Alternatively, the wireless connection can be set up via a network scan:
Press "Scan" in the Wireless Overview tab and chose the desired WiFi network by selecting join.
Enter the password in the Joining Network tab and confirm by pressing Submit (Pic.16). Save the WiFi settings by selecting
SAVE & APPLY in Associated Stations tab (Pic.15).
6. Configuration of the NANOCOLOR
This chapter describes the configuration and connection of the spectrophotometer.
Make sure to perform the installation steps in
ration. A successful connection to the mount should be given before you proceed with this chap-
ter.
6.1 Network configuration
Connect the spectrophotometer to the eth1 interface of the Network Adapter with the help of a LAN cable (CAT 5E).
For the configuration of the NANOCOLOR
The menu to enter the IP addresses can be reached via Main menu
The following settings and conditions are required:
Software version: ≥ 1.10.1
Operating system: ≥ 7.0.2864.2
IP setting: static (the spectrophotometer will perform an automatic restart after confirming the setting)
IP Address: 192.168.2.200
Subnet Mask: 255.255.255.0
Pic.16: Joining WiFi after a network scan
®
spectrophotometer
NOTICE
chapter 4
before starting the photometer configu-
®
UV
VIS II or
/
II, refer to the manual of the instrument.
VIS
Connectivity
Network
LAN.
14/21