D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 1 . FOREWORD Please refer to this manual for detailed instructions on warnings, cautions, specifica- tions, and additional information. Important: Users should read this entire manual before operating the Zen-O lite™ Port- able Oxygen Concentrator. Failure to do so could result in personal injury and/or death. If you have questions about the information in this user manual or about the safe opera- tion of this system, contact your distributor.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN • ISO 7637-2:2011 Road Vehicles - Electrical disturbances from conduction and coupling- Part 2: Electrical transient conduction along supply lines only. • Medical Device Directive 93/42/EEC. This equipment is classified as: • Class II •...
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 3 . SAFETY INSTRUCTIONS WARNINGS OVERVIEW 3 . 1 . 1 . The device must be used in the carry bag to provide protection from liquid intrusion from rain and/or spills. 2 . There is a risk of fire associated with oxygen equipment and therapy. Do not use near sparks or open flames.
Page 10
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 15 . Do not use lubricants on this device or any of its accessories. 16 . Use of this device at an altitude above 4,000 m (13,000 feet), or outside the tempera- ture range of 5°C (41°F) to 40°C (104°F), or outside the humidity range of 5% to 93% may adversely affect the flowrate and percentage of oxygen and consequently the quality of therapy.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 3 .2 . CAUTIONS OVERVIEW Keep away from heat sources (fireplaces, radiant heaters, etc.) that could cause the operating temperature at or near the device to exceed 40°C (104°F). 2 . The display may be difficult to read under bright lighting conditions (sunlight, interior lights, etc.), move away from direct light for viewing the display.
Page 12
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 16 . Always follow the cannula manufacturer’s instructions for proper use. 17 . Replace the cannula on a regular basis. Check with your home oxygen provider or clinician to determine how often the cannula should be replaced. 18 .
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 3 .3 . IMPORTANT INFORMATION If an extension cord is necessary, use a UL listed 15 amp or higher cord. Do not con- nect any other devices on the same extension cord. Do not use a multisocketed extension cord.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 5 . PRODUCT DESCRIPTION 5 . 1 . SCHEMATIC DESCRIPTION Breath detect LED Alarm mute indicator Red LED - warning Display screen Green LED - power ON Yellow LED - warning Pulse setting Power button selection buttons Battery indicator...
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 6 . GENERAL INSTRUCTIONS BEFORE USE A variety of accessories can enhance the portability and use of the Zen-O lite™ Portable Oxygen Concentrator. In addition to the device, the package contains accessories to get started and a user manual. Contact your home oxygen provider for a complete list of available accessories.
Page 16
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 6 .2 . BATTERY Zen-O lite™ Portable Oxygen Concentrator can always be used when directly connected to a power source. However, to enhance its portability, the concentrator is equipped with a rechargeable lithium-ion internal battery. WARNING: Always ensure the battery is inserted before using this device .
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN IMPORTANT: Ensure power status icon (see Fig. 7) indicates power is connected. If not, check that cord is plugged in completely (see Chapter 10 - Trouble shooting for more information) To maximise battery life and run time, avoid letting the battery deplete and use while connected to a power source whenever possible.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN CAUTION: Only use approved accessories with this device. Refer to the approved ac- cessories guide for a complete list of accessories and cannula approved for use with this device. Using unapproved accessories or cannula may impair the performance of this device, including flow rate or oxygen purity.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN 7 .2 . TURNING ON • To turn the device on, press the power button • The concentrator will chirp and the green, yellow, and red LEDs will flash once, while the screen displays the device name. Red LED - indicates a warning danger and/or a need for urgent action Yellow LED - indicates caution or attention required Flashing green LED - indicates oxygen is being delivered.
D.SPEC.26.4 Zen-O lite - Mode d'emploi_EN Zen-O lite™ operating on battery Fig. 7 Zen-O lite™ running on external power. power. 7 .5 . ECO MODE While the Setting screen is shown, press and hold the Battery button for approximately four seconds until it changes from “Pulse XX” to “Eco Mode”. Press and hold the Battery button again to return to Pulse Mode.