Page 1
Voluson® P8 / Voluson® P6/ Proprietary Service Manual • Voluson® P8 systems • Voluson® P6 systems Document Number: 5459672-100 Revision: 6 0459...
Page 2
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND (DE) NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 3
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
Page 4
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL VAROITUS (FI) ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ (EL) FIGYELMEZTETÉS (HU) • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ...
Page 5
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL VÝSTRAHA (CS) • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
Page 6
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL ĮSPĖJIMAS (LT) • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
Page 7
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL ОСТОРОЖНО! (RU) UPOZORNENIE (SK) • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA...
Page 8
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL (JA) (ZH-CN) (KO)
Page 9
GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 10
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of Voluson®...
Page 11
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Revision History Revision Date Reason for change June 2012 Initial release August 2012 Included instructions to convert low version to high version...
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 1 Introduction Overview............1 - 1 Purpose of Chapter 1 .
Page 13
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 2 Site Preparation Overview ............2 - 1 Purpose of Chapter 2 .
Page 14
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
Page 15
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Scanner Shutdown ........3 - 27 Transducer Connection .
Page 16
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Stand-alone Voluson® P8 / Voluson® P6 ..... . 3 - 66 Voluson®...
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
Page 18
Control Console Positioning ........4 - 49 Assembly procedure for converting Voluson P6/P8 from Low Version to High Version .
Page 19
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
Page 20
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL SonoNT ..........5 - 17 Description of Hardware Options .
Page 21
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Power Distribution Block Diagram ......5 - 40 Power Up Sequence Description .
Page 23
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
Page 24
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter Voluson®...
Page 25
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Apertur Lateral Min ..........7 - 36 Check for missing channels or elements.
Page 26
No Connection to the Network at All ......7 - 59 GE remote service connection ....... 7 - 59 Tech Tips .
Page 27
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
Page 28
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Manpower ..........8 - 23 Tools .
Page 29
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Replacement of SIDE COVER L ASSY ........8 - 40 Manpower .
Page 30
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Tools ............8 - 52 Preparations .
Page 31
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Manpower ..........8 - 65 Tools .
Page 32
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Removal Procedure ......... . 8 - 77 Installation Procedure .
Page 33
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Replacement of FAN with Cable ASSY II ....... . 8 - 104 Manpower .
Page 34
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
Page 35
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview ............10 - 1 Periodic Maintenance Inspections .
Page 36
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Meter Procedure ......... . 10 - 20 Enclosure Leakage Current Test .
Page 37
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL xxvi Table of Contents...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson® P8 / Voluson® P6 and/or Voluson®...
ANUARY ROPRIETARY ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • GE Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 Models Covered by this Manual Table 1-2 Voluson®...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Voluson® P8 / Voluson®...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
Page 43
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 1-5 Product Icons Voluson® P8 / Voluson® P6 LABEL/SYMBOL PURPOSE/MEANING LOCATION "Mains OFF" Indicates the power off position of the mains Rear of system at mains switch power switch.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-4 Electrical Safety 1-3-4-1 Safe Practices To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
Page 46
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-5 Auxiliary Devices Safety (cont’d) Everybody who connects additional equipment to the signal input portion or signal output portion configures a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 60601-1-1.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-6 Labels Locations The Voluson® P8 / Voluson® P6 comes equipped with product labels and icons.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-6-1 Main Label . The Main Label is located on the rear of the Voluson® P8 / Voluson® P6 system.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-6-2 UL Label The UL label is located on the rear of the system. Figure 1-3 UL Label...
ERVICE ANUAL 1-3-6-3 Multi Caution Label The Multi caution label is located on the rear of the Voluson P8/Voluson P6 system Figure 1-4 Multi-Caution Label Symbol “Follow instructions for use” Symbol “Do not use the following devices near this system”...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-3-9 Returning/Shipping System, Probes and Repair Parts (cont’d) NOTE: The US Department of Transportation (DOT) has ruled that “items what were saturated and/or dripping with human blood that are now caked with dried blood;...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL This page was intentionally left blank. Chapter 1 - Introduction 1-17...
Page 55
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1-18 Section 1-5 - Customer Assistance...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson® P8 / Voluson®...
The Voluson® P8 / Voluson® P6 is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE- specified Cable length, material, and routing are all important; do not make any changes that do not meet harnesses and peripherals.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2-2-4 Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C.
The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements. Dedicated Power Outlets VOLUSON P6/P8 Hospital Network Cabinet for Software and Manuals (optional) 35.5 IN.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2-3-4 Networking Setup Requirements 2-3-4-1 Stand-alone Unit (without Network Connection) None 2-3-4-2 Unit Connected to Hospital’s Network Supported networks: •...
Page 65
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the Voluson® P8 / Voluson® P6. Included are procedures to receive, unpack and configure the equipment.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the Voluson® P8 / Voluson® P6 weighs approximately 70 kg (154.3 lbs.) without peripherals, two people are required to unpack it.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-2-2-4 Brake Pedal Operation REMEMBER: If the front wheel brakes are engaged for transportation, release brake...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment Please read this section carefully before unpacking the Voluson® P8 / Voluson® P6 and its CAUTION (optional) peripherals.
Page 70
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) The envelope with delivery address, packing list and invoice is located on the front panel of the crate.
Page 71
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Remove the top cover and top plate. Remove the three Plastic Joints from the outer sleeve..
Page 72
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Cut the tape to unfold the package and remove the antistatic vinyl cover Disengage the brakes, grasp the system at the handle, then slowly move unit down the ramp.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-4 Preparing for Set Up 3-4-1 Verify Customer Order 1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have been received.
Page 74
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices NOTE: Normally auxiliary devices and peripherals come pre-installed with the system.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-1 Connecting the LCD Monitor NOTE: The LCD monitor comes pre-installed with the system. Figure 3-5 Connection Scheme - LCD Monitor...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-2 Connecting the Black & White Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-27.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-3 Connecting the Color Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-27.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-4 Connecting the DeskJet Color Printer 3-5-4-1 Connection via Bluetooth Adapter The DeskJet Color Printer can be connected to an external, non-isolated power source. The bluetooth...
Page 82
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-4-1 Connection via Bluetooth Adapter (cont’d) Please observe that the complete Bluetooth Printer Assembly has to be located outside of the CAUTION patient environment (according to IEC 60601-1 / UL 60601-1).
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-5 Connecting the Wireless Network Adapter The Wireless Network Adapter can be connected to any accessible USB port of the Voluson® P8 / Voluson®...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-6 Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the Voluson® P8 / Voluson®...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-7 Connecting the USB Flash Memory Stick NOTICE Before connecting an USB device, please read General Remarks and Hints when using external USB- Devices on page 3-21.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-5-9 General Remarks and Hints when using external USB-Devices Do not connect or disconnect any external USB-devices to or from the system while...
Page 87
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL To stop the external device, select it and then click the STOP DEVICE button. Figure 3-12 Device can now be safely removed Confirm the ‘Stop Device‘...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-6 Completing the Set Up 3-6-1 Connecting the Unit to a Power Source The connection of the Voluson® P8 / Voluson® P6 ultrasound unit to a power source should be performed by a qualified person who has completed basic Voluson®...
Page 89
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Switch ON the Circuit at the rear of the system. Circuit Main Power Figure 3-13 Circuit and main power cable at rear of system...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-6-2-2 Back End Processor Boot Up When AC power is applied to the scanner, the ON/OFF button on the control panel illuminates amber, NOTICE indicating that the System (including the Back-end Processor) is in Standby mode.
Figure 3-16 below. Figure 3-16 GE Healthcare wallpaper with progress bar 6.) The software initiates and sets up the Front-End electronics and the rest of the scanner. 7.) The Keyboard backlight is lit. 8.) As soon as the software has been loaded, the 2D screen is displayed on the monitor.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-6-3 Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again.
Page 93
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF button.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-6-4 Transducer Connection NOTE: When the probe is connected, it is automatically activated. Once connected, the probes can be selected for different applications.
The Bluetooth Printer Connection set as well as the Color Deskjet printer MUST NOT be installed CAUTION by the user! !! CAUTION: For installation please contact your local distributor or GE service representative. 3-7-1 Installing Digital Black & White Printer Sony UP-D897 1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-27.
1.) Turn ON the system and wait till the system has booted. 2.) Connect the USB "standard GE" service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up. Remove the dongle.
Page 97
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8.) Confirm the following dialog window with NEXT. The "License Agreement" window appears: Figure 3-21 Welcome to Install... and License Agreement 9.) Select "I accept the terms in the license agreement"...
16.)Remove the DVD, close all open windows and restart the system (turn off and on the system). 3-7-3-2 Configuration of the Bluetooth Software 1.) After the system has restarted, connect the USB "standard GE" service dongle on the front USB connector. 2.) Enter the "Standard GE Revolving" password, click ENABLE.
Page 99
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4.) Press CTRL + ESC to open the Windows Start menu and select "My Bluetooth Places".
Page 100
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: Display of this window might be different. It depends on installed Bluetooth Adapter. deselect ALL services...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-3-3 Prepare the Printer and connect the Bluetooth Printer Adapter NOTICE Preparations as well as Printer Adapter connection depend on the used Bluetooth Line Printer.
Page 102
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Select "View or modify configuration". Add a Bluetooth Device View or modify configuration Figure 3-30 My Bluetooth Places “View or modify configuration”...
