GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Page 1-2...
Page 4
GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall:Tel: +49 (0) 212 28 02 207 Deutschland GmbH & Co KG Beethovenstrabe 239, Postfach 11 05 60 D-42655 Solingen, Germany...
Page 5
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change 2002.11.21 Initial Release 2009.6.13 Correction for adding Functional Check Procedure after FRU replacement List of Effected Pages PAGES REVISION PAGES...
Page 6
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Page A-4...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 8
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Electrical Requirements ........2 - 3 EMI Limitations .
Page 9
Possible Behavior of the First Power Up After Application Load ..3 - 35 GE USB Control Panel Driver Loss ......3 - 35 Restore Patient, Report, and User-Defined Presets .
Page 10
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Backend Processor ......... . .5 - 16 Patient I/O (Option) .
Page 11
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 12
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL OP Panel LCD Encoder Assy ........8-29 Covers .
Page 13
Loading Applications Software ........8-105 Standard Configurations and Functional Checks for LOGIQ 5 Software . . 8-108 Mandatory Drivers .
Page 14
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL DMC (DMC User Only) ........8-149 USB Printer (USB Printer User Only) .
Page 15
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Isolated Patient Lead (Sink) Leakage-Isolation Test ....10 - 18 Page 13...
Page 16
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Page 14...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ 5 scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 LOGIQ™...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
Page 20
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons LABEL/SYMBOL...
Page 22
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "CAUTION - Dangerous voltage" (the lightning flash with arrowhead in equilateral triangle) is Various used to indicate electric shock hazards.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-3-3 Mechanical Safety (cont’d) NOTE: Special care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position. • Lock the wheels (brake) •...
Page 25
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-3-5 Label Locations NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. Figure 1-1 OUTSIDE MARKINGS OF LOGIQ™ 5 (Monitor) 1.) Caution for Possible injury : “Never put any device onto the Monitor”...
Page 26
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-3-5Label Locations (cont’d) 220-240V 500VA Max Including front printer panel Europe/USA Console (220V) 100~120V 500VA Max Including front printer panel Japan/USA/Asia Console (100~120V) Circuit Breaker Circuit Breaker...
Page 27
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-3-5Label Locations (cont’d) GE ULTRASOUND KOREA LTD. LOGIQ 5 1250VA LOGIQ 5 GE ULTRASOUND KOREA 1250VA 120V 1250VA GE ULTRASOUND Korea Korea 220V Console China 220V Console Figure 1-3 OUTSIDE MARKINGS OF LOGIQ™...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
Milwaukee, WI 53219 1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: +49 212 2802 208 GE Ultraschall Deutschland GmbH& Co. KG +49 212 2802 207 BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Germany Fax: +49 212 2802 431 Tel: +65 291-8528...
Page 31
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 1-5-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer FAX Number GE Ultrasound Korea 462-120 65-1, Sangdaewon-dong, Joong-won-Gu, 82-31-740-6436 Seongnam-Si, Gyeonggi-do, Korea Chapter 1 Introduction 1-15...
Page 32
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 1-16 Section 1-5 - Customer Assistance...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 5.
2-2-2 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Page 36
2-2-2-5 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 30° C Storage:--10° to 60° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals. 2 - 8 Section 2-3 - Facility Needs...
Scale : Each square equals one square foot FILM VIEWER LINEN SUPPLY SINK PROBES/SUPPLIES FILM SUPPLIES EXTERNAL PERIPHERALS DEDICATED POWER OUTLETS FOOT LOGIQ 5 CONSOLE DEDICATED ANALOG TELEPHONE STOOL LINE FOR CONNECTION TO INSITE EXAMINARION GE CABINET TABLE 24 IN. FOT SOFTWARE 76 IN.
Page 42
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
Page 43
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) LOGIQ™ 5 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default...
Page 44
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 2 - 12 Section 2-3 - Facility Needs...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the LOGIQ™ 5 weighs approximately 180 kg. (397 lb) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
Page 47
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
Page 48
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
Page 49
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP COVER up and off. Top Cover Figure 3-8 Removing the top Cover Chapter 3 Installation...
Page 50
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 3.) Remove the MONITOR CAP up and off.. Monitor Cap Figure 3-9 Removing the monitor cap 3 - 6...
Page 51
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three PLASTIC JOINTs from the OUTER SLEEVE 5.) Remove the OUTER SLEEVE. 6.) Remove the INNER SLEEVE.
Page 52
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 7.) Remove the PLASTIC BAG. 8.) Lift the Monitor up by pressing the UP/Down Release Button. 9.) Remove the MONITOR SUPPORTER.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-3-1 Moving into Position Do not lift the unit by the Keyboard. CAUTION Do not tilt the unit more than 5 degrees to avoid tipping it over.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
Page 56
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Scanner Power On (cont’d) 4.) Press the ON/OFF key at the front of the System once. Composite B/W Printer Power 220-240V 500VA Max Including rear panel...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-5-1-3 Power Up Sequence 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. 2.) Then Start Loader display will be shown on the Monitor display. Then the scanner is booted up automatically.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-5-3 Transducer Connection 1.) Connect a transducer to one of the three rightmost transducer receptacle as follows: A.) Ensure that the transducer twist lock lever to the horizontal position.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 System Settings Table 3-1 System Configuration Configuration Category Description Enables the user or service personnel to set the date, time, unit, language, basic information about...
Page 60
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-1-1 System Settings (cont’d) 5.) Select Utility > System. 6.) Set the Hospital name, Department, Date and Time, Language, and Units. Figure 3-19 Setting Display Click on Utilities to terminate the utility function.
Page 61
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-1-2 Physical Dimensions The physical dimensions of the LOGIQ™ 5 unit are summarized in Table 3-2 on page 3-17 . The Size of LOGIQ™ 5, with monitor and peripherals Table 3-2 Physical Dimensions of LOGIQ™...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-2 Electrical Specifications Table 3-4 Electrical Specifications for LOGIQ™ 5 System Voltage Current Frequency 110 ~ 120 VAC 10~12A 50~60Hz 220 ~ 240 VAC 5~5.5A 50~60Hz 3-6-3...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-4 External I/O Connector Panel Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector power connector and control connections for VCR, printer, and service tools.
Page 64
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-4-1 External I/O Pin Outs Pin No. Signal Pin No. Signal Table 3-6 Pin Assignments of RS232C for Remote 1 and Remote 2 Pin No. Signal Pin No.
Page 65
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-4-1 External I/O Pin Outs (cont’d) Pin No. Signal Pin No. Signal +5 VDC DATA + DATA - Table 3-9 Pin Assignments of USB Pin No. Signal Pin No.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 3-6-5 Video Specification Video specifications may be needed to be able to connect laser cameras or other devices to the LOGIQ™ 5. Table 3-14 Video Specifications 1024x768...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-7 Available Probes See Specifications in the LOGIQ™ 5 User Reference Manual for probes and intended use. See Chapter 9, for part numbers to be used when ordering new or replacement probes.
Page 68
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-9 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-10 Loading Base System Software 3-10-1 Loading Base system Software For more information on loading base system software, refer to Section 8-10 "Software Loading Procedure" on page 8-166.
Page 70
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 3-11 System Setup 3-11-1 System Setup For more information on System setup, refer to Section 8-11 "Installing Base Software Load Image" on page 8-178. 3 - 26...
Page 71
User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-12-2-1 Reference off-board peripherals and options None.
Page 72
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 3 - 28 Section 3-10 - Loading Base System Software...
Page 73
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 5 scanner diagnostics by using the built-in service software, and power supply adjustments.\...
Page 74
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
Page 75
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 5.) Press the ON/OFF key at the front of the System once. Composite B/W Printer Power 220-240V 500VA Max Including rear panel Figure 4-2 Power On/Off Standby Switch Location...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-2-2 Switch off the System 1.) Switch OFF the Circuit Breaker at the back of the scanner. 2.) Disconnect the Mains Power Cable is necessary. For example: Servicing or relocating the scanner.
Page 77
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-3-1 Control Panel (cont’d) 10.)Mode/Gain Keys 11.)Imaging/Measurement Keys 12.)Depth 13.)Imaging Feature Keys 14.)Freeze and Print Keys 15.)Probe and Cord Holder 16.)Gel Holder 17.)Power Button 4-3-3-2 Power On/Off SwitchTouch Panel Figure 4-5 Touch Panel Display Tour 1.) Patient: Enter Patient screen...
Page 78
1. Institution/Hospital Name, Date, Time, Operator 2. Probe Identifier, Exam Study Identification 3. Patient Name, Patient Identification 4. GE Symbol : Probe Orientation Marker. Coincides with a probe orientation marking on the probe. 5. Image Preview 6. Gray/Color Bar 7. Cine Guage 8.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
Page 80
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Table 4-3 B/M-Mode Functions Provides intuitive representation of non-quantitative B Flow hemodynamics in vascular structures. Sensitivity/PRI Adjusts the sample rate for the flow signal. Background On lets you view the anatomy roadmap;...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-5 System CFM and PWD Checks Figure 4-8 CFM and PWD Functions Table 4-4 CFM and PWD Functions Adjusts the baseline to accommodate faster or slower blood Baseline flows to eliminate aliasing.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-6 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays.
Page 83
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-6-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. The worksheets are selected from the Measurement Touch Panel.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-7 Probe/Connectors Usage 4-3-7-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-8 Using Cine 4-3-8-1 Activating CINE Press FREEZE, then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/ Stop. To stop CINE Loop playback. press Run/Stop.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-10 Using the CD-RW / MOD (Magneto-Optical Drive) 4-3-10-1 Using the CD-RW Never move the unit with a disk in the CD-RW because the drive actuator will not be locked NOTICE and the CD-RW could break.
