Reading out the MOBILETT XP error memory
5Function Check
•
Switch off the system and disconnect the power plug if necessary. Set the XP Hybrid
operating mode switch to "0". Please observe all safety information.
•
Remove the upper cover. Remove the XP Digital upper covers and place them on their
sides on a table or similar surface. Do not disconnect any cables.
•
Move switch SW2B on CPU board D916 to position 2 (service)
•
Switch on the system.
•
Service program Pr 1 is displayed on the control panel. Use the mAs +/- keys to select
service program 3. Read out the error memory and make a note of the data (see the
service programs section of these instructions).
•
Subsequently set switch SW2B of D916 to position 1 (user)