General information
1
General information
1.1
Training
•
Training workshop or at least 1 installation
1.2
Required documents
•
Instructions for Use MAMMOMAT 1000/3000 Nova
•
Supplement to the Instructions for Use MAMMOMAT 3000-
Opdimar
(Included in the Opdimar delivery)
•
MAMMOMAT 1000/ 3000/ 3000 Nova - Opdimar
Service Instructions
(Included in the Opdimar delivery)
1.3
Required tools, measurement and auxiliary devices
NOTICE
•
Protective ground wire tester (44 15 899 RV090)
•
Stereo calibration phantom (included in the Opdimar delivery) (part No. 64 30 701)
•
Acrylic plastic (PMMA = Poly-methyl-met-acrylate) plates:
3 plates of AEC calibration plexiglas 150 mm x 150 mm x19 mm (part No. 65 61 232) and
one plate calibration plexiglas, 150 mm x 150 mm x 9,7 mm (part No. 65 61 224)
•
Grid object table
•
Mammography bar pattern (5-20 lp/mm),
for instance Nuclear Associates Resolution Test Pattern (07-555), http://www.nucl.com
•
Cleaning agents* specified in Instructions for Use MAMMOMAT 1000/3000 Nova ,
chapter Cleaning, Disinfection
•
Standard service tools*
Siemens AG
Medical Solutions
All tools, measurement and auxiliary devices listed, with the
exception of those marked with " * ", are specified in the STC
(Service Tools Catalogue).
Register 8
SPB7-230.101.03
Rev. 04
SPB7-230.201...
SPB7-230.203.13...
SPB7-230.061.11 ...
Page 1 of 2
07.03
CS PS 24
1 - 1
Opdimar
Maintenance Instructions