Page 87
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-10-2 Using the MOD (Magneto-Optical Drive) The 3.5 inch Magneto-Optical disk drive supports the following densities: Standard: 640MB; 540MB; 230MB; and 128MB Overwrite: 640MB; and 540MB 1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MOD.
Page 88
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-11 Backup and Restore Database, Preset Configurations and Images Avoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator NOTICE unlocked and the MOD susceptible to damage if moved. If forced to use this method, reboot the system, then insert and eject a known good disk using one of the other manual ejection methods.
Page 89
4.) On the monitor display, select BACKUP/RESTORE. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive, System Configuration and User Defined Configuration.
Page 90
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-11-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct MOD or CD-R.
Page 91
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-11-4 Archiving Images To export an exam(s) to a compatible Ultrasound system: 1.) Format the removable media (MOD or CD-ROM). Label the removable media. Answer Yes/OK to the messages.Press Patient. Deselect any selected patient(s) in the search portion of the Patient screen.
Page 92
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-11-4 Archiving Images (cont’d) 4.) In the patient list at the top of the Patient menu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select all the names.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 5 Connect the ECG Harness and check: Table 4-5 ECG Harness Check Step Task to do...
Page 94
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks Step Task to do Expected Result(s) Press (FREEZE) Stop image acquisition. Press (PRINT 1) or (PRINT 2) on the Control The image displayed on the screen is printed on B&W or Color printer,...
Page 95
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-6Monitor Function Checks 4-6-1 CRT Assy function check procedure 1.) Turn on the system and check the completion of Echoloader loading. 2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in Figure 4-19.
Page 96
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-6-4 Rear panel Assy function check procedure 1.) Turn on the system and check the completion of Echoloader loading. 2.) Refer to See “Peripheral Checks” on page 4-22.
Page 97
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-7Keyboard Function Checks 4-7-1 Main Keyboard Assy validation 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Click 'Start' and go to 'Run' .
Page 98
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-1 Main Keyboard Assy validation (cont’d) 6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the scfpapitest program. Figure 4-24 scfpapitest program 7.) Press the keyboard button and rotate rotary button.
Page 99
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Button or Button or Rotary Name The Name displayed in Log Rotary Name The Name displayed in Log Harmonic Harmonics Button PDI Button Reverse Reverse Button PW (Button)
Page 100
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-1 Main Keyboard Assy validation (cont’d) 8.) Check the 'Fast' radio button into Indicator box on scfpapitest program and press 'Set All' button. Figure 4-25 Fast button After pressing Fast radio button, indicators should be blanked on the frontpanel.
Page 101
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-2 OP Panel Encoder Assy Validation 1.) Follow up the procedure 8-4-1 to run the scfpapitest program. 2.) Check the each Encoder in Table 4-8 on page 26.
Page 102
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-8 OP Panel LCD Assy Validation 1.) Boot up the system. 2.) Follow up the process from step1 to step7 of Section 4-7-1. 3.) Touch the touchpanel. Figure 4-26 Touchpanel In case of touch point is displayed in log of scfpapitest.exe while you press the touchpanel, LCD assy...
Page 103
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-9 OP Panel LCD Encoder Assy Validation 1.) Boot up the system and run the scfpapitest program. Follow up the procedure from step 1 to step 7 of Section 4-7-1 to run the scfpapitest program.
Page 104
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-10 Speaker Assy Validation 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Move to trackball to the bottom of main monitor.
Page 105
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-7-10 Speaker Assy Validation (cont’d) 6.) At the Balance in Volume Control, move the gauge to the max left. Figure 4-30 Volume control for left speaker 7.) Move the Volume gauge up and down.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-8 Mechanical Function Checks 4-8-1 Operator I/O Movement Table 4-10 Display platform Maneuverability check Step Task to do Expected Result(s) Pull keyboard lever under the OP Panel to The height of the Monitor and OP Panel will release the lock.
Page 107
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2 Cover Parts Function Validation Table 4-11 Cover parts of the FRU Item Right and Left cover Front cover Rear cover Rear Door cover Top cover Front base cover...
Page 108
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2-1 Task Lamp Assy Check if task lamp is operated. Figure 4-33 Task Lamp Assy 4-8-2-2 Keyboard Front Grip • Check if the keyboard front grip is fastened so that it won't move.
Page 109
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2-4 Gas Spring Assy Check if the up and down mechanism operates rightly. Figure 4-36 Gas Spring Assy 4-8-2-5 Front caster/Rear caster • Check caster locks for proper operation.
Page 110
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2-6 Bumper Set • Check if the bumper set is fastened so that it won't move. • Check if the bumper set can not be removed easily. Figure 4-38 Bumper Set...
Page 111
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2-8 OP Panel Encoder Assy, TGC Knob Set, OP Panel LCD Encoder Assy, Menu Encoder Assy • Check if the each button operates properly. • Check if encoder knobs are in the center.
Page 112
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-8-2-9 Monitor Cover Set • Check if the FRU parts are assembled tightly by naked eye and hands. • Check if there are dents, scratches, or cracks on the FRU parts.
Page 113
3.) Log into "Service Login". Refer to Section 5-10-2 - Global Service User Interface (GSUI) on page 5-33. 4.) Click "Diagnostics" on top menu. 5.) On left folder window, select 'LOGIQ 5 Diagnostics'=>'FRU test'=>'RLY'. 6.) Select 'RLY Full Test' and execute test. 7.) All diagnostic test items should be passed. Refer to the figure below.
Page 114
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'PMP Full Test' 4.) Execute the diagnostic test.
Page 115
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'RDS Full Test' 4.) Execute the diagnostic test.
Page 116
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'ATD Full Test' 4.) Execute the diagnostic test.
Page 117
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'FECII Full Test' 4.) Execute the diagnostic test.
Page 118
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'CWDOP Full Test' 4.) Execute the diagnostic test.
Page 119
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-8 Backplane Assy function check procedure 1.) Turn on the system and check the completion of Echoloader loading. 2.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4- 7.
Page 120
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-11 BEP Assy function check procedure 4-9-11-1 HDD, Memory, CPU Diagnositc Test 1.) The Start up Screen will be shown on the Monitor display when the system is Turned ON.
Page 121
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-11-1 HDD, Memory, CPU Diagnositc Test (cont’d) 6.) Memory Test Select Memory test manu, start to testing. Check display 'Pass' massage Figure 4-51 Memory Test 7.) CPU Test Select CPU test manu, start to testing.
Page 122
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-11-2 Basic Function Test 1.) Turn on the system and check the completion of Echoloader loading. 2.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4- 7.
Page 123
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'VIC Full Test' 4.) Execute the diagnostic test.
Page 124
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-13 SMPS, HDD, ODD Assy function check procedure 4-9-13-1 SMPS Assy Test 1.) Turn on the system and check the completion of Echoloader loading. 2.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4- 7.
Page 125
1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-33. 3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'PC2IP Full Test' 4.) Execute the diagnostic test.
Page 126
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 4-9-15 AC Power Assy function check procedure. 1.) Turn on the system and check the completion of Echoloader loading. 2.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4- 7.
Page 127
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 4-10 Site Log Table 4-12 Site Log Date Service person Problem Comments Chapter 4 Functional Checks 4-55...
Page 128
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 4-56 Section 4-10 - Site Log...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Fuctions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ™ 5’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
Page 130
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-2 General Information LOGIQ™ 5 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and Phono. A Doppler probe may also be connected and used.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-3 Block Diagram SIG Assy PMP Assy VCA2613 ATD Assy CWDOP VCA2613 Assy Pulse Drive r Tx Pulse Driver X16 VCA2613 X 16 channe l (Option)
Page 132
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-3 Block Diagram (cont’d) SIG Assy RDS Assy FEC ASSY Memory Control Buffer 12 Bits ADC 16 MB CWDOP Buffer 12 Bits ADC Buffer 12 Bits ADC...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-4 Front End Figure 5-4 The Front End The Front End generates the strong burst transmitted by the probes as ultrasound into the body. It also receives weak ultrasound echoes from blood cells and body structure, amplifies these signals and converts them to an digital signal.
Page 134
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-4-1 RLY Assy RELAY Probe #1 RELAY To LMT Probe #2 Probe #1 DCWD Probe ID Control Control Temperature Probe I/F FPGA Circuit Sensor JUSC I/F Figure 5-5 RLY Block Diagram RLY ASSY contains of 3 phased array probe connectors and provides switchable connection between probes and 64ch transmitters/receivers.
Page 135
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Detects resistance of a thermistor in a probe head as voltage when the probe is selected.(PTEMP) This signal is connected to FEC. • LED Blinking The LED in a probe blinks when the probe is selected.
Page 137
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-4-4 THV Hi g h HV Switch Circuit THV Low TPG2 ECL CLK 40M Q1 CLKQ1 FET Driver x 16 ch ECL to TTL +TxP[15:0] Tx Pulse[63:48]...
Page 138
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-4-5 5-4-5-1 Overview 1 2 B its C h a n n e l 0 B U F A D C O Q C A R D...
Page 139
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-4-5-2 RDS (cont’d) Reference Voltage Generator : This mode is for imaging. The parameters specific to scan line number are sent to OQCARDs and the start triger for receiving signals is generated.
