2 - 2
1 Required reference documents
- DHHS Test certificate 1a+ and Acceptance Test Protocol (in Register 3 of the
Logbook)
2 Required test equipment
- PTW-Diados
- Fluke
- Calculator
3 Safety Notes
CAUTION
4 Notes for the Maintenance Instructions
- Compliance with the applicable sections of the Federal Performance Standards is
assured when all procedures described in these instructions have been performed and
tolerance values are within the specified range.
- The unit movements required for these procedures must be performed cautiously to
prevent the unit from being damaged in the event of a malfunction.
WARNING
- When performing safety checks or adjustments with radiation ON, radiation protection
clothing must be worn.
4.1 Calibration of the option Integrated Dose Measurement Device
A calibration of the optional installed Integrated dose measurement device
(air kerma measurement device or area dose product measurement device)
is not necessary.
4.2 Information about air kerma strength determination
The air kerma strength is determined with a measuring chamber integrated into the sys-
tem. The reference location for determining the air kerma strength is 30 cm in front of the
image intensifier input (70cm distance from the focal spot).
NOTICE
ARCADIS Varic
General information and overview
Serial-No.: ..........
Serial-No.: ..........
All local safety regulations must be observed.
Checks or adjustments to be performed with radiation ON are
marked with the radiation warning symbol
This value is used because in typical applications the object to be
examined is located approx. 30cm in front of the I.I.
SPR2-310.662.01
Rev. 02
Date of calibration ..........
Date of calibration ..........
Page 2 of 2