Page 103
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 6.) Remove check mark at "Secure Connection" and then click OK. Figure 3-32 "Client Applications" page / remove checkmark at "Secure Connection"...
Page 104
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL b.) If it looks like the right image, enter the passkey (0000) and then click on NEXT.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 13.)If the "Add Printer Wizard" window appears, select the HAVE DISK button. [Browse] for Printer Driver on F:\ (where F: stands for the system´s DVD drive)
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-3-5-1 HP Officejet H470 - Printer Properties After driver installation, the “HP Officejet H470 series BT Properties” page appears.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-4 Printer Installation manually 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 108
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) 7.) Select the corresponding Printer Port (e.g., Figure 3-41 = USB001) and click NEXT.
Page 109
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) Figure 3-43 Select Printer Driver path (R:\system\periph\..) 10.)Click OPEN, select the “xxx.inf” file and click OPEN again.
Page 110
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) 13.)Assign a name, decide if printer should be used as default printer and confirm with NEXT.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-5 Adjustment of Printer Settings 1.) After system restart, press the F2 key on the control panel.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-5-3 UP-D25MD - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings"...
Page 116
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-5-3 UP-D25MD - Printer Settings (cont’d) 3.) Select the GRAPHICS page. From the “Color Adjust” pop-up menu select: a.) Color Balance: Cyan = 0;...
Page 117
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-5-3 UP-D25MD - Printer Settings (cont’d) e.) Adjust Value : Select nomal. Figure 3-54 Graphics page (Adjustment by color) 4.) For saving the adjusted printer settings click APPLY and then OK.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-7-6 Remote Control Selection To assign an auxiliary device (e.g., printer) to the remote keys P1, P2, and/or P3/DVR 1.) Press the F2 key on the control panel.
• Patient Info Display: Drop Down Management, Capitalize Letter in Patient Names, etc. • Service tab: conncet the USB “standard GE” dongle and enter the “standard GE revolving” password (change every six months_ to get access to the Service Tools functions.
Page 120
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL right trackball key (left mouse button): sets, fixates markers and activates pages/buttons etc. marked by the pointing device...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-8-1-2 How to enter Date and Time 1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-54.
If password is unknown click EMERGENCY. This enables standard - but limited - operation. NOTICE The Screen Lock password cannot be reset by the user! Please contact your GE service representative. !! NOTICE: 5.) Click SAVE&EXIT to save Settings and exit System Setup.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL PERIPHERALS tab. 3.) Select desired function of the Footswitch Left, Middle and Right. Figure 3-58 Click SAVE&EXIT to save Settings and exit System Setup.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-8-2 Measure Setup Modifications of system parameters are supported by diverse dialog pages and windows on the...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-8-3 On-Board Optional Peripherals Mains outlet: AC outlet for on-board peripherals Output voltage : 100~250V. Modification of voltage setting only by an authorized service person!
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-8-4 External I/O Connectors Figure 3-59 External I/O Connectors - on Rear of System Table 3-9...
Page 127
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 3-12 Network Connector, RJ45 Modular 8 Pin Pin No Output Signal Description ETHER TD Ethernet RD+...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-8-4-1 External I/O Pin Outs (cont’d) Table 3-13 USB Connectors Pin No Output Signal Description USB Power Supply...
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-11 Connectivity Setup The Voluson® P8 / Voluson® P6 can be connected to various connectivity devices.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-11-1-2 Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-11-1-3 Stand-alone Voluson® P8 / Voluson® P6 If digital images or 3D/4D data sets are stored, they should be saved in the Archive (Image Management System software).
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-11-2 Wireless Network Configuration To configure the Voluson® P8 / Voluson® P6 ultrasound unit to work with WLAN, the hospital’s NOTICE network administrator has to provide the required information.
Page 133
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-11-2 Wireless Network Configuration (cont’d) 5.) Select the country where you will use the wireless USB adapter from the “pull-down menu” and then click AGREE.
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-11-2 Wireless Network Configuration (cont’d) After a few seconds waiting, following information appear in the status bar at the bottom of the page.
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 3-12 Network IP Address Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson®...
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-12-1 Map Network Drive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 137
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3-12-1 Map Network Drive (cont’d) 7.) Select the CONNECT button to establish the connection to the remote machine.
Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 - Setup Instructions...
Page 141
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL This page was intentionally left blank. 3-76 Section 3-14 - Paperwork...
Page 142
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson® P8 / Voluson®...
Page 143
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place.
Page 144
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 5.) Hold down the On/Off button (see: Figure 4-2) on the control panel for ~3 seconds.
Page 145
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-3-2 Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again.
Page 146
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 6.) Press on the brakes to block the front caster wheels. 7.) Disconnect probes. DO NOT disconnect a probe while running (Live Scan “Write” mode)! CAUTION A software error may occur.
Page 147
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-3-3-2 Monitor Display Figure 4-6 Monitor Display Tour Table 4-2 Monitor Display Features 1. Logo 15. Measurement results 2.
Page 148
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, following pages will familiarize you with image optimization for: •...
Page 149
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-1 2D Mode Checks Figure 4-7 2D Main and 2D Sub Menu Table 4-3 2D Mode Functions...
Page 150
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results Focus Depth To select the depth position of the actual focus zone(s). Arrows at the left edge of the 2D-Image mark the active focal zone(s) by their depth position.
Page 151
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results While scanning a certain number of frames (2D images of the last CINE Mode examination sequence), this will be stored in the cine memory automatically.
Page 152
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results TINT MAP A Tint map determines the Chroma value (color tone and saturation) of an echo in relationship to its amplitude.
Page 153
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-2 M Mode Checks NOTE: The AMM menu is only displayed if the Anatomical M-Mode option is installed.
Page 154
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-4 M Mode Functions Step Task Expected Results ENHANCE Due to this function a finer, sharper impression of the image is produced.
Page 155
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-3 Spectral Doppler Mode Checks NOTE: Different menus are displayed depending on which Spectral Doppler Mode (PW or CW) is selected.
Page 156
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Step Task Expected Results Display Format Doppler spectrum is displayed below the B mode image or on its right side.
Page 157
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-4 Color Doppler Mode Checks NOTE: Different menus are displayed depending on which Color Doppler Mode (CFM, PD, HD-FLow) is selected.
Page 158
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-6 Color Doppler Mode (CFM, PD, HD-Flow, TD) Functions Step Task Expected Results After FREEZE you can adjust the Color Threshold. It eliminates small color noise or motion THRESHOLD artifact signals in the color image.
Page 159
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-5 Volume Mode Checks NOTICE 4D Real Time 4D, VCI, VCI Omni View are Options. If options are not part of the system configuration, checks can be omitted.
Page 160
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-5-1 Pre-Volume Mode Functions 1. Select the desired Volume Mode: The Volume Mode function is switched on, the “3D Pre” respectively “4D Pre”...
Page 161
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-7 Pre-Volume Mode Functions Step Task Expected Results • 4D REAL TIME Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered images Volume Contrast Imaging - improves the contrast resolution and the signal / noise ratio and •...
Page 162
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-5-2 Functions after the Acquisition Figure 4-14 Sectional Planes and Image Rendering Table 4-8 Functions after the 3D Acquisition...
Page 163
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-8 Functions after the 3D Acquisition Step Task Expected Results Speckle Reduction Imaging (SRI) to reduce speckle which interferes with the SRI / SRI 3D sectional planes or rendered 3D image.
Page 164
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-5-3 3D/4D Sub Menu Figure 4-15 3D/4D Sub Menu Table 4-9 Sub Menus Step Task Expected Results PROBE ORIENTATION Adjust and activate the display of direction markers at border of the image.
Page 165
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-6 Using Cine 4-4-6-1 Activating Cine Press FREEZE, then roll the TRACKBALL to display the images of the stored sequence one by one.
Page 166
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-6-8 Activating Cine Calc 1.) After 3D Volume or Real Time 4D acquisition select the CINE CALC key on the.
Page 167
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-7 Generic Measurements NOTE: Different menus are displayed depending on: • the currently selected Application and Display Mode, •...
Page 168
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: To exit Generic measurements, select the EXIT key on the Voluson® P8 / Voluson® P6, press the MEASUREkey or the EXIT key on the control panel.
Page 169
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: To erase the results, select the CLEAR key on the control panel. 4-4-7-3-2 Measurements in Spectral Doppler Mode NOTE: The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis).
Page 170
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-8 Calculations The Voluson® P8 / Voluson® P6 system supports calculation packages and application-oriented patient Worksheets (Reports) for following applications: •...
Page 171
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-8-1 Worksheet (Report) Pages Press the REPORT key on the control panel to view the “application dependent” patient worksheet pages that contain the results of calculations and measurements.
Page 172
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-9 Probe/Connectors Usage 4-4-9-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case.
Page 173
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-10 Patient Archive (Image Management) The Voluson® P8 / Voluson® P6 provides an Patient/Image Management System that allows fast and extremely easy patient, exam and image management.
Page 174
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-4-11 Erasing DVD/CD 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 175
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images It is recommended to Backup the Full System Configuration...
Page 176
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-1 Save Image Settings Only The Image Settings contains: • Application Settings • User Programs •...
Page 177
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-2 Load Image Settings Only The loading procedure overwrites the existing image settings on the local hard drive.
