10 Technical Specification

Operating Modes

non-synchronized (defibrillation on demand)
synchronized (cardioversion)

Energy Selection

by preselection, energy to be delivered into 50 Ohms is indicated as a numerical value:
energy adjustable in steps, energy values as energy delivered into 50 Ohms (internal defibrillation:
energy limited to 50 Joules)
2
5
7
10
20
30
50
100
150
200
300
360 Joules
Autosequence
possible deviation from selected energy less than permitted by IEC and AAMI DF-2

Defibrillator Charging

via capacitor, capacitor charging from battery
capacitor charging time for energy setting of 360 J: typically 10 s

Defibrillation Shock

capacitor discharge via induction coil (Lown / Edmark); exponential pulse shape, monophase, damped
sinusoidal halfwave:
pulse duration for external resistance of 50 Ohms approx. 2.6 ms in compliance with AAMI DF-2
Page 50 of 72
Servicing Manual Dash Responder V1.0
Revision F
2004768-001