Page 140
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-4-6 (cont’d) • CLK Generator Block: Generates 40MHz (two phases), and 26.6MHz clock using 160MHz master clock. • Real Time Controller (RTCL): The TRIG cyclical Real Time Control is done by SH4 RISC processor. SH4 is a one of MID BUS agent through SH-PCI bridge.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-5 Backend Processor 5-5-1 BEP with Intel Motherboard To Monitor Patient IO To RAI B'D, RDI B'D in Rear Panel (Option) Battery Power Supply On B'D...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-5-2 VIC Card 50p C o n n e c to r T o R e a r P a n e l fo r re m o te c o n tro l o f V C R...
Page 144
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-6 Top Console The Top Console includes a On/off switch, a keyboard, different controls for manipulating the picture quality, controls for use in Measure & Analyze (M&A), and loudspeakers for stereo sound output (used during Doppler scanning, inside the CRT monitor cover).
Page 145
15 inch high resolution RGB monitor with non interlace scan includes loudspeakers for stereo sound output and microphone to record talking of operator. Monitor have also task lamp to light up the A/N key on keyboard assy Figure 5-15 LOGIQ 5 Monitor Chapter 5 Components and Functions 5 - 17...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-7 External I/O (Rear Panel) The External I/O is the interface between the scanner and all external items, located at the rear side of the scanner.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2 Peripherals and Cable connection A VCR, a Black & White Video Printer and a Color Printer may be installed onboard the scanner. These devices are connected to the External I/O (Rear Panel) or Front Panel in the case of B/W Printer.
Page 148
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2-2 Cable Connection - Color Printer Connect the cables listed below between the Color Printer and the console. RS232C CABLE BNC CABLE for VIDEO SIGNAL (supplied with color video printer)
Page 149
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2-2 Cable Connection - Color Printer (cont’d) Connect the cables listed below between the Color Printer and the console. RS232C C ABLE BNC CABLE for VIDEO SI G NAL...
Page 150
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2-2 Cable Connection - Color Printer (cont’d) Connect USB cable listed below between the SONY UP-D21MD Color Printer and the console. Rear Panel of JUPITER SYNC S-VIDEO...
Page 151
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2-3 Cable Connection- VCR Connect the cables listed below between the Sony VCR and the console. RS232C CABLE Rear Panel of JUPITER S. Video CABLE for VIDEO SIGNAL...
Page 152
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-7-2-3 Cable Connection- VCR (cont’d) Connect the cables listed below between the Pansonic VCR and the console. RS232C CABLE Rear Panel of JUPITER VIDEO IN AUDIO VIDEO OUT...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-8 Power Diagrams 5-8-1 Overview The AC Power assy’s main tasks are to supply the various internal subsystems and to galvanically isolate the scanner from the on site Mains Power System and to control power on/off . The voltage to peripherals can be confederated to either 110 VAC or 220 VAC.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-8-3 DC Power (Low Voltage) DC Power are supplied by LV unit in AC Power assy. DC Power of LV unit are power source of all assys in nest and nest fan, etc JPC assy in AC power assy is DC power source also.
Page 156
The REAR ASSY and REAR CONN 220V ASSY are the rear-panel assemblies which interconnect external signals and power with the LOGIQ 5 system. Both of the assemblies are the same except the circuit breaker’s current ratings depending on voltage of the system power source (See Safety described below).
Page 157
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-8-4-2 Safety • Circuit Breaker Table 5-15 Circuit Breaker Specification Power Voltage Rating Peripheral Current Trip Time Reset Time Assembly Rating (A) 1 hour max. @ 5.4A, 25...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 5-9 Air Flow 5-9-1 Air Flow Distribution Figure 5-25 Air Flow Inside the Scanner The four air flow passes allow the scanner to be cooled down as shown in the figure above.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-9-3 Fans Fan for PC Box Fans for AC Power Assy Fans for Nest Assy Figure 5-26 Fans The scanner contains the eight fans at the following positions for producing an air flow.
The Service Platform contains a set of software modules that are common to all ultrasound and cardiology systems containing a PC backend. This web-enabled technology provides linkage to e- Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform will increase service productivity and reduce training and service costs.
Page 161
Global Service User Interface (GSUI) 5-10-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-10-2-2 Service Login Select the wrench icon in the status bar at the bottom of the scan display screen.
Page 162
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-3 Service Home Page The navigation bar at the top of the screen allows the user to select from several tools and utilities. Figure 5-28 Customer Service Home Page...
Page 163
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-4 Error Logs Tab From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users.
Page 164
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-4-1 Logs The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-29 provides a graphical example of the log entries for the System Logs.
Page 165
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-4-2 Utilities The two sub-menus of the Utilities category are Plot Log, and Plot Page. Plot Log Plot Page Figure 5-30 Utilities Sub-Menus • Plot Log Allows for the color coded plot of all Log contents with the package on the ‘x’ axis and incident count on the ‘y’...
Page 166
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-4-4 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page. Figure 5-32 Exit Log Sub-Menu 5 - 38 Section 5-10 - Service Platform...
Page 167
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-5 Diagnostics Detailed Diagnostic information is found in 7 Diagnostics/Troubleshooting. Figure 5-33 User Diagnostic Page 5-10-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level diagnostic selection based on the user’s level and login access permissions.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. Figure 5-34 Image Quality Page 5-10-7 Calibration The Calibration page is intended to contain the tools used to calibrate the system.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorization control for remote access to the configuration subsystem.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 5-10-10 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system. Figure 5-38 Part Replacement Page 5-10-11 The PM page is intended to contain the tools used in periodic maintenance of the system.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-3-1 LV Unit Disassembly This system contains three power supply modules; HV unit, LV unit, and ATX PS. However, the LV unit only can be adjusted. 6-3-1-1 Tools •...
Page 173
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-3-1-4 Removal Procedure (cont’d) 4.) Unscrew four 4 screws and disconnect twenty (20) connectors from the front side of the LV unit using long-nose gripper. Screw Screw...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-3-2 Adjustments Procedures 1.) Using the following VRs, DC output can be adjusted. Table 6-20 DC Output Specification for LV Unit Specifications 3.3V -15 V VR6 VR7...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 6-4 Reloading the Probe Data This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details, refer to Section 7, Troubleshooting.
Page 176
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 6-4 Reloading the Probe Data (cont’d) 5.) Then the Maintenance Access Menu will be shown on the monitor display. Enter the password to logon the system. The User Name is not necessary.
Page 177
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 6-4 Reloading the Probe Data (cont’d) 7.) Windows Desktop screen is displayed. Figure 6-8 Windows 2000 Desktop Chapter 6 Service Adjustment 6 - 7...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 6-5 Monitor and LCD Adjustments 6-5-1 Contrast and Brightness Adjustment To adjust the contrast and brightness: 1.) Press the Toggle button (1) for contrast and brightness. Confirm that the contrast (or brightness) indicator is displayed on the monitor.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-5-2 LCD Adjustment 6-5-2-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench • Long-nose gripper 6-5-2-2 Needed Manpower • 1 person, 15 minutes + travel 6-5-2-3 Preparations •...
Page 180
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-5-2-4 Removal Procedure (cont’d) 3.) OSD Menu is shown on the LCD screen. Refer to Figure 6-11 on page 6-10 . Contrast Adjust Menu Color Adjust Menu...
Page 181
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 6-5-2-4 Removal Procedure (cont’d) 6.) Use Increase or Decrease button to adjust the Brightness. Refer to Figure 6-12 on page 6-11 . Figure 6-12 Brightness Adjustment 7.) Move the cursor to execute Auto Adjust.
Page 182
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 6 - 12 Section 6-3 - Power Supply Adjustment...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 7-3 Gathering Trouble Data 7-3-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
Page 185
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-3-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
Page 186
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-4-1 Check and Record the P4 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. The destinations list displays the following information:...
Page 188
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk, MOD or CD-R.
Page 190
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 7-5 Diagnostics As described in - Components and Fuctions (Theory), the service platform uses a web-based user interface (UI) to provide access to common service components.
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 7-6 Common Diagnostics Common Diagnostics is the first main heading under the GSUI heading of Diagnostics. Figure 7-7 shows the top level menu choices under Common Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections below.
Page 192
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-6-2 PC Diagnostics (Non-Interactive Tests) 7-6-2-1 CPU Tests 7-6-2-2 Hard Drive Tests 7-6-2-3 Memory Tests 7-6-2-4 CD/DVD Drive Test 7-6-2-5 Video Test 7-6-2-6 USB Test 7-6-2-7 PCI Board Configuration Test...
Page 193
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 7-6-3 PC Diagnostics (Interactive Tests) 7-6-3-1 Keyboard Test 7-6-3-2 Audio Test 7-6-3-3 Monitor Test Chapter 7 Diagnostics/Troubleshooting 7 - 11...
Page 194
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 7-7LOGIQ5 Diagnostics When you select a test to run, the display changes as shown in Figure 7-8 on page 7-13 1.) Instructions Frame Displays either a test-specific text or the default instructions.
Page 195
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL • Cable to RDi board in Rear Panel and Keyboard for Power switch. Figure 7-8 Diagnostic Group Selections - SCB Tests • Connect J4 connector in RDi board with CON4 connector in keyboard.
Page 196
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 7 - 14 Section 7-7 - LOGIQ5 Diagnostics...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Section Description Page Number Overview...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-3 Monitor 8-3-1 CRT Assy Purpose: This is a description on how to remove and replace the CRT Assy. 8-3-1-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 199
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-1-4 Removal procedure (cont’d) 3.) Remove 3Connectors (CN1 - CN3) from the Monitor. Refer to Figure 8-2. Figure 8-2 Disconnect three(3) connectors 4.) Lift the Monitor Assy upward. Refer to Figure 8-3.