Page 178
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-2-2 Load “Complete Backup” NOTE: Following procedure should be used, if the user prefers factory default settings, which are adapted for the installed Application Software version.
Page 179
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-2-3 Load only parts of the “Complete Backup” NOTE: Following procedure should be used, if the user has saved own User Programs, 3D/4D Programs or AutoText entries, but new presets have to be added to match the installed Application Software version (e.g., User Programs and 3D/4D Programs for new probe, etc.).
Page 180
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-3 Save Full System Configuration (Full Backup) A backup of the Full System Configuration always contains the following data •...
Page 181
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-3 Save Full System Configuration (Full Backup) (cont’d) When the “Full Backup” is saved on a network drive (to map a network drive see: Section 3-12-1 "Map...
Page 182
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) It is recommended to backup data before an upgrade; see: Section 4-5-3 on page 4-39.
Page 183
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) (cont’d) 1.) Press the F2 key on the control panel.
Page 184
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-4 Load Full System Configuration (Full Backup) (cont’d) 9.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure.
Page 185
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-6 Archiving Images It is recommended to Save Image Archive 4-44) and Backup the Full (Section 4-5-6-1 on page...
Page 186
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-6-1 Save Image Archive (cont’d) 5.) Choose archiving span time from the pop-up menu. If for example “All images older than” 1 Day is chosen (see...
Page 187
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-5-6-2 Load Image Archive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 188
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 4-6 Software Configuration Checks 4-6-1 System Setup Press the F2 key on the control panel and then select SYSTEM SETUP.
Page 189
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-6-2 Measure Setup Press the F2 key on the control panel and then select MEASURE SETUP.
Page 190
The control console can be adjusted to a fixed height (910 mm) with an Extender Kit (H46882LG). 4-7-2-1 Assembly procedure for converting Voluson P6/P8 from Low Version to High Version Table 4-15 Part list of LCD cable and LCD Arm cover...
Page 191
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 1.) Remove the Monitor rear cover Monitor Rear cover Figure 4-37 2.) Remove the LCD cable...
Page 192
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 4-16 Part list of Speaker plastics GEPN Description 5433007 COVER LEFT SIDE, SPEAKER, PLASTIC 5433008...
Page 193
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Remove keyboard base fasteners (5459100, OPIO CAPTIVE SCREW TETRIS) Figure 4-41 Keyboard base fasteners 6.) Remove 4 Probe holders from the keyboard assembly...
Page 194
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Pull it downwards Figure 4-43 Keyboard assembly 8.) Remove upper Keyboard cover Upper keyboard cover Remove 4 screws 2159625, SCREW W/SP PH M4X8 Figure 4-44 Keyboard cover 9.) Remove the cable tie...
Page 195
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Cable tie 2160955 Figure 4-45 Cable tie 2160955 10.)Pull the LCD Cable through the LCD arm and Monitor holder...
Page 196
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Keyboard lower plastic Keyboard assembly USB cables OPIO cable Power cable Figure 4-47 OPIO, Power and USB cables 12.)Disconnect Speaker cables...
Page 197
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 14.)Disconnect the Power Cable and USB Cables from Keyboard upper cover USB cables Power cable Figure 4-50 USB and power cables 15.)Disassemble the keyboard upper and lower covers...
Page 198
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 16.)Hold the cast base firmly and lift it up as shown in the images below Hold the cast base firmly and lift it...
Page 199
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Assemble Extender as shown and pull all the cables through the Extender Extender Fasten 4 screws, 4 spring...
Page 200
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 19.)Assemble the keyboard base Fasten 4 screws, 4 spring washers and 4 flat washers Figure 4-56 Assemble the keyboard base 20.)Re-Assemble LCD Cable...
Page 201
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 22.)Route the LCD cable as shown below. Insert the connector in the slot and fasten the screws.
Page 202
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 24.)Connect the ground cable Ground cables, fasten 2 screws Figure 4-60 Ground cables 25.)Connect the Power cable and USB Cables...
Page 203
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Fasten 2 screws and assemble the keyboard upper cover Figure 4-63 Keyboard upper cover Connect OPIO, Power and USB cables 28.)Reassemble Keyboard Lower cover...
Page 204
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 29.)Reassemble 4 probe holders Figure 4-65 4 probe holders Figure 4-66 Fasten 9 screws 30.)Reassemble left and right speaker covers...
Page 205
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Insert 2 rubber caps Fasten the screw Assemble clamp and fasten a screw Route the LCD cable...
Page 206
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4-7-3 Brakes and Direction (Swivel) Locks Check the brakes and swivel locks function as described below.
Page 207
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 4-8 Site Log Table 4-20 Voluson® P8 / Voluson® P6 - Site Log (Paper Documentation)Voluson® P8 / Voluson®...
Page 208
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson® P8 / Voluson® P6’s system concepts, component arrangement, and subsystem function.
Page 209
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-2 General Information Voluson® P8 / Voluson® P6 is a digital beamforming curved-, linear- and phased array ultrasound imaging system.
Page 210
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-2 General Information (cont’d) Among other significant features of the Voluson® P8 / Voluson® P6 are the following:...
Page 211
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-2 General Information (cont’d) Figure 5-2 Basic Block diagram of Voluson® P8 / Voluson® P6...
Page 212
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-2 General Information (cont’d) The Voluson® P8 / Voluson® P6 used digital beamforming technology which provides high resolution and high penetration performance.
Page 213
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-1 Description of Voluson® P8 / Voluson® P6 Operating Modes 5-2-1-1 B-Mode or 2D-Mode B-Mode or 2D-mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes.
Page 214
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-1-3 Color Doppler Modes Color Doppler is used to detect motion presented as a two-dimensional display.
Page 215
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-1-4 Pulsed (PW) Doppler PW Doppler processing is one of two spectral Doppler modalities. In spectral Doppler, blood flow is presented as a scrolling display, with flow velocity on the Y-axis and time on the X-axis.
Page 216
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-1-8 3D Rendering The 3D (volume) rendering is a calculation process to visualize certain 3D-structures of a scanned volume by means of a 2D-image.
Page 217
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-2 Block diagram Voluson® P8 / Voluson® P6 Figure 5-4 Voluson® P8 / Voluson® P6 - Block diagram...
Page 218
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-2-1 Front End Block diagram For further details : see: "Front End Processor - Block diagram" on page 5-20 see: "PIR - Blockdiagram"...
Page 219
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-3 Data Flow Control Description This section describes the functions of the boards vs. different operation modes.
Page 220
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL b.) XBEAM CRI (CrossBeam - Compound Resolution Imaging): Does not need any special functions of RFS.
Page 221
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-3-5 CFM-Mode (Color Flow Mode) 1.) RFS PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12 After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
Page 222
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-4 Description of Software Options To activate the software options: 1.) Press the F2 button on the OPIO.
Page 223
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-4-1 3D/4D Advance 5-2-4-1-1 Real Time 4D Real Time 4D mode is obtained through continuous volume acquisition and parallel calculation of 3D rendered images.
Page 224
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL tissue scatters, image speckle characteristics such as brightness, density or size have no apparent value in determining tissue structure or related properties. The elimination of or significant reduction in speckle improves the quality or diagnostic potential of the image.
Page 225
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-2-5 Description of Hardware Options Table 5-3 Hardware Options HW-Options Description CW-Doppler 5-2-5-1 CW - Continuous Wave Doppler...
Page 226
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Printer Drivers LINUX partition: (not visible in Windows) Linux operating system for rescue functionality Figure 5-5 Data Location - distribution of partitions...
Page 227
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-3 FrontEnd Processor Figure 5-6 Front End Processor - Block diagram Table 5-4 below outlines the Voluson® P8 / Voluson® P6 Front End components described in the sub- sections.
Page 228
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-3-1 PIR - Probe Interface board Switches the Probe Connectors (3 RS -Connectors) and recognizes Probes •...
Page 229
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-3-2 BF64 - Beam former board BF64 board has 64 TX/RX channels, respectively. beamformer board is a complete 64 channel ultrasound Front-end that includes the following features and functions:.
Page 230
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-3-3 RFS - Radio Frequency interface and System controller Figure 5-9 RFS-Board - Block diagram 5-3-3-1 RFS board - FPGA A.) Beamformer interface...
Page 231
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL • Stores this data in the Line Memory buffer (2x SDRAM) • Transfers buffered data by DMA-transfer via PCI and PCI-Express to the PC Non-TimeMotion data (B, C): DMA0 TimeMotion data (M, PW): DMA1.
Page 232
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-4 BackEnd Processor Figure 5-10 BackEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory)
Page 233
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 5-5 Voluson® P8 / Voluson® P6 - RFS components Sub-section Description Page Number 5-4-1 5-26...
Page 234
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Control DVD drive (SATA), User Interface (USB), DVR (USB) Figure 5-11 SOM (System On Module) 5-4-3...
Page 235
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-5 Internal I/O Figure 5-12 Voluson® P8 / Voluson® P6: Internal I/O 5-28 Section 5-5 - Internal I/O...
Page 236
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-6 OPIO (User Interface) 5-6-1 OPIO Block Diagram A/N Key USB Cable 8 wire (USB (4) + LED backlight)
Page 237
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-6-2 OPIO layout Figure 5-14 Voluson® P8 / Voluson® P6 - OPIO Layout 5-6-3 Button Description...
Page 238
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 5 - Components and Functions (Theory) 5-31...
