MN1-5064 Rev.16 Introduction This is an instruction for model UST-52105, an ultrasound probe. Read the manual carefully before using the instrument. Take special note of the items in section 1, "Safety Precautions". Keep this manual securely for future reference. The CE mark on the probe indicates that this probe is valid when it is connected to equipment bearing the CE mark that is specified as available in section 2 of this document.
MN1-5064 Rev.16 1. Safety Precautions 1-1. Intended use This probe is intended for use by a doctor or other qualified operator when placed into direct contact with the skin making ultrasonic observations of surrounding organs. Caution Do not use this equipment for other than its intended purpose. Use for other purposes can cause burns or other injuries to the patient or operator.
MN1-5064 Rev.16 1-2-1. Warnings and safety information Warning Follow the information in this manual and the documentation supplied with any equipment used together with this probe. Use that is not in accordance with the supplied documentation can result in a serious or moderate injury, equipment breakdown, or physical damage that impairs operation.
MN1-5064 Rev.16 Scan for the minimum length of time necessary for the diagnosis and at the lowest suitable output. Overuse can adversely affect the internal tissues of the patient. For details about the acoustic output, please refer to the documentation supplied with the ultrasound diagnostic instrument.
Page 11
MN1-5064 Rev.16 Label 2 This instrument complies with Directive 93/42/EEC relating to Medical Device and Directive 2011/65/EU relating to RoHS. IPX7 mark IPX7 See section 2-2, “Specifications”. Type BF applied part Do not waste the instrument as general waste. Comply with a local regulation.
Page 12
MN1-5064 Rev.16 (2) Storage case Label A Label B...
Page 13
MN1-5064 Rev.16 Label A(Examples) Model Serial No. Label B(Examples) Hitachi Medical Systems GmbH Otto-von-Guericke-Ring 3 D-65205 Wiesbaden, Germany 2016-09 P-3258F This instrument complies with Directive 93/42/EEC relating to Medical Device and Directive 2011/65/EU relating to RoHS. DATE OF MANUFACTURE (in case of 2016-09)
MN1-5064 Rev.16 2. Specifications and Parts name 2-1. Principles of operation This probe and the ultrasound diagnostic instrument enable image diagnosis using ultrasonic waves. These instruments operate under the principles described below. (1) When an electric pulse signal is applied from the transmitter to the transducer of the probe, the transducer operates by converting electrical vibrations to mechanical vibration energy for emitting pulse-shaped ultrasonic waves into the body part contacting the transducer or into liquid or other medium.
MN1-5064 Rev.16 2-2. Specifications Application regions: Circulatory organs, general areas Form of application to patient: Surface Connectable instruments: SSD- α5, SSD- α10, Prosound α 7, Prosound F75, ARIETTA 70 Field of view: 90° Frequency: 1.5 to 4.3 MHz Cable length: 1.8 m Weight: 900 g...
MN1-5064 Rev.16 2-3. Performance For measurement tolerances, operating tolerances and other data, refer to the instruction manual for the ultrasound diagnostic instrument. 2-4. Name of each parts Probe tip Cable This area is held during operation. This cable propagates the ultrasonic signals that are sent and received.
MN1-5064 Rev.16 2-5. Environmental conditions Use and store the probe under the following conditions. 2-5-1. Operating environmental conditions Ambient temperature: 10°C to 40°C 50°F to 104°F Relative humidity: 30% to 75% Atmospheric pressure: 700 hPa to 1060 hPa Altitude: 3,000 m or less 2-5-2.
MN1-5064 Rev.16 3. Preparations for Use 3-1. Start up check 3-1-1. Visual check Visually check the probe tip, ultrasonic irradiation area, cable and connector. If any holes, indentations, abrasion, cracks, deformation, looseness, discoloration, or other abnormalities are found, do not use the equipment. 3-1-2.
MN1-5064 Rev.16 4. Usage 4-1. Operation Bring the ultrasonic irradiation area of the probe into contact with the skin surface. An image of the region of interest is displayed on the monitor of the ultrasound diagnostic instrument. For details on displaying and adjusting the screens, see the documentation supplied with the ultrasound diagnostic instrument.
MN1-5064 Rev.16 4-2. Connecting to the ultrasound diagnostic instrument The lock lever of the connector moves over the range shown in the figure at right. Align the ○ mark with the LOCK or RELEASE position and lock or release the probe connector. Connect the probe to the electronic probe connecting socket of the diagnostic instrument ( probe connector ) by following the procedure below.
MN1-5064 Rev.16 4-3. Removing from the ultrasound diagnostic instrument The lock lever of the connector moves over the range shown in the figure at right. Align the ○ mark with the LOCK or RELEASE position and lock or release the probe connector. Use the procedure below to remove the probe from the electronic scan probe connector of the ultrasound diagnostic instrument ( probe connector ) .
MN1-5064 Rev.16 4-4. Actions to be taken when an abnormal state is detected 4-4-1. Ensuring safety of patients Immediately move the probe away from the patient and quit operation. Keep the patient in safe condition and administer the required medical treatment. 4-4-2.
Applicable cleaning, disinfection and sterilization methods for each product are listed in the Table 1. The detail of each method is described in Chapter 5-2. Table 1 Applicable cleaning, disinfection and sterilization methods Cleaning Disinfection Sterilization Model UST-52105 Note: X means “Applicable” : Automated Need waterproof cover : Liquid sterilization USA only -19-...
