Summary of Contents for Siemens MULTIMOBIL 5C

  • Page 1 MULTIMOBIL 5C Pre-Installation © Siemens AG 2004 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility...
  • Page 2 Pre-installation Instructions Multimobil 5C Multimobil 5C Pre-installation Version 5.0 Copyright © SIEMENS LTD., Med INDIA Siemens Ltd. Med India Version 5.0 Page 1 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 3: Table Of Contents

    ..........................7 ACKAGING ..........................8 NLOADING MM5C......................8 NPACKING THE ................... 10 NPACKING THE MONITOR TROLLEY ..................... 11 NSTALLATION ONITORS Siemens Ltd. Med India Version 5.0 Page 2 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 4: Pre-Installation

    Anti-isowatt 0.5 to 5 mA Contrast 0.5 to 7mA Fluroscopy Power 550W continous Timer Fluoroscopy time display with timer Siemens Ltd. Med India Version 5.0 Page 3 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 5 Continuous operation Focus-II distance 860mm for 9’’ system 900mm for 6’’ system Max. cassette size at 900 10’’X12’’ cm SID Siemens Ltd. Med India Version 5.0 Page 4 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 6 II Chain Resolution Nominal 1.2 lp/cm Zoom 1.6 lp/cm for 9’‘ system Nominal 1.4 lp/cm Zoom 2.0 lp/cm for 6’‘ system Siemens Ltd. Med India Version 5.0 Page 5 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 7 EN 60601-1:1990 Standards EN 60601-2-7:1998 EN 60601-1-3:1994 EN 60601-1-2:2001 Compliance to AERB Type Approval Certification Compliance to BIS test certification. Siemens Ltd. Med India Version 5.0 Page 6 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 8: Information On Room Planning

    Humidity in range of 30% to 95% RH provided no condensation occurs. 1.3 Packaging The Multimobil 5C unit is delivered in two wooden boxes. One box consists of the Monitor trolley. This is a wooden box with the trolley mounted on the wooden base and is fixed using clamping brackets.
  • Page 9: Unloading

    The complete system must be inspected for physical damage and / or material short-fall and factory must be informed accordingly. The Multimobil 5C consists of fragile components like camera, Image intensifier, X-ray Tube etc. Sufficient care must be taken while unloading / unpacking the unit. Refer the unpacking instructions provided on the packing.
  • Page 10 15) Now the unit will be resting on the wooden platform. Release the brakes and push the unit down to the ground through the ramp. Siemens Ltd. Med India Version 5.0 Page 9 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 11: Unpacking The Monitor Trolley

    8) Lift the trolley gently by holding at the base and supporting it from top. Place it on the floor. 9) Release the brakes and the trolley can be shifted to the OT. Siemens Ltd. Med India Version 5.0 Page 10 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 12: Installation Of The Monitors

    4) Make use of the connector provided on the monitor trolley for providing power supply to the monitors. 5) Ensure that the radiation indication lamp is available on the Monitor Trolley. Siemens Ltd. Med India Version 5.0 Page 11 of 11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...