Page 239
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-32 Section 5-6 - OPIO (User Interface)
Page 240
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Function of the Trackball at Diverse Dialog Pages In general operations at diverse dialog pages and windows on the system desktop (e.g., Patient Data Entry, Usage of EUM, System Setup, Measurement Setup, etc.) are executed by...
Page 241
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-34 Section 5-6 - OPIO (User Interface)
Page 242
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-7 Monitor right > < left On screen Display (OSD) increase decrease Menu key Figure 5-15 Monitor Adjustment buttons Menu key [OSD](Press the OSD (middle) key of the monitor controls over 10 sec.):...
Page 243
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-8 External I/O Figure 5-16 External I/O Connectors - on Rear of System (GES) Item...
Page 244
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-9 Peripherals 5-9-1 Recording Tools 5-9-1-1 DVD Recorder The DVR recorder is implemmented by installing the DVR Module on RFS incorporating with DVD Drive.
Page 245
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 5-10 Power Distribution 5-10-1 ACFE - Primary Power Module Circuit breaker Equipotential connection Mains voltage indicator...
Page 246
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-10-1-1 CPS Block Diagram Figure 5-19 CPS Block Diagram Chapter 5 - Components and Functions (Theory)
Page 247
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-10-1-2 Power Distribution Block Diagram Figure 5-20 CPS Block Diagram 5-40 Section 5-10 - Power Distribution...
Page 248
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-10-1-3 Power Up Sequence Description The Power Up Sequence can be divided in the following steps: •...
Page 249
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-10-1-3-2 Power Down Sequence Description Figure 5-22 Power Down Sequence 1.) On/Off Switch is pressed on OPIO.
Page 250
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: All DC-supply voltages for built in peripherals are generated in DCDC module not inside ACFE.
Page 251
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-10-2 ACDC - primary DC Power Supply 5-10-2-1 Function of ACDC With the AC input voltage from the ACFE, the ACDC module generates 12Vdc...
Page 252
4.) Enter the password SHE and then click the ACCEPT button. 5.) Click the CSD button. 6.) As soon as the GEHC Service Home Page appears, select “GE Service” from the pull-down menu, enter the “Standard GE Revolving” password (password changes every six months) and then click [Okay].
Page 253
1.) If not already in read mode, FREEZE the image. 2.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the “Extended Access” login window pops up..
Page 254
Section 7-5-7 "CSD: Configuration" on page 7-25. !! NOTICE: This offers GE technicians the possibility to view the entire customer´s desktop and operation system. Remote access to the Voluson® P8 / Voluson® P6 scanner requires permission and customer input to run diagnostics.
Page 255
5-12-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
Page 256
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-12-2 Error Logs When the Error Logs page is selected, different log viewing options are available.
Page 257
Figure 5-29 Common Service Desktop - Configuration Remote access is ONLY possible if the service platform is properly configured (either by the user NOTICE or a GE technician at site). Operation see: Section 7-5-7 "CSD: Configuration" on page 7-25. !! NOTICE:...
Page 258
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-12-6 Utilities The Utilities page contains a variety of Windows utility tools to indicate the status of the system, in addition to various other tools.
Page 259
2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
Page 260
5-45. 5-13-3-5 Request for Service Click the RFS button, fill out the “Request For Service” form and then send the problem description to GE Service/Application representatives. Operation see: Section 7-2-1 "Request for Service (RFS)" on page 7-4 5-13-3-6 Delete all Patients 1.) Click the DELETE ALL PATIENTS...
Page 261
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-13-3-8 Keyboard Layout To change the keyboard layout to different languages. Operation see: Section 6-4 "Modification of Keyboard Layout" on page 6-3.
Page 262
2.) As soon as the “Boot Screen” appears (see: Figure 5-36 down] key on the keyboard until the GE Service item is highlighted, then press ENTER. (case sensitive) and confirm with ENTER. The system boots into 3.) Enter the Password Rudi LINUX.
Page 263
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-14-1-3 Rollback This function offers the possibility to simply restore the previously used system configuration (rollback), which was stored on R:\ during “FMI from DVD”.
Page 264
• * via Boot Screen Functions - GE Service: functions are secured by service password (class C) * via Windows Start Menu (for GE Field Engineers ONLY): by means of service dongle and GE password System is NOT responsive: (booting from System DVD is required) •...
Page 265
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-3 Before Recovering or Initializing the Hard disk If the system is still responsive (system starts up without problems and the application software CAUTION is running as usual, try to print out or write down System Setup and Measure Setup settings;...
Page 266
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL - Click the GENERAL tab and print out or write down the settings of this page.
Page 267
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-4 Voluson Maintenance System Enter Linux as described in Section 5-15-2 "Access/Security" on page 5-57. The “Voluson Maintenance System” main menu with following functions appear:...
Page 268
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-5 Media Menu The “Media Menu” is only accessible if the System DVD (see: Table 9-13) is inserted into the system’s...
Page 269
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Operation: 1.) Select [Restore] in the displayed “Media Menu” (see: Figure 5-39 above). NOTE: If the message: “Recovery system (Linux) needs to be updated ...” appears, confirm with YES.
Page 270
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4.) In the displayed menus select [Exit] and then [Exit/Reboot]. Figure 5-42 Exit -> Exit/Reboot 5.) Remove the System DVD from the DVD/CD+R/RW Drive drive.
Page 271
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8.) If following message appears, skip the “BT warning” with CONTINUE. Figure 5-44 Skip the BT warning with CONTINUE NOTE: Options will be recovered at a later step using “Full Backup”...
Page 272
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTICE However, if system functionality allows it: • save the patient images and corresponding patient data...
Page 273
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Use “Arrow” keys [left] or [right] on the alphanumeric keyboard to select the [Boot] tab.
Page 274
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7.) Select Primary Master and then press ENTER. again. Figure 5-46 1st Boot Device Chapter 5 - Components and Functions (Theory)
Page 275
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8.) Select 3rd Boot Device and then press ENTER. 9.) Select Third Master and then press ENTER. again.
Page 276
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) In the displayed menus select [Media] -> [NewDisk] and then press ENTER. Figure 5-48 Media -> NewDisk...
Page 277
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL below) appears and then press ENTER. Please wait until the final window (Figure 5-50 Figure 5-50 New Harddisk setup OK 6.) Remove the System DVD from the DVD/CD+R/RW Drive.
Page 278
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10.)Wait until all processes are finished and the final “DOS” window (Figure 5-52 below) appear. Figure 5-52 press the “A” key to start UlSApp 11.)Press A key to activate the Ultrasound Application Software [UIS].
Page 279
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-6 Advanced Functions Menu 1.) If system functionality allows it, collect vital system information as described in Section 5-15-7 "After...
Page 280
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-6-3 SSHD By selecting this item, a SSH-daemon is started on the system. The [SSHD] function is reserved for extended troubleshooting. Not intended for use in the field.
Page 281
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-6-9 USBTree The USB Tree displays connected devices and their respective hub. This enables easy control, whether all devices are correctly wired.
Page 282
1.) If not currently done, restart the system (with new HDD) and boot up completely. 2.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
Page 283
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 12.)Press F2, select the MEASURE SETUP item and check and match settings of all the pages.
Page 284
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-15-8 Memory Check in LINUX 1.) Turn system OFF and then back ON. 5-55), press the PG DN As soon as the “Boot Screen”...
Page 285
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5-78 Section 5-15 - LINUX - Voluson Maintenance System...
Page 286
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
Page 287
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 6-3 LCD Monitor Adjustment The Voluson® P8 / Voluson® P6 has a free adjustable LCD monitor in relation to the user interface.
Page 288
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 6-4 Modification of Keyboard Layout NOTE: Configuring the layout of the keyboard is only useful by changing the concerned keys also.
Page 289
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 6.) Select default input language from the drop down menu. default input languages (drop down menu)
Page 290
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter VOLUSON® P8 / VOLUSON® P6 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
Page 291
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-2 Collect Vital System Information 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 292
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-2 Collect Vital System Information (cont’d) System type Serial number Hardware ID Application Software Backup filename...
Page 293
!! NOTICE: There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen. Move the cursor to the GE icon and press the left trackball key (= left-click) to display the “Request For Service”...
Page 294
5.) Write down and keep the Reference Number for follow up procedures, then click OK. If the service platform is not configured an Error message is displayed. The request is NOT sent! NOTE: The request is saved in QUEUE for reviewing and/or resending. Figure 7-4 Contact GE - Queue Chapter 7 - Diagnostics/Troubleshooting...
Page 295
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-2-2 Shortcuts List Press the CTRL + H key simultaneous to display the shortcuts list and a description of what they do.
Page 296
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 How to check power 7-3-1-1 Common Power Supply (CPS) Turn on mains switch and check green LED inside Power.
Page 297
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center.
Page 298
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-4-1-1 How to use a Capture Filter Use filters to limit the captured data. Filter by IP and port number by IP only.
Page 299
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-4-1-2 How to use a Screen Filter In addition to the “pre” capture filter, use the Filter tool in the screen to filter what is displayed.
Page 300
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-4-2 Export Log’s and System Data There are two possibilities to export system data (and log´s): 1.) by pressing the ALT + D key to save a snapshot and “Full Backup”...
