Concluding steps ........2 - 16 3 _______Changes to previous version _____________________________________ 3 - 1 Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
1 kg can 73 95 445 • Optimol Optiflex ELM paste 1 kg can 15 90 293 • Silicon paste 100 g tube 20 49 914 Siemens AG RLL5-310.101.02 Page 1 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
When performing service work and tests adhere to: CAUTION - the product-specific safety information in the document, - the safety instructions in RA0-000.012.40... in the UROSKOP D - Logbook, as well as - the general safety information contained in the TI folder / ARTD Part 2 (CD-ROM) .
System Quality Check Software Maintenance The items identified by these abbreviations are contained in the maintenance certificate and should be checked off upon completion. Siemens AG RLL5-310.101.02 Page 3 of 4 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Page 8
1 - 4 General information This page intentionally left blank. UROSKOP D RLL5-310.101.02 Page 4 of 4 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...
Remove the covers from the lifting base, the spotfilm device and the X-ray tube support arm / rotation drive. 2.2.2 System base Torque Check the torque (85 Nm) of the four mounting screws on the lifting base. Siemens AG RLL5-310.101.02 Page 1 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Check the chain coupler position on the outside of the lifting base (2/Fig. 1/2); the bolt for the chain coupler (K1) must firmly connect with the outer surface of the lifting base (F) (Fig. 2). • Check that the safety screw (S) is secured tightly. UROSKOP D RLL5-310.101.02 Page 2 of 16 Siemens AG Rev. 02 01.97...
(from the edge of the tabletop on the physician’s side to the white plastic piece in the table rail) and check for noise and smooth travel. Siemens AG RLL5-310.101.02 Page 3 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Insert the tabletop extension on the head end. • Move the table headward to the maximum end stop. • Check the safety distances to the wall or to other fixed installations. UROSKOP D RLL5-310.101.02 Page 4 of 16 Siemens AG Rev. 02 01.97...
- Error 0050 must appear on board D1 in M20. - system movements must be blocked. • Unlock the emergency OFF switches on the system - the system must automatically reset. Siemens AG RLL5-310.101.02 Page 5 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Indicator “Radiation on” • Release an exposure with 40 kV and 2 mAs and check whether the radiation ON indicators light up on the system and the main operating panels. UROSKOP D RLL5-310.101.02 Page 6 of 16 Siemens AG Rev. 02 01.97...
Service Instructions RLL5-310.091.06. Rotating belt • After five years of service or when worn, replace the belt for the cassette loading according to Service Instructions RLL5-310.091. Siemens AG RLL5-310.101.02 Page 7 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Visually inspect themotor drive belt at 7 for wear and tear. Transverse table drive - ball bearing guide • Visually inspect the ball bearing guide for wear and tear. UROSKOP D RLL5-310.101.02 Page 8 of 16 Siemens AG Rev. 02 01.97...
Lightly lubricate the chain guide (5/Fig. 3) where accessible with LONGTIME PD2. Toothed rack • Lightly grease the toothed rack (6/Fig. 3)of the potentiometer drive (r102) with LONGTIME PD2. Siemens AG RLL5-310.101.02 Page 9 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Remove any contrast agent residues and other contamination from points on the system normally not accessible to the customer. Cables and hoses • Clean all accessible cables and hoses. UROSKOP D RLL5-310.101.02 Page 10 of 16 Siemens AG Rev. 02 01.97...
Inspect all air vents on the system cabinets and make sure that there are no obstructions, e.g. clogged by dirt, etc. (Fig. 4). • If necessary, clean any vents that are clogged. Siemens AG RLL5-310.101.02 Page 11 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
- Release an exposure using values 45 kV, 3.2 mA. - Develop the film. - Measure the deviations between the light and radiation fields; they must be smaller than 0.5 cm. UROSKOP D RLL5-310.101.02 Page 12 of 16 Siemens AG Rev.
If the tension was adjusted, reference switch S51 for the cassette height measurement must be recalibrated. Refer to Adjustment instructions RLL5-310.071.01.., chapter 8, page 8-3. Siemens AG RLL5-310.101.02 Page 13 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
The cassette must be retracted into the spotfilm device and then ejected back into the loading position. The spotfilm device should sound an acoustic warning. NOTICE The 43 cm horizontal format has not been included in the cassette program. UROSKOP D RLL5-310.101.02 Page 14 of 16 Siemens AG Rev. 02 01.97...
Do not use abrasive cleaners, (due to possible material incompatibility) organic cleaners or cleaning agents containing organic substances (e.g. benzine, alcohol, spot remover). Siemens AG RLL5-310.101.02 Page 15 of 16 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Clean and reinstall the covers. Reconnect all protective conductors. 2.11 Concluding steps Protective conductor test • Perform the final protective conductor test according to TI 236, section VII. Maximum total resistance: 0.2 ohm. UROSKOP D RLL5-310.101.02 Page 16 of 16 Siemens AG Rev. 02 01.97 TDSP 1...
Page 25
New layout (converted to new templates) Work steps expanded and sequence changed. TDSP 2 / Fleischter TDSP 1 / Groß SMS Iselin / O’Donnell Siemens AG RLL5-310.101.02 Page 1 of 2 UROSKOP D Medical Engineering Rev. 02 01.97 TDSP 1...
Page 26
3 - 2 Changes to previous version This page intentionally left blank. UROSKOP D RLL5-310.101.02 Page 2 of 2 Siemens AG Rev. 02 01.97 TDSP 1 Medical Engineering...