hit counter script

Advertisement

Quick Links

Notes for operators and responsible maintenance personnel
★ Please read through this Instruction Manual as well as the separate Instruction
Manual "Safety (MN1-5990)" and "Cleaning, Disinfection and Sterilization
(MN1-6161)" carefully prior to use.
★ Keep this Instruction Manual together with the ultrasound diagnostic instru-
ment for any future reference.
© Hitachi, Ltd. 2016, 2017. All rights reserved.
CC41R1 Probe
Instruction Manual
Specification
MN1-6264 Rev.2
i
MN1-6264 Rev.2

Advertisement

Table of Contents
loading

Summary of Contents for Hitachi CC41R1

  • Page 1 ★ Please read through this Instruction Manual as well as the separate Instruction Manual “Safety (MN1-5990)” and “Cleaning, Disinfection and Sterilization (MN1-6161)“ carefully prior to use. ★ Keep this Instruction Manual together with the ultrasound diagnostic instru- ment for any future reference. © Hitachi, Ltd. 2016, 2017. All rights reserved.
  • Page 2 MN1-6264 Rev.2...
  • Page 3 MN1-6264 Rev.2 Introduction This is the instruction manual for CC41R1 probe. The probe is available by connecting to Hitachi's ultrasound diagnostic instrument and can be used as a transrectal probe for observation of prostate and surrounding organs or used as a vaginal probe for observation of uterus and surrounding organs. It can also be used for ultrasound-guided puncture under the condition that the optional puncture adapter is attached to it.
  • Page 4: Table Of Contents

    MN1-6264 Rev.2 CONTENTS 1. General Information ...........................1 1-1. Intended use ...................................1 1-2. Classification of ME equipment ............................1 1-3. Standard components ..............................1 1-4. Options ..................................2 2. Specifications and Parts name ........................3 2-1. Specifications .................................3 2-2. Name of each parts ................................4 3. Preparations before use ..........................5 3-1.
  • Page 5: General Information

    “Cleaning, Disinfection and Sterilization” 1-3. Standard components The standard components of CC41R1 probe are as follows. CC41R1 Probe ······················································· 1 pc Storage case ··························································· 1 set Instruction Manual • Specification (MN1-6264) ······································· 1 copy • Safety (MN1-5990) ··············································· 1 copy...
  • Page 6: Options

    MN1-6264 Rev.2 1-4. Options The following options are available for CC41R1 probe. • Puncture Please use one of the options listed in Table 1 for performing a puncture. Please refer to the section 4-4 for how to attach the sterile puncture adapter.
  • Page 7: Specifications And Parts Name

    At the time of publication of this manual, the connectable diagnostic ultrasound instrument or instrument software version available with this probe is different for each country, please refer to the instrument instruction manual or contact your local Hitachi representative. Field of view: 180°...
  • Page 8: Name Of Each Parts

    MN1-6264 Rev.2 2-2. Name of each parts The name of each part is shown in Figure 2 and the explanation for each part is listed in Table 5. Probe cover Figure 2 Name of each parts Table 5 Name of each part and its explanation Name Explanation Ultrasound is radiated from this part.
  • Page 9: Preparations Before Use

    MN1-6264 Rev.2 3. Preparations before use This chapter describes preparations needed to use the probe safely. Please prepare the probe prior to each use by following the instructions below. 3-1. Visual check Visually check the ultrasonic radiation part, insertion portion, handle, cable, and connector. If any holes, indentations, abrasion, cracks, deformation, looseness, discoloration, or other abnormalities are found, do not use the probe.
  • Page 10: Visual Check For The Sterile Puncture Adapter (Ezu-Pa5V)

    MN1-6264 Rev.2 3-4. Visual check for the sterile puncture adapter (EZU-PA5V) Visually inspect the envelope of the sterile puncture adapter for any break, deformation, crack, or denaturalization. If you find any damage, do not use them and contact a service support immediately.
  • Page 11: Operation

    MN1-6264 Rev.2 4. Operation This chapter describes the operation of the probe, the image direction, and how to attach the sterile puncture adapter, the puncture guide fixture, magnetic position sensor and Magnetic Position Sensor Attachment. 4-1. Operation Mount a probe cover on the probe and insert the probe into the body cavity. An image of the region of interest is displayed on the monitor of the ultrasound diagnostic instrument.
  • Page 12: How To Mount/Remove The Probe Cover

    Mount the probe cover on the insertion portion. If the probe cover is not mounted, residual pathogens on the probe could infect the patient. Use Hitachi-approved probe covers only. Use of non biocompatible probe covers can cause an adverse reaction.
  • Page 13: How To Remove The Probe Cover

    (4) Display the puncture guideline before inserting the needle into the puncture adapter. Refer to the section for the guideline. The needle insertion direction is shown in Figure 6. EZU-PA5V CC41R1 Probe Cover Figure 5 How to attach Sterile Puncture Adapter Biopsy needle is not...
  • Page 14 MN1-6264 Rev.2 Warning Mount a probe cover on the insertion portion. If a probe cover is not mounted, residual pathogens on the probe could infect the patient. Use only biocompatible probe covers. Use of non biocompatible probe covers can cause an adverse reaction. Use only sterile probe covers.
  • Page 15: Display Of Puncture Guideline

    MN1-6264 Rev.2 4-5. Display of Puncture Guideline The puncture guideline can be displayed by dot marks. Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to display the puncture guideline on the image. Guideline Figure 7 Puncture Guideline Warning During puncture operation, display an appropriate puncture guideline on the monitor of the ultrasound diagnostic instrument.
  • Page 16 MN1-6264 Rev.2 Use sterile physiological saline solution as the acoustic medium. Non sterile acoustic medium can cause infection to the patient. Use a needle which is applicable to the puncture adapter. Use of the needle not applicable to the puncture adapter can result in puncture at unintended parts and injury to the patient.
  • Page 17: How To Attach/Release The Magnetic Position Sensor And The Magnetic Position Sensor Attachment

    MN1-6264 Rev.2 4-6. How to attach/release the magnetic position sensor and the magnetic position sensor attachment This section provides how to attach the magnetic position sensor and the magnetic position sensor attachment to the probe. 4-6-1. How to attach the magnetic position sensor and the magnetic position sensor attachment (1) Confirm that the magnetic position sensor attachment is sterilized or disinfected.
  • Page 18: How To Release The Magnetic Position Sensor And The Magnetic Position Sensor Attachment

    MN1-6264 Rev.2 4-6-2. How to release the magnetic position sensor and the magnetic position sensor attachment (1) Release the magnetic position sensor attachment from the probe as shown in Figure 10. (2) Release the magnetic position sensor from the magnetic position sensor attachment as shown in Figure 11. Figure 10 How to release Magnetic Position Sensor Attachment Figure 11 How to release Magnetic Position Sensor from the attachment -14-...
  • Page 19 MN1-6264 Rev.2 -15-...
  • Page 20 +81-3-6284-3668 http://www.hitachi.com/businesses/healthcare/index.html Overseas Offices: Hitachi Medical Systems GmbH Otto-von-Guericke-Ring 3 D-65205 Wiesbaden, Germany EU Importer: Hitachi Medical Systems Europe Holding AG Address: Sumpfstrasse 13 CH-6300 Zug, Switzerland US Importer: Hitachi Healthcare Americas Corporation Address: 1959 Summit Commerce Park, Twinsburg, Ohio 44087 Distributor MN1-6264 Rev.

Table of Contents