Page 301
2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
Page 302
7-5-1 General This offers GE technicians the possibility to view the entire customer´s desktop and operation system. Using VCO (Virtual Console Observation) a service technician or the OnLine Center can access and modify all PC settings and programs or run diagnostics on the customer´s ultrasound scanner.
Page 303
SSO and password. NOTE: ONLY registered GE Service personnel have access to this feature. A password is required. All actions and changes should be documented. 2.) Search for the Voluson® P8 / Voluson® P6 system which should be observed. (Refer to...
Page 304
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 7-15 enable Disruptive Mode 6.) To Enable “Disruptive Mode” select YES. NOTE: A request is sent to the scanner or workstation asking permission to diagnose the system remotely.
Page 305
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL The screen below confirms the Disruptive Mode status change. Figure 7-17 Disruptive Mode enabled 7.) After “Disruptive Mode” is enabled, “Virtual Console Observation” can be activated by selecting Utilities ->...
Page 306
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8.) To terminate “Virtual Console Observation” click on the STOP button. Figure 7-19 Virtual Console Observation is stopped NOTICE After performing any diagnostics, disable “Disruptive Mode”...
Page 307
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-5-4 How the FE remotely Connect to Device (Total Access Session) 1.) On the remote workstation start the Internet Explorer and navigate to the enterprise server: https://plt2-insite2.health.ge.com/qss/gelogin.jsp...
Page 308
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL A screen with additional system data and information appears. Connect to Device Figure 7-21 Connect to Device 5.) In the right frame (see:...
Page 309
Figure 7-22 “TotalAccess Session” window The Total Access Session window shows the different connection types supported by the device. Virtual Network Computing (VNC) enables the GE Service technician to see and inter- UL_VNC act with information displayed on the monitor of the Voluson® P8 / Voluson® P6 ultra- sound system remotely.
Page 310
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-5-5 Virtual Network Computing (VNC) 1.) In the Total Access Session window (see: Figure 7-22 on page 7-20) find the item labeled UL_VNC and click on the [Connect] button.
Page 311
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL The tool bar on top of the virtual monitor screen contains useful options. Figure 7-25 Tool bar on top of virtual screen...
Page 312
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: For complete shortcut list refer to Section 7-2-2 "Shortcuts List" on page 7-6. Press the shortcut ALT + S on the alphanumeric keyboard, to display the Console Emulator.
Page 313
5-48, the service platform uses a web-based user interface (UI) to provide access to common service components. The Service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface.
Page 314
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-5-7 CSD: Configuration 7-5-7-1 To configure Service Platform If a Proxy server is available, the information MUST be properly entered, otherwise remote control NOTICE does not work.
Page 315
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: 1.) It requires all bold fields should be filled out. The mandatory are : 1.) Device Name (Not Change) 2.) CRM No...
Page 316
2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up.
Page 317
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-7 How to use the Auto Tester program (cont’d) The following message box appears. Figure 7-31 Message Box 8.) Click OK.
Page 318
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-7 How to use the Auto Tester program (cont’d) The following screen will appear. “DVD/CD Burn”...
Page 319
Before starting this tool, all programs that access the databases MUST be closed! CAUTION 1.) Connect the USB “standard GE” service dongle on the front USB connector(s). As soon as the dongle is detected, the login window pops up. Remove the dongle.
Page 320
(cont’d) 8.) After terminating all processes mentioned above, press CTRL + ESC to open windows start menu. 9.) Using the [Arrow] keys on the keyboard, select “Programs” -> “GE FIELD ENGINEER” and then double-click the “Compact and Repair Database” application.
Page 321
2.) In the F2 menu, press SYSTEM SETUP button to invoke the setup desktop on the sceen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the system. As soon as the dongle is detected, the login window pops up.
Page 322
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Automatic/Not-Automatic diagnostic tests are executed from this System Diagnostics Tool. Figure 7-41 System Diagnostics Tool Table 7-1 below outlines Voluson®...
Page 323
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-1 Audio(System) Interface Test. Test of the Audio path from system to Speaker and AudioOutput can be performed.
Page 324
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-3 PatternGen - Pattern Generator Test. This test can help to find out where image problems are caused from (FrontEnd of BackEnd processing).
Page 325
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-4 Apertur Lateral Min This test is used for detection of missing channels on Probe Connector Board (PIR) or missing elements on probe, which are typically shown as dark Ultrasound Lines.
Page 326
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-5 Temperature Test This test is used to check the operation of Temperature Warning Message by simulating warning temperature level.
Page 327
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL After 15 minutes, a second warning message is displayed. (see : Figure 7-46 below). Figure 7-46 A second high system temperature message 5.) Click OK to exit the warning message.
Page 328
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6 Automatic Diagnostic Test. Automatic diagnostic test can be performed by selecting the desired test item from "Automatic Test Selection Box"...
Page 329
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-1 System Overall Test System Overall Test is root-top level test that executes all the sub diagnostic tests. It executes each sub diagnostic test one by one, and summarizes the test result at the end of the test execution.
Page 330
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-2-5 FPGA(RFS) Configuration Test It verifies FPGA(RFS) configuration on RFS board by reading/writing test data, and displays the test result as PASSED/FAILED.
Page 331
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-3 BF (Beamformer) Overall Test This diagnostic test performs the self-diagnosis of every electric component and interface in BF board.
Page 332
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-4 PIR Overall Test This diagnostic test performs the self-diagnosis of each electric component and interface in PIR board.
Page 333
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-5-2 IIC(CPS) Interface Test It verifies I2C path named I2C_PS by using EERPOM on CPS board, and displays the test result as PASSED/FAILED.
Page 334
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-9-6-7-2 Tx Signal Path Test It verifies Tx Signal from TxPulser on BF board by firing test Tx pulse wave and reading back from corresponding Rx channel, and displays each channels diagnostic result, and the overall test result as PASSED/FAILED.
Page 335
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-10 Minimum Configuration to Boot/Scan 7-10-1 Booting without using the Ultrasound Hardware If the low-level ultrasound hardware is failing, the application can not be initialized correctly.
Page 336
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 7-49 H Key pressed during startup 4.) Click on CONTINUE. The "Probe Selection" menu appears on the screen.
Page 337
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 7-50 Reboot Machine 7-48 Section 7-10 - Minimum Configuration to Boot/Scan...
Page 338
7-11-1 Console Test 1.) Connect the USB "standard GE" service dongle on the front USB connector(s). As soon as the dongle is detected the login window pops up. 2.) Enter the "Standard GE Revolving" password (password changes every six months) and then click ENABLE.
Page 339
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9.) The "TetrisConsole - Console Tester" window is displayed. Status Window Figure 7-52 Console Tester NOTE: For description of keys and their functions as well as the key code numbers refer to: Figure 5-6-2: OPIO layout on page 5-30The "TetrisCosole - Console Tester"...
Page 340
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-12 Troubleshooting Trees, Instructions and Tech Tips Table 7-10 below outlines Voluson® P8 / Voluson® P6 troubleshooting trees and instructions shown in the sub-sections.
Page 341
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-1 System does not boot up System doesn’t boot Verify that Is Standby Switch - Power Cord is connected to Mains...
Page 342
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-2 Noise disturbs the Image Noise is shown in image. Is there any electrical device that affect the...
Page 343
ROPRIETARY ERVICE ANUAL 7-12-3 Trackball - Impaired Sensitivity The trackball has low sensitivity! Contact GE service representative Is it difficult to roll the to replace the trackball module. trackball? Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page).
Page 344
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-4 Printer Malfunction Figure 7-56 Printer Malfunction Chapter 7 - Diagnostics/Troubleshooting 7-55...
Page 345
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-5 Monitor Troubleshooting Fault symptom Check these items Check the main panel cable is properly connected to Monitor.
Page 346
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-6 DVD/CD+R/RW Drive Test NOTE: Only DVD+R/RW and CD+(R)W medias are available. 1.) Insert an empty DVD/CD+R/RW into the Drive.
Page 347
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-6 DVD/CD+R/RW Drive Test (cont’d) 6.) Export images of the selected exam to DVD/CD+R/RW: a.) Click the EXPORT button.
Page 348
Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK. 7-12-7-2 GE remote service connection The customer gets visual information about GE remote service status (shown in the status area on the right bottom of the screen). Gray = Idle State Remote access is inactive.
Page 349
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7-12-8 Tech Tips Table 7-11 below to access Voluson® P8 / Voluson® P6 (troubleshooting) Technical Tips listed.
Page 350
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 7-13 Error Messages NOTICE If the problem (error message) still exists after performing the described actions, call technical support.
Page 351
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions BC Ensemble Out Of Range press ok and save this user-setting once again BC lines_per_sequenz <...
Page 352
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions cannot read file restart the system Cannot import volume data to 3D dll Load volume files from other storage medium.
Page 353
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions restart the system CW quant TxFreq out of range Press ok and save this user-setting once again.
Page 354
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions Enhance Out Of Range press ok and save this user-setting once again Error in apertureShape, lateral start pos of the aperture incorrect!
Page 355
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions File Volume size not consistent. load volume files from other storage medium Action: Load volume files from other storage medium.
Page 356
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions iPWCW_CloseProcessing in RT_PWorCWFilterBlock::execute() returned error restart the system code xx !! Action: restart the system, If after restart not ok then call technical support...
Page 357
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions check the connection and the Power cable - plug the cable No CD Writer found off and on and try again.