Page 200
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-1-6 Monitor Calibration Purpose: This procedure describe the monitor calibation procedure for newly installed monitor or CRT assembly. The newly installed monitor should be adjusted the gain uniformity for each RGB analog singal which come from VGA port of BEP assy.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-2 Monitor Cable Assy Purpose: This is a description on how to remove and replace the Monitor Assy. 8-3-2-1 Tools • Common pillips screwdrivers • Cutter 8-3-2-2 Needed Manpower •...
Page 202
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-2-4 Removal procedure (cont’d) 4.) Disconnect two connectors on the other end (4-5). Refer to Figure 8-8. Figure 8-8 Disconnect two(2) Connectors (5-6) 5.) Pull the Cable Assy out from the OP Rear Cover.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-3 Monitor Cover Set Purpose: This is a description on how to remove the Monitor Cover Set. 8-3-3-1 Tools • Common pillips screwdrivers 8-3-3-2 Needed Manpower • 1person, 10 minutes + travel...
Page 204
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-3-4 Removal Procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-4 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-4 Monitor S/W Assy Purpose: This is a description on how to remove and replace the Monitor S/W Assy. 8-3-4-1 Tools • Common pillips screwdrivers 8-3-4-2 Needed Manpower •...
Page 206
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-4-4 Removal Procedure (cont’d) 4.) Remove the Monitor S/W Assy.Refer to Figure 8-12. Figure 8-12 Removing Monitor S/W Assy 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-5 Speaker Assy Purpose: This is a description on how to remove and replace the Speaker Assy. 8-3-5-1 Tools • Common pillips screwdrivers 8-3-5-2 Needed Manpower • 1person, 10 minutes + travel...
Page 208
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-5-4 Removal Procedure (cont’d) 2.) Unscrew four screws (3-6) on the right side and (7-10) on the left to remove the speakers. Refer to Figure 8-14. Figure 8-14 Speaker Assy Disassembly 3.) Unscrew four screws (7-10) to remove the speaker on the right.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-6 Task Lamp Assy Purpose: This is a description on how to remove and replace the Task Lamp Assy. 8-3-6-1 Tools • Common pillips screwdrivers 8-3-6-2 Needed Manpower •...
Page 210
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-3-6-4 Removal Procedure (cont’d) 2.) Disconnect two (3-4) connectors. Refer to Figure 8-16. Figure 8-16 Disconnect two connectors 3.) Unscrew two (5-6) screws to remove the TaskLamp Assy. Refer to Figure 8-17.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-4 Keyboard Block 8-4-1 KeyBoard Assy Purpose: This is a description on how to remove and replace the Keyboard Assy. 8-4-1-1 Tools • Common pillips screwdrivers. 8-4-1-2 Needed Manpower •...
Page 212
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-1-4 Removal Procedure (cont’d) 4.) Disconnect three connectors (6-8) from the HUB board. Refer to Figure 8-20. HUB Board Figure 8-20 Disconnect 3 connectors from the HUB 5.) Disconnect two connectors (9-10). Refer to Figure 8-21.
Page 213
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-1-6 Mounting procedure 1.) Confirm the configuration jumper setting The new keyboard with engineering assy P/N rev ”-7” have a jumper for USB device power configuration for keyboard function keys. Before install the new keyboard , the jumper cap should be attached in proper place.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-2 KeyBoard Front Grip Purpose: This is a description on how to remove and replace the Keyboard Front Grip. 8-4-2-1 Tools • Common pillips screwdrivers 8-4-2-2 Needed Manpower •...
Page 215
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-2-5 Mounting procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-9 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-3 Probe Holder Purpose: This is a description on how to remove and replace the Probe Holder. 8-4-3-1 Tools • Not Required 8-4-3-2 Needed Manpower • 1persons, 1 minute + travel...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-4 ECG Cable Hook Purpose: This is a description on how to remove and replace the TGC Cable Holder. 8-4-4-1 Tools • Common pillips screwdrivers 8-4-4-2 Needed Manpower •...
Page 218
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-4-4 Removal Procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-11 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-5 OP Panel Encoder Assy Purpose: This is a description on how to remove and replace OP Panel Encoder Assy. 8-4-5-1 Tools • Common pillips screwdrivers. • Longnose.
Page 220
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-5-4 Removal Procedure (cont’d) 2.) Using longnose, pull out encoder nuts and encoder washers from each encoder. Refer to Figure 8- Figure 8-28 Pulling Out encoder nuts 3.) Turn the keyboard to back side.
Page 221
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-5-4 Removal Procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-12 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-6 OP Panel TGC Assy Purpose: This is a description on how to remove and replace OP Panel TGC Assy. 8-4-6-1 Tools • Common pillips screwdrivers. 8-4-6-2 Needed Manpower •...
Page 223
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-6-4 Removal Procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-13 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-7 TGC Knob Set Purpose: This is a description on how to remove and replace TGC Knobs. 8-4-7-1 Tools • Common pillips screwdrivers. 8-4-7-2 Needed Manpower • 1persons, 5 minutes + travel...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-8 OP Panel A/N Keyboard Assy Purpose: This is a description on how to remove and replace OP Panel A/N Keyboard Assy. 8-4-8-1 Tools • Common pillips screwdrivers...
Page 226
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-8-4 Removal Procedure (cont’d) 3.) Unscrew eleven (11) screws (1-11) to remove the A/N keyboard prefix bracket. Refer to Figure 8-35. (10) (11) Figure 8-35 Unscrew 11 screws 4.) Remove the A/N keyboard prefix bracket...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-9 OP Panel Trackball Assy Purpose: This is a description on how to remove and replace OP Panel Trackball Assy. 8-4-9-1 Tools • Common pillips screwdrivers. 8-4-9-2 Needed Manpower •...
Page 228
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-9-4 Removal Procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-16 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-10 OP Panel HUB board Assy Purpose: This is a description on how to remove and replace HUB board Assy. 8-4-10-1 Tools • Common pillips screwdrivers 8-4-10-2 Needed Manpower •...
Page 230
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-10-4 Removal Procedure (cont’d) 3.) Unscrew six (6) screws (7-12) to remove HUB Board Assy. Refer to Figure 8-41. Figure 8-41 Unscrew 6 screws 4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-11 OP Panel LCD Assy Purpose: This is a description on how to remove and replace LCD Assy. 8-4-11-1 Tools • Common pillips screwdrivers 8-4-11-2 Needed Manpower •...
Page 232
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-11-4 Removal Procedure (cont’d) 2.) Disconnect one connector (11) from the main circuit board in the keyboard. Refer to Figure 8-43. (11) Figure 8-43 Disconnect one connector 3.) Unscrew four (4) screws (12-15) from the LCD unit. Refer to Figure 8-44.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-12 OP Panel LCD Encoder Assy Purpose: This is a description on how to remove and replace OP panel encoder Assy. 8-4-12-1 Tools • Common pillips screwdrivers 8-4-12-2 Needed Manpower •...
Page 234
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-4-12-4 Removal Procedure (cont’d) 3.) Unscrew four (4) screws (7-10) to remove the LCD encoder Assy. Refer to Figure 8-46. (10) LCD Encoder Circuit Board Figure 8-46 Unscrew 4 screws from LCD Encoder Circuit Board 4.) Disconnect one connector (14) from the HUB Board.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-5 Covers 8-5-1 Left Cover Purpose: This is a description on how to remove and replace the Left Cover. 8-5-1-1 Tools • Common pillips screwdrivers 8-5-1-2 Needed Manpower •...
Page 236
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-1-4 Removal Procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-20 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-2 Right Cover Purpose: This is a description on how to remove and replace the Right Cover. 8-5-2-1 Tools • Common pillips screwdrivers 8-5-2-2 Needed Manpower • 1person, 5 minutes + travel...
Page 238
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-2-4 Removal Procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-21 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-3 Rear Cover Purpose: This is a description on how to remove and replace the Rear Cover. 8-5-3-1 Tools • Common pillips screwdrivers 8-5-3-2 Needed Manpower • 1 persons, 15 minutes + travel...
Page 240
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-3-4 Removal Procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-22 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-4 Rear Door Cover Purpose: This is a description on how to remove and replace the Rear Door Cover. 8-5-4-1 Tools • Common pillips screwdrivers 8-5-4-2 Needed Manpower •...
Page 242
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-4-4 Removal Procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-23 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-5 Top Cover Purpose: This is a description on how to remove and replace the Top Cover. 8-5-5-1 Tools • Common pillips screwdrivers 8-5-5-2 Needed Manpower • 1 person, 20 minutes + travel...
Page 244
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-5-4 Removal Procedure (cont’d) 5.) Unscrew four screws (3-6) from the bottom of the Top cover. Refer to Figure 8-53. Figure 8-53 Unscrew 4 screws from the bottom 6.) Perform the following functional tests.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-6 Front Base Cover Purpose: This is a description on how to remove and replace the Front Base Cover. 8-5-6-1 Tools • Common pillips screwdrivers 8-5-6-2 Needed Manpower •...
Page 246
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-6-4 Removal Procedure (cont’d) 4.) Remove the Front Base Cover. Refer to Figure 8-55. Figure 8-55 Removing the Front Base Cover 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-7 Front Cover Purpose: This is a description on how to remove and replace the Front Cover. 8-5-7-1 Tools • Common pillips screwdrivers 8-5-7-2 Needed Manpower • 1person, 20 minutes + travel...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-8 OP Side L Cover Purpose: This is a description on how to remove and replace the OP Side L Cover. 8-5-8-1 Tools • Common pillips screwdrivers •...
