Storing The Error Log In Ascii Format; Prerequisites - Siemens ARCOSKOP Troubleshooting Manual

Table of Contents

Advertisement

4 - 4

Storing the error log in ASCII format

When storing the error log using the backup function of the service program, it will be
stored as a binary file which cannot be read by a text editor.
For easier evaluation of the error log, it can be stored in ASCII format as well.
For easier service support from the USC or HSC, please send the error log as an ASCII
file.
NOTE

Prerequisites

The service PC is connected and the service program is started.
Select the "Diagnostics..." menu.
Select the "Error log" menu.
Click the "Get from unit" button. Wait until the error log is transferred to the service PC.
Click the "Export to file" button.
Save the file to your hard drive.
For the file name, please enter the current date, the part number and the serial number of
the unit as described in the example.
Example:
The current date is (DD-MM-YY): 29-03-01
The unit part no. is: 4774019
The unit serial no. is: 01234
Enter this file name: 22-03-01_4774019_01234.txt
Save the file to the hard drive of your service PC under a path of your choice.
Exit the service program.
Copy the previous stored file to an emty, formatted floppy disk.
ARCOSKOP
System Manual
The parameter disk must be in drive A before starting the service
program and during service work. Otherwise, the text "no error
text available" will display behind the error codes instead of the
actual error messages.
SPR2-250.840.01