Page 358
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet on left probe connector (A). Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet on middle probe connector (B).
Page 359
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions RepresentationManager::vCreate: Dimension unknown, arguments of restart the system vSetReplayParams() incorrect!! RepresentationManager::vCreate: Nr of Units = xx, wrong arguments in Constructor...
Page 360
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions RT_MColorFilterBlock::Memory allocation failed. restart the system Action: restart the system, If after restart not ok then call technical support...
Page 361
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions RT_RFProcessingBlockTissue::vCheckRFDataSizeAndUpdateBuffers: Memory allocation failed. restart the system Action: restart the system, If after restart not ok then call technical support RT3D.cpp rot phased: unexpected O_AxisAngle ?
Page 362
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions if possible, • save “Full Backup” (see: Section 4-5-3 on page 4-39) to The System detected a problem with your harddisk.
Page 363
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Error Messages Actions Unhandled Probe-EEPROM data type on middle probe connector (B). Connect another probe Unhandled Probe-EEPROM data type on right probe connector (C).
Page 364
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems.
Page 365
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-15 8-37 Replacement of the AIR FILTER...
Page 366
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-1-2 Returning/Shipping System, Probes and Repair Parts When returning or shipping the Voluson® P8 / Voluson® P6 system in the original packaging: •...
Page 367
One person ~ 1 hour (depends on contents of System DVD, peripherals, etc.) 8-2-2 Tools System DVD - Base image software part number for Voluson P6/P8 - 5438266 1.) Insert the CD/DVD 2.) The CD/DVD is located as shown below. Ensure to remove the Front cover and DVD Bezel assembly before performing the steps illustrated below.
Page 368
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Select Boot option Figure 8-2 Boot option 4.) Select Media Select Media Figure 8-3 Select Media...
Page 369
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Select New Disk Select New Disk Figure 8-4 New Disk 6.) Figure below shows the Installing screen Figure 8-5 Installing 7.) After installation, remove the DVD from the system.
Page 370
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Select A Figure 8-6 Activate Software 10.)After UISApp is running, check the installed version by following these steps: A.) Press F2 (Utility) hard key on the console...
Page 371
DVD being used is SW 13.0.0 or higher, calculate new software specific “Permanent key” in OKOS. Please contact your local distributor or GE service representative to get the necessary key. NOTE: It is NOT necessary to save Full System Configuration (Full Backup) prior to the upgrade.
Page 372
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-4 Preparations (cont’d) Table 8-1 System Function Mode Task Expected Result Connect an abdominal probe, press the PROBE key, select the “Abdomen”...
Page 373
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL which Software/Hardware version is installed in the unit. Voluson P8 currently installed Ultrasound Application Software version...
Page 374
If the currently installed software has to be upgraded to a newer version, and the system DVD being !! NOTICE: used is SW 13.0.0 or higher, a new software specific “Permanent key is required. Please contact your local distributor or GE service representative. Chapter 8 - Replacement Procedures 8-11...
Page 375
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-5 System Software - Installation Procedure (FMI from DVD) The system software installation procedure starts with saving and recording the settings present on the system (silent “Rollback”).
Page 376
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-5 System Software - Installation Procedure (FMI from DVD) (cont’d) 9.) Click the FMI FROM DVD button (see:...
Page 377
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-5 System Software - Installation Procedure (FMI from DVD) (cont’d) 12.)Check disk is performed automatically - restart.
Page 378
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-5 System Software - Installation Procedure (FMI from DVD) (cont’d) 14.)Booting auto..Figure 8-17 Boot screen - auto 15.)Updating will take some time..
Page 379
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-3-5 System Software - Installation Procedure (FMI from DVD) (cont’d) Figure 8-19 automatic processes are running 18.)Please wait until all processes are finished.
Page 380
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-4 Software and Functional Checks after Installation/Upgrade Procedure 1.) Press the F2 key on the control panel.
Page 381
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-5 Image Settings Only - Loading Procedure 8-5-1 Introduction The Image Settings contain: • Application Settings •...
Page 382
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-8 Replacement or Activation of Options Following Software options are available: Table 8-2 Software options...
Page 383
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-8-1 Operation for activating Options 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
Page 384
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-9 Replacement of Monitor 8-9-1 Manpower 2 people, 10 minutes 8-9-2 Tools Philips screwdriver, Allen Wrench...
Page 385
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL The monitor arm can spring up and cause serious damage or personal injury while separating the...
Page 386
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-10 Replacement of the MONITOR STAND ARM 8-10-1 Manpower Two persons, 20 minutes 8-10-2 Tools...
Page 387
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-11 Replacement of OPIO 8-11-1 Manpower One person, 15 minutes 8-11-2 Tools Philips screwdriver, Nipper, Allen Wrench...
Page 388
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Unscrew 9 screws located in the bottom of the OPIO and lift up the MAIN OPIO. Refer to the figure below.
Page 389
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Remove the ribbon cable. Ribbon cable 8-11-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
Page 390
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-12 Replacement of the SPEAKER 8-12-1 Manpower One person, 30 minutes 8-12-2 Tools Philips screwdriver...
Page 391
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Disconnect the 2 SPEAKER CONNECTORS. Refer to the figure below. Figure 8-30 Disconnecting SPEAKER CONNECTOR...
Page 392
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-13 Remove the Pedestal Bottom Cover PEDESTAL BOTTOM COVER PEDESTAL FRONT COVER Figure 8-31 Unscrewing 1 screw 3.) Unscrew 2 screws located in the Pedestal Front Cover and remove the part.
Page 393
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Remove the REAR COVER ASSY. Refer to the figure below.. Figure 8-33 Unscrewing 5 screws and Remove REAR COVER ASSY 6.) Unscrew 7 screws to remove the CABLE COVER BRACKET.
Page 394
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7.) Untie the CABLE CLAMP with finger. Refer to the figure below. Figure 8-35 Untying the CABLE CLAMP...
Page 395
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8.) Unscrew 2 screws to remove the MONITOR CABLE BRACKET. Then unscrew the screw of MAIN PANEL LONG CABLE and disconnect the CABLE.
Page 396
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-14 Replacement of the NEST 8-14-1 Manpower One person, 10 minutes 8-14-2 Tools Philips screwdriver, Allen Wrench...
Page 397
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 8-3 Cable Connections with FRU information Cables LCD Monitor cable 5444914 OPIO cable 6020400 Peripheral Cable Assembly...
Page 398
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Set the 3 RS KNOBS to unlock position to separate from the SIDE COVER R. and unscrew one screw from the each 3 RS KNOB.
Page 399
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Unscrew 2 hexagonal socket screws. Refer to the figure below. Figure 8-42 Unscrewing 2 hexagonal socket screws 6.) Unscrew 7 screws to open the rear bracket.
Page 400
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-15 Replacement of the AIR FILTER 8-15-1 Manpower One person, 1 minutes 8-15-2 Tools None...
Page 401
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-15-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-14: Replacement of the AIR FILTER - Functional Tests.
Page 402
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-16 Replacement of the Probe Holder(KIT) 8-16-1 Manpower One person, 1 minutes 8-16-2 Tools None...
Page 403
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-17 Replacement of SIDE COVER L ASSY 8-17-1 Manpower One person, 5 minutes 8-17-2 Tools...
Page 404
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-17-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-16: Replacement of SIDE COVER L ASSY - Functional Tests.
Page 405
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-18 Replacement of SIDE COVER R ASSY 8-18-1 Manpower One person, 5 minutes 8-18-2 Tools...
Page 406
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Unscrew one screw from the each 3 RS knobs. Refer to the figure below. Figure 8-51 Removing the 3 RS KNOBS...
Page 407
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-19 Replacement of the Trackball 8-19-1 Manpower One person, 10 minutes 8-19-2 Tools Philips Screwdriver...
Page 408
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Unscrew 9 screws located in the bottom of the OPIO and lift up the MAIN OPIO. Refer to the figure below.
Page 409
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4.) Disconnect the TRACKBALL CABLE from the OPIO HUB PWA. Refer to the figure below. Figure 8-55 Disconnecting the TRACKBALL CABLE 5.) Unscrew 2 screws and then pull out the TRACKBALL from the MAIN OPIO.
Page 410
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-20 Replacement of the TGC BOARD 8-20-1 Manpower One person, 10 minutes 8-20-2 Tools Philips Screwdriver...
Page 411
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Disconnect TGC BOARD CABLE. Refer to the figure below. Figure 8-58 Disconnecting TGC BOARD CABLE 4.) Unscrew 4 screws.
Page 412
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-21 Replacement of AN KEYBOARD ASSY 8-21-1 Manpower One person, 15 minutes 8-21-2 Tools Philips Screwdriver...
Page 413
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-21-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-20: Replacement of AN KEYBOARD ASSY - Functional Tests.
Page 414
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-22 Replacement of the Caster Cap 8-22-1 Manpower One person, 2 minutes 8-22-2 Tools None...
Page 415
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-23 Replacement of the CASTER 8-23-1 Manpower One person, 10 minutes 8-23-2 Tools Monkey Wrench...
Page 416
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 8-64 Unfasten the nut 8-23-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
Page 417
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-24 Replacement of the SIDE TRAY 8-24-1 Manpower One person, 2 minutes 8-24-2 Tools Philips screwdriver...
Page 418
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-25 Replacement of the REAR HANDLE 8-25-1 Manpower One person, 5 minutes 8-25-2 Tools Philips screwdriver...