MN1-5064 Rev.16 5-1. Precautions for cleaning, disinfection and sterilization The following warnings and cautions must be observed when cleaning, disinfecting and sterilizing the probe and accessories. Warning Wear protective gloves and other protective gear during cleaning, disinfection and sterilization. Handling of the probe with your bare hands before sterilization can result in an infection. After finishing soaking the probe in cleaning agents, thoroughly wash it with running water.
MN1-5064 Rev.16 5-2. Reprocessing instruction according to ISO 17664 Take care about clean circumstances before using the probe on the next patients. If processors reprocess this equipment, refer to these instructions. Table 1 ・ The probe is delivered unsterile. Prior to the first use, reprocess the probe. ・...
Page 28
MN1-5064 Rev.16 Flowchart of reprocessing process of this probe is as follows: Point of use (Pre-cleaning) Containment and transportation Manual cleaning Rinsing after manual cleaning Drying Manual disinfection Rinsing after manual disinfection Drying Maintenance, inspection and testing Packaging Sterilization -22-...
MN1-5064 Rev.16 5-3. Point of use (Pre-cleaning) In the operating room after use of the probe Probe 1) Flush patient’s blood or fluid by tap water directly after use until the surface looks visually clean. 2) Wipe the whole surface of the probe by gauze pad and remove superficial visible impurities until the surface looks visually clean.
MN1-5064 Rev.16 5-5. Manual cleaning and disinfection Prepare following items before manual cleaning and disinfection. Probe 1) Detergent: ENZOL /Cidezyme (Johnson & Johnson, #2258) or another cleaning agent with approved ® ® material compatibility for this medical device. 2) Disinfectant: Cidex OPA (Johnson &...
MN1-5064 Rev.16 5-5-2. Manual disinfection Probe 1) Before immersing the equipment, it is recommended to test the concentration of disinfectant solution before each usage. The solution Cidex OPA is ready for use and does not need to be diluted. ® Test strips to verify that the appropriate concentration of Cidex OPA is correct are available by ®...
MN1-5064 Rev.16 5-7. Applicable cleaners and disinfectants / Suppliers List The applicable chemical solutions are listed below. General name Trade name Manufacturer ® ADVANCED STERILIZATION PRODUCTS ENZOL /Cidezyme ® ® Enzyme cleaning agent A Johnson & Johnson company Practical liquid 0.8V/V% Division of Ethicon, Inc.
Page 34
MN1-5064 Rev.16 High-level disinfection General name Trade name Manufacturer PERASAFE Hydrogen peroxide ANTEC INTERNATIONAL Practical liquid 1.62W/V% Acecide ® Peracetic acid Saraya Co., Ltd. Solution 6% ® WAVICIDE -01 * Glutaraldehyde Medical Chemical Corporation Solution 2.65% STERANIOS * Glutaraldehyde Laboratoires ANIOS Solution 2.0% ®...
MN1-5064 Rev.16 5-8. Drying Probe 1) Wipe the probe with single use, fluff free wipe or towel for removing moisture on the surface of the equipment. 2) If using drying heater for medical equipment, the temperature limit is a maximum of 60 °C [140 °F]. Dry until no visible moisture is left.
MN1-5064 Rev.16 5-11. Sterilization See “Table 1. Applicable cleaning disinfection and sterilization methods” for available sterilization methods Follow the instructions of the sterilizer manufacturer regarding usage, temperature and sterilization-time etc. Handling and maximum input to chamber of sterilizer should be according to operation manual of the sterilizer.
MN1-5064 Rev.16 6. Storage 6-1. Actions before storing the probe When the probe will not be used for an extended period of time, perform the procedures described in section 5 “Cleaning, disinfection and sterilization" and then store it in its storage case. 6-2.
MN1-5064 Rev.16 7. Moving and Transporting 7-1. Moving and transporting In this section, moving refers to "carrying of the probe within a facility" and transporting refers to "transferring using a vehicle or sending the probe for repairs". 7-2. Preparing the probe and accessories for moving Store in the storage case after performing the procedure in section 5 “Cleaning, disinfection and sterilization".
MN1-5064 Rev.16 8. Periodic Inspection 8-1. Safety tests The safety tests should be conducted at least once a year by a qualified technician. The test record should be stored for future reference. Remarks 1 Qualified technician: personnel for conducting safety tests of medical electrical equipment. If the user requires an appropriate qualified technician, service personnel trained by us can conduct a test at the user’s expense.
MN1-5064 Rev.16 8-2. Testing of measurement tolerances Perform the measurements specified below using an ultrasonic phantom* at least once per year. The test record should be stored for future reference. • Sensitivity • Resolution Remarks Make a copy of the Measurement accuracy inspection data sheet provided in the instruction manual for the ultrasound diagnostic instrument.
MN1-5064 Rev.16 10. Disposal of the Device Recycle or dispose this equipment properly in compliance with the Waste Management and Public Cleansing Law. Caution Before disposing the equipment, disinfect or take other infection-prevention measures. Disposal of the equipment without taking the proper preventative measures can lead to infection. Waste Electrical and Electronic Equipment (WEEE) Directive The illustration on the right is required by the EU WEEE Directive to appear on all electrical and electronic equipment.
Page 48
Hitachi, Ltd. 2-16-1, Higashi-Ueno, Taito-ku, Tokyo, 110-0015, Japan ■Contact +81-3-6284-3668 http://www.hitachi.com/businesses/healthcare/index.html Overseas Offices: Hitachi Medical Systems GmbH Otto-von-Guericke-Ring 3 D-65205 Wiesbaden, Germany EU Importer: Hitachi Medical Systems Europe Holding AG Address: Sumpfstrasse 13 CH-6300 Zug, Switzerland Distributor -42- MN1-5064 Rev.16 ’17.01.11...