Parameterization
7.4
Description of parameters
7.4.1
Menu: Easy Setup
Menu / Parameter
Easy Setup
Language
Q (Flowrate) Unit
Qmax
Totalizer/Pulse Unit
Operation
Pulses per Unit
Fullscale Frequency
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
90
Value range
Deutsch, English, Français,
Español, Italiano, Dansk,
Svenska, Polski, Russki,
Zhongweng, Turkce
l/s; l/min; l/h; ml/s; ml/min;
m3/s; m3/min; m3/h; m3/d;
Ml/d; ft3/s; ft3/min; ft3/h; ft3/d;
ugal/s; ugal/min; ugal/h;
ugal/d; Mugal/d; igal/s;
igal/min; igal/h; igal/d; bls/s;
bls/min; bls/h; bls/d; hl/h; g/s;
g/min; g/h; kg/s; kg/min; kg/h;
kg/d; t/min; t/h; t/d; lb/s;
lb/min; lb/h; lb/d; custom/s
Min. flow range: 0 ... 0.2 m/s
(0 ... 0.02 x Q
DN)
max
Max. flow range: 0 ... 20 m/s
(0 ...2 x Q
DN)
max
m3; l; ml; ft3; hl; g; kg; t; lb;
igal; ugal; bls; Ml; Mugal;
custom
Pulse Mode, Fullscale
Frequency
-
0 ... 5250 Hz
FEX300, FEX500
Description
"Easy Setup" Menu
Select the menu language.
Select the unit for the flow indicator.
Default setting: l/min
Select the flow range for forward and reverse flow.
Default setting: 1 x Q
max
Select the unit for the flowmeters.
Default setting: l
Select the operating mode for the digital output.
There are two operating modes available:
•
„Pulse Mode": In pulse mode, pulses per unit are
output (e.g., 1 pulse per m
•
„Fullscale Frequency": In frequency mode, a
frequency proportional to the flowrate is output.
The maximum frequency corresponding to the flow
range end value is configurable (max. 5.25 kHz).
Default setting: „Pulse Mode"
Display of the pulses per unit output by the digital
output. The max. possible number of pulses is 5250
per second.
Set the frequency for the flow range end value in
Fullscale Frequency operating mode.
DN.
3
).