Page 419
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4.) Unscrew 2 hexagonal Socket screws to remove the REAR HANDLE. Refer to the figure below.
Page 420
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-26 Replacement of the OP COVER L,R 8-26-1 Manpower One person, 5 minutes 8-26-2 Tools...
Page 421
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-26-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-27: Replacement of the OP COVER L,R - Functional Tests.
Page 422
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-27 Replacement of the CPS 8-27-1 Manpower One person, 30 minutes 8-27-2 Tools Philips screwdriver, Allen Wrench...
Page 423
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-27-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-28: Replacement of the CPS - Functional Tests.
Page 424
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-28 Replacement of the RFS 8-28-1 Manpower One person, 30 minutes 8-28-2 Tools Philips screwdriver, Allen Wrench...
Page 425
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-28-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perfom 8-28-6 Serial Number - Reprogamming Procedure to reprogram the serial number.
Page 426
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 8-75 Hardware ID on label RFS-board 2.) Call the OLC to get the "Serial Key" and the "Permanent Key"...
Page 427
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Program the System Serial number and re-enter the Permanent Key To program the System SNO, click the OPTIONS tab: Field to enter the “Serial Key”...
Page 428
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-29 Replacement of the SOM 8-29-1 Manpower One person, 30 minutes 8-29-2 Tools Philips screwdriver, Allen Wrench...
Page 429
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL When replacing with new SOM, Push the SOM in both side of RFS ASSEMBLY without putting CAUTION weight on the board.
Page 430
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-30 Replacement of the HDD 8-30-1 Manpower One person, 30 minutes 8-30-2 Tools Philips screwdriver, Allen Wrench...
Page 431
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-30-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Set up the new hard disk as described in 5-15-5-4-2 on page 5-68.
Page 432
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-31 Replacement of the DVR 8-31-1 Manpower One person, 30 minutes 8-31-2 Tools Philips screwdriver, Allen Wrench...
Page 433
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Remove the dummy board from the RFS. Figure 8-82 Installing DVR 8-31-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
Page 434
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-32 Replacement of the BF64 Manpower One person, 30 minutes 8-32-1 Tools Philips screwdriver, Allen Wrench...
Page 435
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Unscrew 6 screws. 3.) Remove BF64 from the EMI Shield cover.. EMI Shield cover Figure 8-85 Unscrewing 6 screws...
Page 436
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-33 Replacement of the PIR 8-33-1 Manpower One person, 30 minutes 8-33-2 Tools Philips screwdriver, Allen Wrench, Hirose ejector...
Page 437
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Unscrew 6 screws. 3.) Remove PIR from the EMI Shield cover. Figure 8-88 Removing the bracket...
Page 438
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-34 Replacement of the CSI 8-34-1 Manpower One person, 30 minutes 8-34-2 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
Page 439
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-34-4 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-35: Replacement of the CSI - Functional Tests..
Page 440
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-35 Replacement of the CBP 8-35-1 Manpower One person, 30 minutes 8-35-2 Tools Philips screwdriver, Allen Wrench...
Page 441
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Unscrew 6 screws. Figure 8-93 Unscrewing 6 screws 8-35-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
Page 442
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-36 Replacement of the DVD 8-36-1 Manpower One person, 10 minutes 8-36-2 Tools Philips screwdriver...
Page 443
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Unscrewing 2 screws to remove the DVD . Figure 8-96 Removing the DVD 8-36-5 Installation Procedure 1.) Install the new parts in the reverse order of removal.
Page 444
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-37 Replacement of the OPIO USB Sub PWA 8-37-1 Manpower One person, 30 minutes 8-37-2...
Page 445
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-37-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-39: Replacement of the OPIO USB Sub PWA - Functional Tests.
Page 446
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-38 Replacement of AN KEY TOP 8-38-1 Manpower One person, 30 minutes 8-38-2 Tools Flat head screw driver...
Page 447
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-39 Replacement of FULL AN KEY 8-39-1 Manpower One person, 30 minutes 8-39-2 Tools Philips screwdriver...
Page 448
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Lift up the OPIO Assy. And then, disconnect 3 cables from OPIO ASSY. Figure 8-101 Disconnecting 3 cables from OPIO ASSY 3.) Separate the cable that connects with HUB Board.
Page 449
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-39-4 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-41 Replacement of FULL AN KEY - Functional Tests.
Page 450
• if just the circle key caps have to be replaced. • if trackball buttons or mode key slices have to be replaced too, please contact your local distributor or GE service representative. 8-40-1 Replacement of Circle Key Caps only...
Page 451
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-41 Replacing optional Peripherals NOTICE Normally auxiliary devices and peripherals come pre-installed with the Voluson® P8 / Voluson® P6 system.
Page 452
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL EMI Gaskets (5448934) EMI Gaskets (5447303) EMI Gaskets (5457475) Figure 8-105 B/W Printer assembly Chapter 8 - Replacement Procedures...
Page 453
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Disassemble front printer cover assembly 6.) Remove dummy printer bracket assembly Figure 8-106 Disassemble printer cover 7.) Disassemble power cord bracket...
Page 454
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9.) Unscrew 7 fasteners and disassemble rear EMI cover. 10.)Unscrew 2 fasteners and disassemble printer EMI cover 1 and 2.
Page 455
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 12.)Assemble Printer assembly. 13.)Push the printer assembly till it makes contact with sheet metal chassis. Push the printer assembly...
Page 456
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 17.)Insert the printer USB cable and printer power cable (5410088 - printer cable set). 18.)Insert the printer power cable in to the nest box.
Page 457
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 20.)Check the printer USB cable and power cable to passing through in EMI cover slot. Then fasten 2 screws and assemble printer EMI cover 1 and 2.
Page 458
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-41-3-2 Color Printer Assembly 1.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate.
Page 459
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 5.) Remove the front printer cover assembly. Figure 8-117 Front cover 6.) Disassemble power cord bracket...
Page 460
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7.) Unscrew 7 fasteners and disassemble rear EMI cover. 8.) Unscrew 2 fasteners and disassemble printer EMI cover 1 and 2.
Page 461
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 13.)Assemble the printer assembly. 14.)Assemble the printer base plate with chassis using 2 screws (2159634 - screw BH M4X10).
Page 462
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Printer Power cable in Nest box Printer USB cable Printer Power cable Figure 8-123 Printer cables 17.)Insert the printer USB cable in to the nest box.
Page 463
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 19.)Fasten the 7 screws and fix the rear EMI cover. 20.)Assemble the rear cover, power cord and power cord bracket.
Page 464
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 22.)Assemble 4 numbers of Holding Catch, female with DVD front cover using 8 screws (2306562 - screw FH M3x6) 23.)Assemble 2 numbers of holding catch male with sheet metal chassis using 4 screws (2306562...
Page 465
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-41-4 HORIZONTAL TV PROBE HOLDER Installation 1.) Install HORIZONTAL TV PROBE HOLDER on the left side bottom of OPIO with 3 screws.
Page 466
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-41-5 VERTICAL TV PROBE HOLDER Installation 1.) Install VERTICAL TV PROBE HOLDER BASE on the right side bottom of OPIO with 2 screws.
Page 467
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-42 Replacement of FAN with Cable ASSY II 8-42-1 Manpower One person, 30 minutes. 8-42-2...
Page 468
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-43 Functional Checks after FRU replacement (Debrief Guidelines) 1.) Perform required Functional and/or Leakage Current tests based upon the FRU being replaced.
Page 469
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-44 Replacement of CWD 8-44-1 Manpower One person, 30 minutes. 8-44-2 Tools Philips screwdriver, Hirose ejector...
Page 470
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 2.) Unscrew 5 screws and remove REAR COVER. Figure 8-135 Rear cover 3.) Remove the REAR EMI COVER.
Page 471
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 4.) Unscrew Monitor Cable fix screw with "-" Driver and Disconnect all cable connectors. Peripherals Power Cable...
Page 472
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 6.) Pull out BF64/PIR Boards. Figure 8-139 BF64/PIR Boards 7.) Unscrew 6 screws and replace BF64(p/n:5396937) to BF64C Boards which have CWD Circuits.
Page 473
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 8-45 Replacement of ECG Kit assy 8-45-1 Manpower One person, 10 minutes. 8-45-2 Tools Philips screwdriver...
Page 474
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Open keyboard assembly. 4.) Remove ECG clamp. Figure 8-142 ECG Clamp 8-45-5 Installation Procedure 1.) Assemble ECG kit on keyboard base.
Page 475
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 3.) Connect the USB cable with Keyboard assembly as shown. 4.) Assemble Keyboard assembly. Figure 8-144 USB cable 5.) Fasten 9 numbers of fasteners below the Keyboard base.
Page 476
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 7.) Attach ECG label. ECG label Figure 8-146 ECG label 8.) Perform: FRU8-47: Replacement of ECG Kit assy - Functional Tests...
Page 477
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-5: Full Backup (Full System Configuration) - Loading Procedure go to 4-5-4 Load Full System Configuration (Full Backup)
Page 478
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-10: Replacement of OPIO Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson®...
Page 479
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-14: Replacement of the AIR FILTER Functional and/or Leakage Current Test - N/A FRU8-15: Replacement of the Probe Holder(KIT)
Page 480
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-20: Replacement of AN KEYBOARD ASSY Functional and/or See Section Leakage Current Test Debrief Script 4-3-1...