Page 249
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-8-4 Removal Procedure (cont’d) 7.) Unscrew five screws (3-7) and remove OP Side L Cover. Refer to Figure 8-58. Figure 8-58 Unscrew 5 screws to remove OP L Cover NOTE: Different screw in length is used in the location Number (7).
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-9 OP Side R Cover Purpose: This is a description on how to remove and replace the OP Side R Cover. 8-5-9-1 Tools • Common pillips screwdrivers •...
Page 251
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-9-4 Removal Procedure (cont’d) 7.) Unscrew two (2) screws (6-7) and remove the OP Curtain R Bracket. Refer to Figure 8-60. OP Curtain R Bracket Figure 8-60 Removing the OP Curtain R Bracket NOTE: Different screw in length is used in the location Number (7).
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-10 OP Rear Cover Purpose: This is a description on how to remove and replace the OP Rear Cover. 8-5-10-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 253
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-10-4 Removal Procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-29 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-11 OP Bottom Cover Purpose: This is a description on how to remove and replace the OP Bottom Cover. 8-5-11-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 255
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-11-4 Removal Procedure (cont’d) 9.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-30 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-12 OP Curtain Cover Purpose: This is a description on how to remove and replace the OP Curtain Cover. 8-5-12-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 257
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-12-4 Removal Procedure (cont’d) 9.) Remove the OP Curtain Cover. Refer to Figure 8-65. Figure 8-65 Remove the OP Curtain Cover 10.)Perform the following functional tests. If all are successful, include the debrief script provided below.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-13 EMI Cover L Purpose: This is a description on how to remove and replace the EMI Cover L. 8-5-13-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 259
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-13-4 Removal Procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-32 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-14 EMI Cover R Purpose: This is a description on how to remove and replace the EMI Cover R. 8-5-14-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 261
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-14-4 Removal Procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-33 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-15 Side Cap Purpose: This is a description on how to remove and replace the Side Cap. 8-5-15-1 Tools • Not required 8-5-15-2 Needed Manpower • 1 persons, 3 minute + travel...
Page 263
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-15-4 Removal Procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-34 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-16 Side Fringe Purpose: This is a description on how to remove and replace the Side Fringe. 8-5-16-1 Tools • Not required. 8-5-16-2 Needed Manpower • 1 persons, 3 minute + travel...
Page 265
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-5-16-4 Removal Procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-35 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-6 Body Block 8-6-1 Rear Handle Purpose: This is a description on how to remove and replace the Rear Handle. 8-6-1-1 Tools • Common pillips screwdrivers •...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-1-4 Removal Procedure (cont’d) 6.) Unscrew two (2) screws (5-6) from the rear side of handle. Refer to Figure 8-71. Figure 8-71 Unscrew two screws 7.) Unscrew four (4) screws (7-10) above top plate surface on the Rear Handle.
Page 268
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-1-4 Removal Procedure (cont’d) 10.)Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-36 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2 Rear Panel Assy Purpose: This is a description on how to remove and replace the Rear Panel Assy. 8-6-2-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 270
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2-4 Removal Procedure (cont’d) 5.) Disconnect 8 connectors (1-8) from the left side of the Panel. Refer to Figure 8-74. External VGA Connector Isolated Power Connector Power Control Connector...
Page 271
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2-4 Removal Procedure (cont’d) 8.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-37 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
Page 272
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2-6 Replace the EEPROM on the RDI Assy in the rear panel assy 1.) Rear panel assy include the eeprom which have the system serial number information. 2.) If the rear panel assy would be replaced with other one, the eeprom in the rear panel assy should be moved to new assy from old one.
Page 273
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2-7 How to Check and write the system serial number on EEPROM (cont’d) Figure 8-77 Window Explorer on OS screen 9.) Run the "PC2IPIIC.EXE" by double click . This execution file is in "C:\Jupiter\target\bin" directory Figure 8-78 “PC2IPIIC.EXE”...
Page 274
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-2-7 How to Check and write the system serial number on EEPROM (cont’d) Figure 8-79 Execute the “PC2IPIIC.EXE” 15.)If the same serial number appear on the next column as below picture, the write procedure is finished.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-3 Gas Spring Assy Purpose: This is a description on how to remove and replace the Up/Down Assy. 8-6-3-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench •...
Page 276
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-3-4 Removal Procedure (cont’d) 3.) Unscrew two (3) screws (4-6) to remove the OP button bracket. Refer to Figure 8-81. OP Button Bracket Figure 8-81 Unscrew 6 screws to remove BRKT 4.) Unscrew four (4) hexagon screws (7-10).
Page 277
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-3-4 Removal Procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-38 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-4 Front Caster Purpose: This is a description on how to remove and replace the Front Caster. 8-6-4-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench • Lifter (180Kg)
Page 279
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-4-4 Removal Procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-39 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-5 Rear Caster Purpose: This is a description on how to remove and replace the Rear Caster. 8-6-5-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench • Lifter (180Kg)
Page 281
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-5-4 Removal Procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-40 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-6 Bumper Set Purpose: This is a description on how to remove and replace the Bumper Set. 8-6-6-1 Tools • Common pillips screwdrivers 8-6-6-2 Needed Manpower • 1persons, 20 minutes + travel...
Page 283
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-6-4 Removal Procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-41 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-6-7 Air Filter Set Purpose: This is a description on how to remove and replace the Air Filter Set. 8-6-7-1 Tools • Not Required. 8-6-7-2 Needed Manpower •...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-7 PCB Boards 8-7-1 RLY Assy Purpose: This is a description on how to remove and replace the RLY Assy. 8-7-1-1 Tools • Common pillips screwdrivers 8-7-1-2 Needed Manpower •...
Page 286
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-1-4 Removal Procedure (cont’d) 4.) Unscrew nine (9) screws (6-14) to remove the RLY Board Cover. Refer to Figure 8-89. (12) (13) (14) (10) (11) Figure 8-89 Removing the RLY board Cover NOTE: Bind head screw is used at the location number (8).
Page 287
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-1-4 Removal Procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-43 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-2 LMT Assy Purpose: This is a description on how to remove and replace the LMT Assy. 8-7-2-1 Tools • Common pillips screwdrivers 8-7-2-2 Needed Manpower • 1 person, 25 minutes + travel...
Page 289
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-2-4 Removal Procedure (cont’d) 3.) Pull the LMT Board out. Refer to Figure 8-92. Figure 8-92 Pulling Out LMT Board 4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-3 BackPlane Purpose: This is a description on how to remove and replace the Backplane. 8-7-3-1 Tools • Common pillips screwdrivers 8-7-3-2 Needed Manpower • 1 person, 30 minutes + travel...
Page 291
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-3-4 Removal Procedure (cont’d) 5.) From the right view, disconnect two connectors (9-10). (11) (12) (10) (13) (14) Figure 8-94 Disconnect 2 connectors from the right 6.) Unscrew four (4) screws (11-14) to remove the Backplane Assy.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-4 PCB Boards Purpose: This is a description on how to remove and replace the PCB Boards. PMP Assy, SIG Assy, FEC Assy, RDS Assy, CWDOP Assy, ATD Assy, HV Assy and VIC Card.
Page 293
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-4-4 Removal Procedure (cont’d) 6.) Disconnect one connector (1) if removing the FEC Assy. Otherwise skip this step. Refer to Figure 8-95. SIG Board Figure 8-95 Disconnecting Cable 7.) Move the upper and lower stopper in the direction indicated by the arrow (1).
Page 294
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-4-4 Removal Procedure (cont’d) 9.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-46 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-5 DC Fan Assy Purpose: This is a description on how to remove and replace the DC Fan. 8-7-5-1 Tools • Common pillips screwdrivers 8-7-5-2 Needed Manpower •...
Page 296
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-7-5-4 Removal Procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-47 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-8 PC Block 8-8-1 BEP (Back End Processor - BEP1, BEP2, BEP3, BT05 BEP3,BEP4) ASSY Purpose: This is a description on how to remove and replace the Parts in the BEP, BEP2 and BEP3 Assy.
Page 298
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-4 General Procedures of BEP2 Assy Disassembly (cont’d) 13.)When replace the BEP assy the option dongle should be moved from old BEP to new one. The FRU BEP doesn’t have the option dongle.
Page 299
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-4 General Procedures of BEP2 Assy Disassembly (cont’d) General Procedures of BT05 BEP3 Assy Disassembly Option Dongle Figure 8-101 BT05 BEP3 Option Dongle replacement NOTE: BEP2 and BEP3 FUR include applicable FRU CD SET, APPLICATION CD [2.06 Version], VGA...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3) 1.) Push the BEP Assy into The System and then Connect PCI Cable.
Page 301
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3) (cont’d) 6.) Connect LI, LO, MI, ETHERNET Connector to Sound/LAN Port on BEP.
Page 302
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3) (cont’d) 9.) Connect CON21 Connector to BEP. CON 21 Cable Figure 1-107 CON21 cable connection 10.)Connection COM1 Connector to Serial Port on BEP.
Page 303
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3) (cont’d) 11.)Connection Keyboard, Printer connector to USB Port on BEP. Keyboard USB Port...
Page 304
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-6 BEP Cable Connection With Console ( In case of change to BTO5 BEP3 FRU on BTO5 SYSTEM ) 1.) Push the BEP Assy into The System and then Connect PCI Cable.
