifting base, the spotfilm device and the X-ray tube support arm / rotation drive. 2.2.2 System base Torque Check the torque (85 Nm) of the four mounting screws on the lifting base. Siemens AG RLL5-310.101.02 Page 1 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 10: Lifting Drive

    Check the chain coupler position on the outside of the lifting base (2/Fig. 1/2); the bolt for the chain coupler (K1) must firmly connect with the outer surface of the lifting base (F) (Fig. 2). • Check that the safety screw (S) is secured tightly. UROSKOP D RLL5-310.101.02 Page 2 of 16 Siemens AG Rev. 02 01.97...
  • Page 11: Tilt Drive

    (from the edge of the tabletop on the physician’s side to the white plastic piece in the table rail) and check for noise and smooth travel. Siemens AG RLL5-310.101.02 Page 3 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 12: Transverse Movement Drive

    Insert the tabletop extension on the head end. • Move the table headward to the maximum end stop. • Check the safety distances to the wall or to other fixed installations. UROSKOP D RLL5-310.101.02 Page 4 of 16 Siemens AG Rev. 02 01.97...
  • Page 13: System General

    - Error 0050 must appear on board D1 in M20. - system movements must be blocked. • Unlock the emergency OFF switches on the system - the system must automatically reset. Siemens AG RLL5-310.101.02 Page 5 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 14: Checking Radiation Protection

    Indicator “Radiation on” • Release an exposure with 40 kV and 2 mAs and check whether the radiation ON indicators light up on the system and the main operating panels. UROSKOP D RLL5-310.101.02 Page 6 of 16 Siemens AG Rev. 02 01.97...
  • Page 15: Periodic Maintenance / Parts Replacement

    Service Instructions RLL5-310.091.06. Rotating belt • After five years of service or when worn, replace the belt for the cassette loading according to Service Instructions RLL5-310.091. Siemens AG RLL5-310.101.02 Page 7 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 16: Inspection For Wear And Tear

    Visually inspect themotor drive belt at 7 for wear and tear. Transverse table drive - ball bearing guide • Visually inspect the ball bearing guide for wear and tear. UROSKOP D RLL5-310.101.02 Page 8 of 16 Siemens AG Rev. 02 01.97...
  • Page 17: Maintenance

    Lightly lubricate the chain guide (5/Fig. 3) where accessible with LONGTIME PD2. Toothed rack • Lightly grease the toothed rack (6/Fig. 3)of the potentiometer drive (r102) with LONGTIME PD2. Siemens AG RLL5-310.101.02 Page 9 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 18: Tilt Drive

    Remove any contrast agent residues and other contamination from points on the system normally not accessible to the customer. Cables and hoses • Clean all accessible cables and hoses. UROSKOP D RLL5-310.101.02 Page 10 of 16 Siemens AG Rev. 02 01.97...
  • Page 19: Inspecting The Internal Heat Dissipation

    Inspect all air vents on the system cabinets and make sure that there are no obstructions, e.g. clogged by dirt, etc. (Fig. 4). • If necessary, clean any vents that are clogged. Siemens AG RLL5-310.101.02 Page 11 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 20: Operating Values - Inspection

    - Release an exposure using values 45 kV, 3.2 mA. - Develop the film. - Measure the deviations between the light and radiation fields; they must be smaller than 0.5 cm. UROSKOP D RLL5-310.101.02 Page 12 of 16 Siemens AG Rev.
  • Page 21: Collimator Blades

    If the tension was adjusted, reference switch S51 for the cassette height measurement must be recalibrated. Refer to Adjustment instructions RLL5-310.071.01.., chapter 8, page 8-3. Siemens AG RLL5-310.101.02 Page 13 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 22: Function Inspection

    The cassette must be retracted into the spotfilm device and then ejected back into the loading position. The spotfilm device should sound an acoustic warning. NOTICE The 43 cm horizontal format has not been included in the cassette program. UROSKOP D RLL5-310.101.02 Page 14 of 16 Siemens AG Rev. 02 01.97...
  • Page 23: Component Quality Inspection

    Do not use abrasive cleaners, (due to possible material incompatibility) organic cleaners or cleaning agents containing organic substances (e.g. benzine, alcohol, spot remover). Siemens AG RLL5-310.101.02 Page 15 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 24: Concluding Steps

    Clean and reinstall the covers. Reconnect all protective conductors. 2.11 Concluding steps Protective conductor test • Perform the final protective conductor test according to TI 236, section VII. Maximum total resistance: 0.2 ohm. UROSKOP D RLL5-310.101.02 Page 16 of 16 Siemens AG Rev. 02 01.97 TDSP 1...
  • Page 25 New layout (converted to new templates) Work steps expanded and sequence changed. TDSP 2 / Fleischter TDSP 1 / Groß SMS Iselin / O’Donnell Siemens AG RLL5-310.101.02 Page 1 of 2 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
  • Page 26 3 - 2 Changes to previous version This page intentionally left blank. UROSKOP D RLL5-310.101.02 Page 2 of 2 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...