Summary of Contents for Siemens SITRANS F M MAG 5000
Page 1
SITRANS F Electromagnetic flowmeters SITRANS F M MAG 5000/6000 Operating Instructions Edition 12/2013 Answers for industry.
Page 3
___________________ SITRANS F M MAG 5000/6000 Introduction ___________________ Safety notes ___________________ Description SITRANS F ___________________ Installing/Mounting Flowmeters ___________________ SITRANS F M MAG 5000/6000 Connecting ___________________ Commissioning Operating Instructions ___________________ Functions ___________________ Alarm, error, and system messages ___________________ Service and maintenance...
Page 4
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
It is the responsibility of the customer that the instructions and directions provided in the manual are read, understood and followed by the relevant personnel before installing the device. Items supplied SITRANS F M MAG 5000/6000 transmitter Calibration report SITRANS F literature CD Quick start guide Inspection 1.
Product information on the Internet The Operating Instructions are available on the CD-ROM shipped with the device, and on the Internet on the Siemens homepage, where further information on the range of SITRANS F flowmeters may also be found: Product information on the internet (http://www.siemens.com/flow)
Page 12
Introduction 1.4 Further Information SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
50% relative humidity from 40°C (104°F) ● Main supply voltage fluctuations up to ±10% of the nominal voltage (see technical specifications) ● Overvoltage category II ● Pollution degree 2 SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
● Electrical connections are in accordance with Elex V (VO in explosion hazardous areas) and EN60079-14 (Installing Electrical Systems in Explosion Hazardous Areas). ● The protective cover over the power supply is properly installed. For intrinsically safe circuits the connection area can be opened. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 15
● EN50281-1-2 is considered for installation in areas with combustible dust. ● When protective earth (PE) is connected, no potential difference between the protective earth (PE) and the potential equalization (PA) can exist, even during a fault condition. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 16
Safety notes 2.2 Installation in hazardous location SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Description System components A SITRANS F M MAG 5000/6000 flowmeter system includes: ● Transmitter (type SITRANS F M MAG 5000/6000) ● Sensor (types: SITRANS F MAG 1100/1100F/3100/3100 P/5100 W) ● Communication module (optional) (types: HART, PROFIBUS PA/DP, MODBUS RTU RS 485, Foundation Fieldbus H1, Devicenet) ●...
Output module The output module converts flow data to analog, digital and relay outputs. The outputs are galvanically isolated and can be individually set to suit a particular application. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
SV version (MAG 6000 only) This version is identical to the standard MAG 6000 transmitters except for the following additional functions: ● Zero point adjustment ● Adjustable excitation frequency up to 44 Hz SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 20
Description 3.5 MAG 5000/MAG 6000 versions SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
No further assembling is necessary. The transmitter can be installed either compact on the sensor or remote. Figure 4-1 Compact installation Figure 4-2 Remote installation SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Reading and operating the flowmeter is possible under almost any installation conditions because the display can be oriented in relation to the sensor. To ensure optimum flow measurement, attention should be paid to the following: Vibrations Figure 4-3 Avoid strong vibrations SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 23
Standard, blind and SV versions Figure 4-5 CT version Remote installation Cable length and type (as described in Cable requirements (Page 72)) must be used. For installation conditions for sensors, see respective sensor operating instructions. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
System will not register flow if black plugs are not connected to connection board. 7. Fit supply and output cables through cable glands and connect to connection plate as shown in Electrical connection (Page 36). SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 25
11. Transmitter is ready to be powered up. NOTICE Exposing transmitter to direct sunlight may increase operating temperature above its specified limit, and decrease display visibility. A sunshield is available as accessory. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
1. Remove terminal box lid. 2. Remove SENSORPROM unit from sensor terminal box and mount it in terminal box of ® wall mounting unit. 3. Fit M20 or ½" NPT cable glands for cables. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 27
Unscreened cable ends must be kept as short as possible. Electrode cable and coil cable must be kept separate to prevent interference. 5. Tighten cable glands well to obtain optimum sealing. WARNING Mount terminal box lid before power up. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 28
Figure 4-6 Wall mounting Figure 4-7 Pipe mounting - vertical Figure 4-8 Pipe mounting - horizontal 2. Ensure that correct SENSORPROM memory unit is mounted in wall/pipe mounting unit. ® SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 29
6. Fit coil, electrode, supply and output cables through cable glands and connect to connection plate as shown in Electrical connection (Page 36). 7. Fix connection plate with the two diagonally opposite screws. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
MAG 6000 CT is installed like a Standard MAG 6000 except for the final sealing. Calibration sealing has been carried out at calibration. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
When key is mounted, access to all menu items is gained. When key is removed, primary settings are blocked in accordance with requirements in authorisation. 4.5.2 Seal device Seal transmitter CAUTION Seal transmitter to prevent unauthorized access. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 32
1. Seal connection plate to prevent access to SENSORPROM memory unit as shown ® below. 1 indicates sealing locations. 2. Drill through marked drilling holes in terminal box and transmitter/lid. Seal transmitter externally as shown below. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
● Compact or remote with max. 10 m (33 ft.) cable Other restrictions may apply (see certificate 22.76/10.02) **insert installation conditions as pasted Turning transmitter/keypad It is possible to alter the standard assembly, e.g. to turn transmitter or keypad. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 34
Terminal box can be rotated ±90° in order to optimize viewing angle of transmitter display/keypad 1. Unscrew the four screws in bottom of terminal box. 2. Turn terminal box to required position. 3. Retighten screws firmly. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 35
