Maintenance; Dicom Bridge; Codonics Printer - Siemens POWERMOBIL SP Maintenance Instruction

Table of Contents

Advertisement

36 Maintenance, Operating Value / Functional Inspection
4-
4
Maintenance, Operating Value / Functional Inspec-
tion
4.1

Maintenance

PMP
Cleaning the system
Clean the entire system:
- Visible cables
- Exterior surfaces
- Contact surface of wheels
- Interior space
PMP
System ventilation
Clean the ventilation slots of the POWERMOBIL and monitor cart.
PMP
POWERPHOS fan
Remove the cover of the POWERPHOS.
Check the visible fan for proper functioning.
Check the fan built into the C-arm for proper functioning. To do this, briefly block the vis-
ible fan mechanically. The fan built into the C-arm must run audibly.
PMP
Memoskop ventilation
Clean the ventilation slots of the Memoskop Fast.
Check that the fan is functioning properly.
4.1.1

DICOM bridge

PMP
Ventilation
Clean the ventilation slots of the DICOM bridge.
Check that the fan is functioning properly.
4.1.2

Codonics printer

PMP
Ventilation
Clean the ventilation slots of the printer and Framegrabber.
Check that the fan is functioning properly.
PMP
Cleaning
POWERMOBIL
SPR2-240.831.01.02.02
07.05
Page 36 of 44
CS PS 24
Siemens AG
Medical Solutions