Inspection and maintenance
If the value for the calculated fluctuation coefficient is higher than 0.045, the reason for
the deviation must be determined and corrective measures must be taken.
Refer to section "P10 - Fast adaptation" in the service instructions.
QSQ
Fluctuation coefficient C
Steps for calculating C
Example:
The values used in the example are not real values.
Release 10 exposures and record the measured values.
Exposure
(n=10)
1
2
3
4
5
6
7
8
9
10
Add the measured values:
Σ = 1.01 + 1.02 + 1.03 + 1.04 + 1.03 + 1.02 + 1.02 + 1.01 + 1.03 + 1.04 => S = 10.25
Calculate the average value:
Fig. 24:
Calculate the standard deviation:
Fig. 25:
Siemens AG
Medical Solutions
Measured val-
ues
(xi)
1.01
1.02
1.03
1.04
1.03
1.02
1.02
1.01
1.03
1.04
SPR8-215.831.04.03.02
07.05
MOBILETT Plus/ -E/ -M
Page 47 of 64
CS PS 24
47