Page 305
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-6 BEP Cable Connection With Console ( In case of change to BTO5 BEP3 FRU on BTO5 SYSTEM ) (cont’d) 5.) Connect the LCD Connector to BEP. LCD connector Figure 1-113 LCD cable connection 6.) Connect LI, LO, MI, ETHERNET Connector to Sound/LAN Port on BEP.
Page 306
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-6 BEP Cable Connection With Console ( In case of change to BTO5 BEP3 FRU on BTO5 SYSTEM ) (cont’d) 7.) Connect CON17 Connector to BEP. CON 17 cable Figure 1-115 Connecting CON17 cables 8.) Check the CON 17 cable is arranged without short before connecting.
Page 307
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-6 BEP Cable Connection With Console ( In case of change to BTO5 BEP3 FRU on BTO5 SYSTEM ) (cont’d) 10.)Connection COM1 Connector to Serial Port on BEP.
Page 308
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-6 BEP Cable Connection With Console ( In case of change to BTO5 BEP3 FRU on BTO5 SYSTEM ) (cont’d) 12.)Connect ATX Power to BEP. ATX Power cable...
Page 309
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-7 System Functional Check (In case of change BEP 1 into BEP2 or BEP3 ) Check the system after BEP is assembled on the console. 1.) Pull Up Circuit Breaker.
Page 310
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Advanced Menu P C I C o n f ig u r a t io n B o o t C o n f ig u r a t io n...
Page 311
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Power Menu P o w e r A C P I A C P I S u s p e n d S ta te [ S 1 S ta te ]...
Page 312
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Advanced menu screen PCI Configuration PCI Slot 1 IRQ Priority [ AUTO] PCI Slot 2 IRQ Priority [ AUTO] PCI Slot 3 IRQ Priority [ AUTO] PCI Slot 4 IRQ Priority...
Page 313
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Event Log Configuration Event Log [ Space available ] -- Not editable View Event Log -- [Enter] Displays the event log Event Logging [ Enable ] Video Configuration...
Page 314
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Security menu screen Supervisor password Installed User password Not Installed Set Supervisor password User Access Level [ No Access ] Set User password Chassis Intrusion [ Disable ]...
Page 315
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL b.) BTO5 BEP3 BIOS set up condition Main menu screen BIOS Version BF86510A.86.A.0069.P21 Process Type Intel Pentium 4 -- Not editable Process Speed 2.8Ghz -- Not editable System Bus Speed...
Page 316
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Peripheral Configuration Serial Port A [ AUTO ] Parallel Port [ AUTO ] Mode [ ECP ] Audio [ Enable ] LAN Device [ Enable ] ASF Support...
Page 317
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL USB Configuration USB Function [ Enable ] High - Speed USB [ Enable ] Legacy USB Support [ Enable ] USB 2.0 Legacy Support [ HI - Speed ]...
Page 318
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Power ACPI ACPI Suspend State [ S1 State ] Wake on Lan from S5 [ stay off ] After Power Failure [ Stay OFF ] Wake On PCI PME...
Page 319
3.) Check the Sound Driver is On board Sound (Sound Max). VIC assy On board sound Figure 8-129 Sound Driver 4.) Check the VIC Assy Driver is GE Medical SuperSonic Diagnostic Device Driver and GEMSK Multimedia Device 5.) Check The USB Device Driver. Chapter 8 Replacement Procedures...
Page 320
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-8 Check the Device Driver Configuration.(BEP2) (cont’d) 6.) L5 Keyboard Communicates with Jupiter Human Interface by USB Protocol. Figure 8-130 Jupiter Human Interface Driver 7.) Check The Network Device Driver.
Page 321
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) After System boot up. confirm the installed device information. Double click Go Device Manager Icon Device Manger Figure 8-131 Device Manager 1.) Disk driver check...
Page 322
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 2.) Display adapter check Secondary display : PCI graphic card Primary display : Onboard 865G chipset Figure 8-133 Display adapter Check 3.) DVD/CD-ROM driver check...
Page 323
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 4.) Human Interface Devices check Trackball device driver A & KEY device driver Figure 8-135 Human Interface Devices check 5.) IDE ATA/ATAPI controllers check...
Page 324
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 6.) Monitor check System Console Monitor Operation Panel LCD Panel monitor Figure 8-137 Monitor check 7.) Multifunction adapters check PCI Bridge device driver...
Page 325
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 8.) Network adapters check On board Network driver Figure 8-139 Network adapters check 9.) Ports (COM <P) check Serial Port device driver ECP ( Enhanced Parallel Port) device driver Figure 8-140 Ports (COM <P) check...
Page 326
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 10.)VIC driver check Video capture driver Sound dummy driver Figure 8-141 VIC driver check 11.)On board sound driver check AC97 controller device driver...
Page 327
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-9 Check the Device Driver Configuration.(BEP3) (cont’d) 12.)USB driver check (logical, Physical, Jupiter Human interface ) General USB Hub Controller driver 82801EB USB Universal Host controller Operation panel driver PCI to USB converter controller driver USB 2.0 controller...
Page 328
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) After System boot up. confirm the installed device information. Double click Go Device Manager Icon Device Manger Figure 8-144 Device Manager 1.) Disk driver check...
Page 329
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 2.) Display adapter check Primary display : Onboard 865G chipset Secondary display : USB to VGA module Figure 8-146 Display adapter check 3.) DVD/CD-ROM driver check...
Page 330
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 4.) Human Interface Devices check Trackball device driver A / KEY device driver Figure 8-148 Human Interface Devices check 5.) IDE ATA/ATAPI controllers check...
Page 331
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 6.) Monitor check System Console Monitor Operation Panel LCD Panel monitor Figure 8-150 Monitor check 7.) Multifunction adapters check PCI Bridge device driver...
Page 332
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 8.) Network adapters check On board Network Driver Figure 8-152 Network adapters check 9.) Ports (COM <P) check Serial Port device driver ECP ( Enhanced Parallel Port) device driver Figure 8-153 Ports (COM <P) check...
Page 333
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 10.)VIC driver check Video capture driver Sound dummy driver Figure 8-154 VIC driver check 11.)On board sound driver check AC97 controller device driver...
Page 334
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-10 Check the Device Driver Configuration.(BT05 BEP3) (cont’d) 12.)USB driver check (logical, Physical, Jupiter Human interface ) General USB Hub Controller driver PCI to USB converter controller driver Operation panel driver USB 2.0 controller...
Page 335
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-1-11 Application S/W Install. After Checking Device Driver and then Install application S/W by inserting CD into CD RW Driver. Figure 8-157 Installation Application Software 1.) Press the "Y" key to continue. Then the program requests confirmation - "Are you sure?" then Press the "Y"...
Page 336
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-2 VIC Assy Purpose: This is a description on how to remove and replace the VIC assy in the BEP Assy. 8-8-2-1 Tools • Common pillips screwdrivers 8-8-2-2 Needed Manpower •...
Page 337
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-2-4-2 Removal of VIC board 1.) Disconnect all of connect in VIC assy. Refer to Figure 8-161. Figure 8-161 Connectors on VIC assy 2.) 5 connectors inside BEP. 14 pins connector to chassi from VIC. 50pin half pitch flat cable connector.
Page 338
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-2-5 Mounting procedure 1.) Install the new parts in the reverse order of removal. 2.) Detail procedure of Harness routing about HDD LED cable and PWR cable. In case of disconnection HDD cable and PWR cable from Mother Board.
Page 339
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-2-5 Mounting procedure (cont’d) Case 2. BEP 2 Figure 8-166 BEP2 Case 3. BEP3 Figure 8-167 BEP3 Chapter 8 Replacement Procedures 8-143...
Page 340
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-2-5 Mounting procedure (cont’d) Case 4. BEP4 Figure 8-168 BEP4 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-48...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3 SMPS Assy, Hard Disk, CD R/W Purpose: This is a description on how to remove and replace the Parts in the BEP Assy. 8-8-3-1 Tools • Common pillips screwdrivers...
Page 342
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-4 Removal Procedure (cont’d) 5.) Unscrew 2 screws on theATX power bracket. 6.) Take out the ATX smps. Figure 8-171 Take out the ATX SMPS 7.) To remove Hard disk and CDRW Unplug the IDE cable for HDD or CDRW drive and power connector.
Page 343
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-4 Removal Procedure (cont’d) 10.)Move the CDRW guide braket to new FRU. Figure 8-173 CDRW Guide bracket 11.)Mode Jumper : Hard disk and CDRW should be set as Master drive. Confirm the mode setting jumper in hard disk and CDRW.
Page 344
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-6 Hard disk copy procedure This section show the patient data duplication procedure for hard disk replacement The purpose od this procedure is to copy the patient data in the used hard diks into the newly installed BEP or hard disk assy.
Page 345
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-6-2 Data copy to new HDD 1.) Go into maintenance mode. 2.) Plug the USB service dongle into usb port on the rear panel. There are 2 usb ports on the rear panel.
Page 346
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-6-3 File Copy (cont’d) 5.) Select the "dbeng7.exe" in the process list and click the "End Process" button. Process Process dbeng7.exe dbeng7.exe End Process End Process Figure 8-178 Task Manager with process 6.) click the "Yes"...
Page 347
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-6-3 File Copy (cont’d) 9.) There should be 8 logical disk totally. Each HDD have 4 partitions, "SYSTEM", "USER", "ARCHIVE", "SWAP". The used HDD which is set to slave is "H" to "K". And the new HDD in BEP which is set to master is "C"...
Page 348
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-8-3-6-4 Check the patient data 1.) Turn on the system 2.) If system boot up, select patient menu and confirm that the all of patient data are available.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-9 Power Block 8-9-1 AC Power Assy Purpose: This is a description on how to remove and replace the AC Power Assy. 8-9-1-1 Tools • Common pillips screwdrivers...
