Chapter 1 Safety
Section 1.1 Energy sources
10
EVISION
Table 1-1
Energy source
Electrical
Electrical
Electrical (batteries)
Pneumatic
Hydraulic
Gas/water/steam
Chemical
Mechanical motion
Gravity
Rear suspension spring
Column spring
Thermal
Stored energy (generator
capacitors)
Air under pressure
Oil under pressure
Water under pressure
Gas under pressure
Steam
Other
Type of equipment and/or method selected to dissipate or isolate stored energy:
-
Allow three minutes for stored energy to dissipate. Stored energy is indicated by safety
lamps on the generator Aux boxes.
-
After waiting period, verify dissipation with volt meter.
Type of equipment and/or method used to ensure disconnections:
-
One lock and tag for each electrical power source
-
Multiple locking device
Yes/No
Location of energy
isolating means
Yes
AC power plug
Yes
Circuit breaker
Yes
Battery board
connector J1
No
No
No
No
No
No
No
Yes
Scroll lock at
top of column
No
Yes
Time discharge and
DVM verification
No
No
No
No
No
No
Chapter 1 Safety
O
XR200
PTIMA
I
NSTALLATION
Magnitude of energy
120 VAC/60Hz
220 VAC/50Hz
156VDC
156VDC
n/a
n/a
n/a
n/a
n/a
n/a
n/a
610 N-m (5400 in-lbs)
n/a
15kW generator: 2200J
30kW generator: 8800J
n/a
n/a
n/a
n/a
n/a
n/a
/XR220
AMX
AMX
M
ANUAL
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
Page 21