12.4.2 Installation in pipe

Depending on process connections, the sensor is connected to the pipe by means of
flanges, terminals or thread. The order code provides information on the process
connections selected.
Risk of injury due to escaping fluids and damage, if fixing materials are
DANGER
inappropriate or not professionally installed
▶ Fixing materials (screws, nuts, terminals, terminal connectors, gaskets, etc.) are not
The clamp connection must be installed as shown in the figure below.
Clamp connection
Fig. 55: Clamp connection
1
2
3
Fixing the flange
92 / 180
included in the delivery and must be provided by the customer. The operator is
responsible for selecting suitable gaskets and defining corresponding torque values.
Protective materials such as protective stickers or caps on process connections must
be removed immediately before installation.
The direction in which the fluid flows through the pipe is indicated by an arrow on the
flow meter. The sensor must be installed in accordance with the flow direction
indicated to ensure optimal measuring results for density measurements.
Otherwise, the parameter flow direction in the transmitter menu must be changed,
see applicable Function Manual.
Terminal
Gasket
Terminal connector
Use screws and nuts suitable for the flanges.
In case the nominal width of the piping deviates from the flow meter, use the
appropriate reductions.
Inner gasket diameters should not fall below the inner diameters of the flange.
A5E52748387-AA, 2023/09
Sensor installation