Page 350
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-1-4 Removal Procedure (cont’d) 3.) Unscrew eight (8) screws (11-18) to remove AC Power assy. Refer to Figure 8-184. Figure 8-184 Removing the AC Power Assy 4.) Remove the AC Power Assy.
Page 351
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-1-4 Removal Procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-50 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-2 LV Unit Purpose: This is a description on how to remove and replace the LV Unit. 8-9-2-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench • Long-nose gripper...
Page 353
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-2-4 Removal Procedure (cont’d) 4.) Unscrew four (4) screws (12-15) and disconnect twenty (20) connectors from the front side of the LV unit using long-nose gripper. Refer to Figure 8-186.
Page 354
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-3 JPC Assy Purpose: This is a description on how to remove and replace the JPC Assy. 8-9-3-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench 8-9-3-2 Needed Manpower •...
Page 355
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-3-4 Removal Procedure (cont’d) 3.) Disconnect all the connectors connected in the JPC Assy. Refer to Figure 8-188. (12) (10) (11) (13) (14) Figure 8-188 Disconnecting all connectors 4.) Remove the JPC Assy.
Page 356
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-3-5 Mounting procedure Case1 Replacing JPC with JPC Replacing JPC2 with JPC2 Install the new parts in the reverse order Case2 Replace JPC with JPC2 J4 connector ( SSR Control signal )
Page 357
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-4 Transformer Purpose: This is a description on how to remove and replace the Transformer. 8-9-4-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench 8-9-4-2 Needed Manpower •...
Page 358
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-4-4 Removal Procedure (cont’d) 6.) Unscrew four (4) ground screws (5-8). Refer to Figure 8-191. (5) (6) (7) (8) Figure 8-191 Unscrew 4 screws 7.) Unscrew one nut in the Center top of the Transformer. Refer to Figure 8-192.
Page 359
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-5 AC Power Box Purpose: This is a description on how to remove and replace the AC Power Box. 8-9-5-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench...
Page 360
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-5-4 Removal Procedure (cont’d) 3.) Disconnect one (1) connector (J1) from the JPC Assy. Refer to Figure 8-194. Figure 8-194 Disconnect 1 connector 4.) Unscrew two (2) ground screws (7-8). Refer to Figure 8-195.
Page 361
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-9-6 Fuse Set Purpose: This is a description on how to remove and replace the Fuse Set. 8-9-6-1 Tools • Common pillips screwdrivers • Allen/Unbraco wrench 8-9-6-2 Needed Manpower •...
Page 362
Section 8-10 Software Loading Procedure 8-10-1 General This describes a full system software loading (Base Software Load Image + LOGIQ 5 Application) procedure. Use this instruction when: - Performing the HDD with new one (Base Software Load Image + Application)
Page 363
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-4 Software Loading Steering Guide Figure 8-197 Flow Chart - Installation of Base and Application CD Chapter 8 Replacement Procedures 8-167...
Page 364
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5 Before Starting Software Loading 8-10-5-1 Check Software Version and BEP Type 1.) Turn ON the system. 2.) Touch the Utility button on the Touch Panel Figure 8-198 Touch the Utility NOTE: Operator Login Window appears.
Page 365
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-1 Check Software Version and BEP Type (cont’d) Image Part Number Software version & Part number Figure 8-200 Check BEP Type and Software Version Chapter 8 Replacement Procedures...
Page 366
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-2 Check BEP Type with BIOS Boot Logo Image NOTE: On booting the system, BIOS Boot Logo Image shows which BEP type is used. BEP Type Boot Logo Image...
Page 367
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-3 Saving Connectivity Before starting upgrade, write down the Connectivity Setting for back-up. 1.) Touch Utility > Connectivity on the touch panel and click the Tcpip tab. 2.) Write down the following parameters:...
Page 368
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-4 Operator Login Password Back-up The Operator Login Password will be deleted. So, write down the password if necessary. Figure 8-202 Operator Login Parameters Descriptions Operator Login Password...
Page 369
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-5 Image Management Guide (cont’d) 5.) The SAVE AS menu appears. Figure 8-203 Save As Menu 6.) Specify Compression and Save As Type and press Save. The image is saved to the CD-ROM.
Page 370
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-5 Image Management Guide (cont’d) Storing Images with More Resolution To store images with more resolution than is available with the JPEG selection, select Save As and select AVI as the Save As Type. You can save single images as .avi files.
Page 371
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-5 Image Management Guide (cont’d) 3.) The [From] combo box is not active. It displays Local Archive. The [To] combo box is active. Select the type of removable media MOD or CD Rom. Then please wait until the patient list is visible.
Page 372
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-5 Image Management Guide (cont’d) 6.) The progress bar appears as the copy is taking place. Once transfer completed, the exported patient list will be shown as Figure 8-210.
Page 373
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-10-5-5 Image Management Guide (cont’d) 4.) The [TO] combo box is not active. It displays Local Archive. The [FROM] combo box is active. Select the type of removable media MOD or CD Rom. Then please wait until the patient list is visible.
Page 374
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-11 Installing Base Software Load Image 8-11-1 Installing Base Software Load Image CD NOTE: While we do not believe it is necessary, It would not hurt to disconnect the system from the network and remove all transducers.
Page 375
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1 Installing Base Software Load Image CD (cont’d) 5.) Using arrow keys, Select Advanced Tab and select Chipset Configuration and then press the Enter key. Verify that ISA Enable Bit is Disabled. If it is enabled, using arrow keys, select ISA...
Page 376
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1 Installing Base Software Load Image CD (cont’d) Figure 8-217 First Boot Device to CD ROM for BEP2 NOTE: If the system is BEP3, select 1st Boot Device to 4M-HL-DT-ST-GCE-8526B Figure 8-218 1st Boot Device to 4M-HL-DT-ST-GCE-8526B 9.) Press the ESC key to return to the Setup Utility screen.
Page 377
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1 Installing Base Software Load Image CD (cont’d) 11.)Insert the disk labeled “Base System Software Load Image Disk 1 of 2” into the CDROM drive. 12.)After restarting the system, when the following screen appears, press the Enter key to continue, to abort remove the CD-ROM from the drive and press CTRL-C.
Page 378
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1 Installing Base Software Load Image CD (cont’d) 13.)The following screen appears. Figure 8-220 Selection Screen 14.)If you choose ‘A’, you will be asked to press ‘Y’ to confirm your selection.
Page 379
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1 Installing Base Software Load Image CD (cont’d) 15.). Approx. ten (10) minutes later, the gray colored screen appears. Eject the “ Base Software Load Image CD Disk 1 of 2 ” from the drive and Insert the “ Base Software Load Image CD Disk 2 of 2”.
Page 380
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks 1.) Approx. two minutes later, the windows XP desktop appears. Then one minutes later, the following screen appears. Clisck on Yes using right button of the trackball to restart the system...
Page 381
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) c.) Right-Click on the “My Computer” icon on the Windows desktop. Select “Manage” you should see a screen like below. Then left click on the driver assigned currently to G:, right click on...
Page 382
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) Select “Computer Management” then you should see a screen like below. Then left click on the device assigned currently to G:, right click on Change Drive Letter and Path...
Page 383
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) e.) Select the drive other than G: (I: is selected in the window shown below for example.) Then click on OK Figure 8-229 Edit Drive Letter f.) Right click on Yes...
Page 384
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) h.) Right click on Edit Figure 8-232 Change Drive Letter i.) Select G: then rignt click on OK Figure 8-233 Edit Drive Letter j.) Right click on Yes...
Page 385
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) k.) Left click on the device which changed the drive number at the step c.). Right click on Change Drive Letter and Path... Figure 8-235 Change Drive Letter and Path l.) Right click on Edit...
Page 386
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-1 System Checks (cont’d) n.)Right click on Yes Figure 8-238 Confirm Window o.) Close all windows 3.) Driver Checks: a.) On the window XP desktop, left click on My Computer then right click on Properties b.) The System Propertiers window appears.
Page 387
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-2 USB Printer Checks (USB Printer User Only) Without any newly additional user invention, LOGIQ5 will recognize the below 9 kinds of USB printer Printer Name Comment HP 990 cxi, HP 6122 (BEP2 Only),...
Page 388
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-2 USB Printer Checks (USB Printer User Only) (cont’d) Serial Right Left 220-240V 500VA Max Including front printer panel Figure 8-240 USB Port - Rear Panel Connector Figure 8-241 USB Port - OP Printer Bracket Back Panel 2.) Make sure that printer properties are proper.
Page 389
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-3 DMC Checks (DMC User Only) 1.) Check DMC Cable (USB to Serial Bridge). NOTE: DMC Cable is supplied only as an option. 2.) Plug in the Cable as shown on the right side of the USB port. Refer to Figure 8-242.
Page 390
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-1-3 DMC Checks (DMC User Only) (cont’d) 10.)Right click on port settings. 11.)Click on advanced .. button. 12.)Specify COM port number is COM4. as shown in refer to Figure 8-244.
Page 391
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-2 Installig Application Software(R4.x.x) 8-11-2-1 Installig Application Software(R4.x.x) 1.) Place the “Application Software” CDROM into the CDROM drive. 2.) Press the Start button on the tool bar at the bottom of the LCD screen and select “Run…”.
Page 392
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-2-1 Installig Application Software(R4.x.x) (cont’d) 8.) After the installation is completed the system will power off automatically. NOTE: Do not operate with using mouse or keyboard during installation procedures.