Page 481
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-26: Replacement of the REAR HANDLE Functional and/or See Section Leakage Current Test Debrief Script 4-3-1...
Page 482
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-30: Replacement of the SOM Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson®...
Page 483
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-34: Replacement of the PIR Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson®...
Page 484
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-37: Replacement of the DVD Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson®...
Page 485
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL FRU8-41 Replacement of FULL AN KEY Functional and/or See Section Leakage Current Test Debrief Script 4-3-1...
Page 486
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Functional and/or See Section Leakage Current Test Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, Voluson®...
Page 487
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 8-124 Section 8-45 - Replacement of ECG Kit assy...
Page 488
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the Voluson® P8 / Voluson® P6.
Page 489
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter A/N Key Alpha Numeric Keys...
Page 490
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-3 Parts List Groups Table 9-2 Mechanical and user accessible parts Item Part Group Name...
Page 491
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-4 Plastics Covers (Front/Sides/Rear) Figure 9-1 Plastic Covers Table 9-3 Plastics Covers Item Part Name...
Page 492
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-5 LCD Monitor Figure 9-2 LCD Monitor Table 9-4 LCD Monitor Item Part Name Part Number...
Page 493
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-6 OPIO Section 9-6 - OPIO...
Page 494
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Figure 9-3 OPIO Table 9-5 OPIO Item Part Name Part Number Description 5461194 OPIO USB Sub PWA Assy...
Page 495
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-7 Nest Box Parts Figure 9-4 Nest Box parts Section 9-7 - Nest Box Parts...
Page 496
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 9-6 Nest Box Item Part Name Part Number Description 5427038 PIR Assy PIR Assy 5396937-2...
Page 497
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-8 Mechanical Parts Figure 9-5 Mechanical Parts Table 9-7 Mechanical Parts Item Part Name Part Number...
Page 498
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 9-7 Mechanical Parts Item Part Name Part Number Description 5400392-2 Castor Dual Lock 2 POLARIS...
Page 499
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-9 Options Figure 9-6 Options Table 9-8 Options Item Part Name Part Number Description 5408092...
Page 500
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-10 Optional Peripherals and Accessories 9-10-1 Printers and options Figure 9-7 Optional Peripherals and Accessories - Printers...
Page 501
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-11 System Manuals Table 9-10 System Manuals - Voluson® P8 / Voluson® P6 Part #...
Page 502
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-12 Probes Table 9-11 below outlines the replacement parts described in the sub-sections. Table 9-11...
Page 503
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9-12-2 2D-Probes - Linear Array Transducer Figure 9-9 2D Linear Array Transducer Table 9-13 2D linear array Transducers...
Page 504
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9-12-3 Real-Time 4D Volume Probes Figure 9-10 Real-Time 4D Volume Probes Table 9-14 Real-Time 4D Volume Probes...
Page 505
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9-12-4 2D-Probes - Phased Array Transducers Figure 9-11 2D Phased Array Probe Table 9-15 2D-Probes - Phased Array Transducers...
Page 506
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-13 Power Cord Figure 9-12 Power Cords Chapter 9 - Renewal Parts 9-19...
Page 507
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 9-16 Power Cord Item Part Name Part Number Description Power Cord for India 5182611 Power Cord for India...
Page 508
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-14 Biopsy Needle Guide 1001 1002 1003 1004 Figure 9-13 Biopsy Needle Guides Table 9-17...
Page 509
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 9-14 Biopsy Needle Guide (cont’d) 1005 1006 1007 1008 1009 1010 Figure 9-14 Biopsy Needle Guides...
Page 510
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 9-18 Biopsy Needle Guides Item Part Name Part Number Description reusable Biopsy needle guide for probe RAB4-8-RS, RAB2-5-RS...
Page 511
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 9-24 Section 9-14 - Biopsy Needle Guide...
Page 512
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Voluson® P8 / Voluson® P6 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Page 513
Your GE Service Representative has an in-depth knowlegde of your Voluson® P8 / Voluson® P6 ultrasound scanning system and can best provide competent, efficient service.
Page 514
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule...
Page 515
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Endocavity Probe Also after corrective •...
Page 516
DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 10-4 Tools Required Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID Description Tool ID Description 9-45358 Pliers Retaining Ring 9-XL9971MM Xcelite-hex Blade 1.27mm 9-4078 Scribe 9-XL9972MM Xcelite-hex Blade 1.5mm...
Page 517
OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-46696 Hex Keys 9-XL100X Wire Stripper/Cutter 5 inch - 100X 9-39829 Torpedo Level, Magnetic...
Page 518
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 10-5 Tools Required 10-5-1 Special Tools, Supplies and Equipment 10-5-1-1 Specific Requirements for Care & Maintenance Table 10-6 Overview of Requirements for Care &...
Page 519
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 10-6 System Maintenance 10-6-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
Page 520
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-6-2 Functional Checks The functional checks take about 60 minutes to perform. Refer to the Voluson® P8 / Voluson® P6 Basic User Manual whenever necessary.
Page 521
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-6-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted.
Page 522
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-6-5 Physical Inspection Table 10-12 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition.
Page 523
The Basic User Manual and/or care card provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of Voluson®...
Page 524
To guarantee that the sensitive probe head does not hit the bottom of Endocavity soaking cup kit RIC-Holder is available at GE Accessory the container, place the probe in the RIC Holder during the disinfection (Order Nr. : E8385TA ) procedure.
Page 525
In case of unexplainable abrupt changes of safety-test results consult experienced !! CAUTION: authorized service personnel or GE for further analysis. To minimize the risk of a probe causing electrical shock, the customer should observe the following recommendations: •...
Page 526
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-2 GEHC Leakage Current Limits for Voluson® P8 / Voluson® P6 The following limits are summarized for IEC60601-1 standard including national deviations and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
Page 527
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Country Conditions Others Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA*** Open Protective earth and close neutral with normal polarity 0.05 mA*...
Page 528
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 10-18 Type CF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.01 mA*...
Page 529
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL NOTE: * Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV ** Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table 19.5DV.1 *** Measurement limits per IEC 60601-1 Ed 2.
Page 530
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
Page 531
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test...
Page 532
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-5 Enclosure Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break.
Page 533
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL ii)Others Hot : Brown Neutral : Blue Ground : Green-yellow Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
Page 534
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-5-3 Data Sheet for Enclosure Leakage Current The test passes when all readings measure less than the value shown in Table 10-14.
Page 535
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground.
Page 536
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 10-24 Typical Data Sheet for Earth Leakage Current Measured Earth Leakage Tester Neutral Tester Polarity Switch...
Page 537
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Neutral : White Ground : Green or Green-yellow ii) Others Hot : Brown Neutral : Blue...
Page 538
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-8-2 Generic Procedure on Leakage Current Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with neutral open and closed.
Page 539
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Table 10-27 Typical Data Sheet for Type BF Patient Auxiliary Current Test Measured Leakage Tester Neutral...
Page 540
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Measured Leakage Tester Neutral Tester Polarity Switch Ground Switch Current Closed Reversed Closed 10-8-12 Total Patient leakage current : Type CF probe to ground(Only for Europe)
Page 541
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL 10-8-13 Total Patient leakage current : Mains to Type CF probe (Only for Europe) Reference the procedure in the IEC 60601-1 Ed 3. Measure leakage current flow from mains to all ECG leads connected together.
Page 542
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Section 10-9 When There's Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Follow the below checklist in case of earth and/or enclosure leakage current failure.
Page 543
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number:...
Page 544
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL ELECTRICAL SAFETY Outlet and Ground Continuity test Max Value Value Outlet/Ground Continuity Comments Allowed Measured Outlet (correct ground &wiring config.)
Page 545
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Type CF Patient Leakage Current Test - ECG Leads ECG lead 1 Tester Tester Polarity Max Value...
Page 546
GE H DRAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Page 547
GE H RAFT ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R DRAFT (J 17, 2013) EVISION ANUARY ROPRIETARY ERVICE ANUAL Total Patient Leakage Current - Type CF probe to Ground Tester Tester Polarity Max Value Value...
Page 548
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Numerics Replacement Procedure, 8-51 2D Mode Checks Replacement Procedure, 8-77 Functional Checks, 4-8 Cine Functional Checks, 4-24 Collecting System Information, 7-2 Color Flow Modes...
Page 549
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Electromagnetic Compatibility, 1-15 7-13 EMI Protection, 3-8 How to install/use the Process Logger, 7-27 Erasing DVD/CD, 4-33 How to use the Auto Tester, 7-27, , 7-30, , 7-32...
Page 550
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL Caster Cap, 8-51 CBP, 8-77 Network IP Address Configuration, 3-70 CPS, 8-59 CSI, 8-75 DVD, 8-79 DVR, 8-69 OPIO Sub PWA HDD, 8-67...
Page 551
GE H ® P8 / V ® P6 EALTHCARE OLUSON OLUSON DIRECTION 5459672-100, R EVISION ROPRIETARY ERVICE ANUAL How to change Keyboard Layout, 3-58 How to change Language, 3-56 Warnings and Cautions, 1-7 How to change the EUM Language, 3-56...
Page 552
GE H EALTHCARE 5459672-100, R IRECTION EVISION ROPRIETARY ERVICE ANUAL 1 - 1...