7.5 Power Cable

Descriptions:
Specifications
Service Amps:
Plug:
5 Wire:
P-116 MSHA, 200 A, a #1/0 4 conductor type W, 600V – 2000V, 90° C, 45'-0"
Cord:
An optional power plug to mate with the Crouse Hinds receptacle is Crouse Hinds # AP20457.
NOTE: The Russellstoll plug IS NOT compatible with the optional Crouse Hinds facility receptacle
The Power Cable could present a trip hazard that could result in personal
injury. Care should be taken to ensure that the cable is routed properly to
minimize its potential as a trip hazard.
The information contained in this booklet is the property of AMST. The contents are confidential. They may not be used,
either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be
reproduced or disclosed without written permission of Advanced Mobility Specialty Vehicles.
150 A
Russellstoll; DS 2504 MP000/DF2032, 600V AC, 200 A
5 pole
and vice versa.

Figure 39: Power Cable