Page 393
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-3 Standard Confirmation & Functional Checks Configure the system settings and perform functional checks after installation of software. Items for configuration and functional checks include: Confirmation of the software version...
Page 394
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-3-2 Setting Computer Name NOTE: If the "Save" is NOT performed, you can NOT enter Patient screen and NOT perform Export/Import. 1.) Touch Utility-> Connectivity 2.) Operator Login Window appears. Select ADM(Administrator) for operator field.Click OK button.
Page 395
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-11-3-3 Functional Checks for Service Platform (Diagnostic Program) NOTE: After performing diagnostics, shut down and restart to make the system working properly. 1.) Make sure that the wrench icon is shown at the bottom of the scan screen. Click on the wrench icon to activate.
Page 396
Service Platform has failed. Reload the application software or Base System Software (OS) + application software. Contact a Technical Support for details. 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform.
Page 397
5.) Shut down the scanner, then restart the scanner. 6.) Verify that the VCR can be properly operated (remote, play...). Be sure to connect the converter and cablebefore booting up the system. Otherwise the LOGIQ 5 NOTICE system does not recognize the peripheral devices...
Page 398
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 8-12 Full Backup / Restore Procedure 8-12-1 General This describes a Full Backup (Database, Image, User preset, Service Data) procedure. Use this instruction when you need to save the data : •...
Page 399
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-12-3 Full Backup Procedure (cont’d) 7.) Press “Backup”. Then the below window pops up.. Figure 8-258 Emergency Disk Making Wizard 8.) Insert a USB HDD. NOTE: It might take about 30 seconds for the system to recognize the USB HDD.
Page 400
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-12-3 Full Backup Procedure (cont’d) 11.)If USB HDD with enough space is not already plugged in, plug in one.Then press Next Figure 8-260 Storage Information 12.)Press next to continue. Then the below windows appear.
Page 401
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-12-3 Full Backup Procedure (cont’d) Figure 8-262 Emergency Disk Making in Progress 13.)Press Finish to complete. 14.)Check if four files are created in the USB HDD. DB_IMG.zip Service_SQL.zip SYSTEMNAME.DAT...
Page 402
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL NOTE: Double click the zip archive files and check if the archived files are the same as the original files. Archive File Original File Path DB_IMG.zip E:\Idunn\GEMS_IMG\ C:\SYBASE\INI\ Service_SQL.zip...
Page 403
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-12-4 Restore procedure 1.) From Figure 8-257, press “Restore” button. Rest of the procedure is the same with “Backup” procedure. The system should be rebooted after finishing of Restore procedures.
Page 404
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 8-208 Section 8-12 - Full Backup / Restore Procedure...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ™ 5. Table 9-1 Contents in Chapter 9...
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-3 Renewal Parts Lists 9-3-1 Equipment Models Covered in this Chapter Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 2332322 100V, NTSC OPERATOR CONSOLE ASSY...
Page 407
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-4 OPERATOR CONSOLE ASSY COLOR MONITOR (FRU 1XX) See Illustration KEYBOARD BLOCK (FRU 2XX) See Illustration COVERS (FRU 3XX) See Illustration CABLES, Peripherals, OPTIONS (FRU 8XX)
Page 408
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-5 MONITOR Figure 9-2 Monitor 9 - 4 Section 9-3 - Renewal Parts Lists...
Page 409
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-5 MONITOR (cont’d) Table 9-2 MONITOR Part Item Part Name Number Description MONITOR CRT ASSY 2299998-3 CRT W/O COVERS MONITOR CABLE ASSY 2299964 CABLE PLASTIC COVER SET WITH...
Page 410
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-6 KEYBOARD BLOCK NOTE : FRU 212 ~223 ARE NOT ILLUSTRATED IN THIS FIGURE Figure 9-3 KEYBOARD BLOCK 9 - 6 Section 9-3 - Renewal Parts Lists...
Page 411
USER DEFINED KEY LABEL GEL HOLDER 2384742 ITV BLOCK 2384743 FREEZE KEY 2384744 KEYBOARD MAIN BOARD 2384745 KEYB’D MAIN PCB BOARD PROBE HOLDER BRACKET 2384746 HUB ASSY 2385577 FOR LOGIQ 5 ONLY Chapter 9 - Renewal Parts 9 - 7...
Page 412
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-7 COVERS Figure 9-4 COVERS 9 - 8 Section 9-3 - Renewal Parts Lists...
Page 413
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-7 COVERS (cont’d) Table 9-4 COVERS Part Item Part Name Number Description INCLUDE LEFT BOTTOM LEFT COVER 2299995 COVER & SIDE MOLDING INCLUDE RIGHT BOTTOM RIGHT COVER 2300004 COVER &...
Page 414
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-8 BODY BLOCK Figure 9-5 BODY BLOCK 9 - 10 Section 9-3 - Renewal Parts Lists...
Page 415
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-8 BODY BLOCK (cont’d) Table 9-5 BODY BLOCK Part Item Part Name Number Description REAR HANDLE 2300016 PLASTIC BAR REAR PANEL ASSY 2300019 110V For US JPN...
Page 416
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-9 PCB BOARDS Figure 9-6 PCB BOARDS 9 - 12 Section 9-3 - Renewal Parts Lists...
Page 417
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-9 PCB BOARDS (cont’d) Table 9-6 PCB BOARDS Part Item Part Name Number Description PROBE INTERFACE RELAY RLY ASSY 2299950 ASSY LMT ASSY 2324741 LIMITER ASSY...
Page 418
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-10 PC BLOCK Figure 9-8 PC BLOCK 9 - 14 Section 9-3 - Renewal Parts Lists...
Page 419
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-10 PC BLOCK (cont’d) Table 9-7 PC BLOCK Part Item Part Name Number Description BEP2 ASSY WITH CABLE AND BEP2 ASSY WITH CABLE AND CD 2379029...
Page 420
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-11 POWER BLOCK Figure 9-9 POWER BLOCK 9 - 16 Section 9-3 - Renewal Parts Lists...
Page 421
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-11 POWER BLOCK (cont’d) Table 9-8 POWER BLOCK Part Item Part Name Number Description 2299962 AC POWER FOR JPN USA AC POWER ASSY 2337036 AC POWER FOR EUROPE...
Page 422
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-12 Options, Peripherals and Cables Table 9-9 CABLES Item Part Name Part Number Description OP CABLE ASSY 2331742 CABLE SET PCI CABLE 2363523 CABLE SET PC CABLE ASSY...
Page 423
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL Section 9-12 Options, Peripherals and Cables (cont’d) Table 9-12 PERIPHERALS Item Part Name Part Number Description SONY UP-895MDW Local SONY UP-D895MD Local MITSUBISHI P91W Local MITSUBISHI P91E...
Page 424
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL 9 - 20 Section 9-3 - Renewal Parts Lists...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ™ 5 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
Page 427
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Monthly Quarterly QA Program Notes •* Clean Probes * or before each use • Clean Probe Holders more frequently depending on •...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance Tool Part Number...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4 System Checks...
Page 430
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-5...
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-5-4-2 Air Filter Cleaning Table 10-9 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter Refer to Chapter 8 for air filter location and removal instructions.
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 10-6 Using a Phantom See the Basic User Manual “Customer Maintenance” for information on using a phantom and quality assurance tests. Section 10-7 Electrical Safety Tests 10-7-1...
Page 435
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
Page 436
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
Page 437
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 10-8Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-1 Chassis Leakage Current Test 10-8-1-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
Page 439
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 7.) Follow the test conditions described for respective test points shown in Table 10-15. Table 10-15 Chassis Leakage Current Test Condition TEST CONDITION Mounting screw for probe receptacle...
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-2 Isolated Patient Lead (Source) Leakage–Lead to Ground 10-8-2-1 Definition This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a patient who is connected to the monitoring equipment and is grounded by touching some other grounded surface.
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Table 10-17 Testing Power Conditions ECG Power Meter’s Polarity Switch Meter’s Neutral Switch REVERSE OPEN 10-8-3 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-LEAD position.
Page 442
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-4-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
Page 443
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-5 Probe Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 444
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-5-3 No Meter Probe Adapter Procedure Figure 10-6 Check Without Probe Adapter Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ 5 unit OFF.
Page 445
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL 10-8-5-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-13 Table 10- 14. Record all data on the PM Inspection Certificate.
Page 446
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL Section 10-9 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 447
Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ Chapter 10 Care &...
Page 448
GE M EDICAL YSTEMS 2300000, R LOGIQ™5 S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Caster...
Page 449
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL INDEX Abbreviations, 9-1 DC Power, 5-13 AC Box Assy, 8-50 DDBF, 6-10 AC Power, 5-12 DICOM Network Function, 2-9 AC Transformer, 8-56 Dimensions, 3-7 Acoustic Noise Output, 3-8...
Page 450
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL INDEX MOTHER Board, 8-46 Handle, 8-10 Hazard Icons, 1-4 HDD and Battery, 9-12 Human Safety, 1-7 OP Panel and Keys, 9-10 HV Power, 5-14 Operation Panel Assy, 8-28...
Page 451
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL INDEX REAR CONN Assy, 8-45 Ultrasound Room Layout, 2-8 Rear Cover, 8-4 Unpacking the Equipment, 3-3 Rear Panel, 5-7 Rear Panel Connector, 3-17 Recording Devices, 3-11,, 9-8...
Page 452
GE M EDICAL YSTEMS 2300000, R LOGIQ™ 5 S IRECTION EVISION ERVICE ANUAL INDEX A - 4 Index...