Installing/Mounting 4.6 Turning transmitter/keypad Keypad 1. Remove outer frame using a screwdriver. 2. Loosen the four screws retaining keypad. 3. Withdraw keypad and turn it to required orientation. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 36
Installing/Mounting 4.6 Turning transmitter/keypad 4. Tighten the four screws until a mechanical stop is felt in order to obtain IP67 enclosure. 5. Snaplock outer frame onto keypad (click). SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
For field wiring installation: Ensure that the National Installation Code of the country in which the flowmeters are installed is met. Note National installation code Observe country specific installation directives for field wiring. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Connecting 5.1 Electrical connection Electrical connection Figure 5-1 Wiring diagram SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 39
Fasten cables/wires in a way that they cannot get into contact with each other, not even in case of an error. Keep wire ends as short as possible. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
91-97. For more information Refer to the relevant BUS communication Quick Start or Operating Instructions available at the SITRANS F literature CD or on the internet, at : www.siemens.com/flowdocumentation (www.siemens.com/flowdocumentation). SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
All changed data will be stored in SENSORPROM memory unit. ® 4. Remove standard transmitter and remount Blind transmitter. 5. Fasten screws holding transmitter. New settings stored in SENSORPROM memory unit will be uploaded in blind transmitter. ® SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Lock field Figure 6-1 Local User Interface Mode field symbols Communication mode Language mode Sensor characteristics Service mode Basic settings Reset mode Operator menu Output Operator-active Product identity External input Operator-inactive SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
2 seconds. The setup menu operates in two modes: ● View mode ● Setup mode View mode is a read-only mode. The pre-selected settings can only be scanned. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Factory setting of password can be re-established as follows: 1. Switch off power supply. 2. While pressing top up key - switch on power supply. 3. Release top up key after 10 sec. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
6. Press lock/unlock key to unlock settings. 7. Use select key and change key to change settings. 8. Press lock/unlock key to confirm new settings. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 46
9. Repeat steps 5-8 to change other settings. 10. Press top up key two times to exit setup mode. Decimal point can be positioned and units set individually for flow rate in totalizer 1 and totalizer 2. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
2. Press lock/unlock key to unlock setting. 3. Use select key to move cursor to upper line. 4. Use change key to select requested reading. 5. Press lock/unlock key to confirm selection. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
4. Press lock/unlock key to unlock language. 5. Use change key to select desired language. 6. Press lock/unlock key to confirm new language. 7. Press top up key two times to exit setup mode. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Alarm level Output 3.5 mA High 22.6 mA For setting of error level, see Error level menu diagram. If current output is not used, it must be set to "Off". SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 50
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT. Note Cleaning If a cleaning unit is installed together with transmitter, relay output must always be used to operate this unit. It cannot be used for other purposes. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
● Calibration factor. ● Correction factor. ● Excitation. See also sensor characteristics menu diagram. Reset mode The reset mode is used for resetting totalizers/counters or for restoring MAG 5000/6000 to its factory settings. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 52
4. Press lock/unlock key to enter reset menu. 5. Press forward key to reach zero adjust menu. 6. Press lock/unlock key to enter the menu. 7. Use change key to select "manual". SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
● Totalizer 1 cannot be reset. ● Totalizer 2 can be reset if the flow velocity in the meter pipe is <0.25 m/s. When the totalizer is reset, the pulse output register will also be reset. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
When this is done, the decrease in measuring accuracy can be kept below 1% o.r. A too high frequency for the sensor used will cause a coil current alarm indication. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
The acceptance level for "error pending" can be individually configured to a particular application. The acceptance level is set in the basic settings menu (Page 43). SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 56
An output can also be selected to activate on a single error number. The alarm field, error output and error pending always operate together. Operator menu Error pending and status log are as default enabled ( ) in the operator menu. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
P41 - Parameter range Switch off and on A parameter is out of range. The parameter could Active Active not be replaced by its default value. The error will disappear at the next power-on SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 58
It is not possible to read from the SENSORPROM Active Active ® unit anymore F70 - Coil current Check cables Coil excitation has failed Active Active F71 - Internal error Replace transmitter Internal convertion error in ASIC Active Active SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
It is a highly advanced instrument, which carries out the complex verification of the entire flowmeter system according to unique SIEMENS patented principles. The verification test is automated and the instrument easy to use, so no human error or influence will affect the verification.
Sensor check list Check list for sensors are included in the respective sensor operating instructions. Technical support NOTICE Repair and service must be carried out by approved Siemens Flow Instruments personnel only. Note Siemens Flow Instrument defines sensors as non-repairable products.
● Information about field service, repairs, spare parts and lots more under "Services." Additional Support Please contact your local Siemens representative and offices if you have additional questions about the device Find your contact partner at: Local contact person (http://www.automation.siemens.com/partner)
You can find the forms on the Internet and on the CD delivered with the device. Recalibration Siemens A/S Flow Instruments offers to recalibrate the sensor. The following calibrations are offered as standard: ● Standard matched pair calibration...
Replace SENSORPROM unit or ® ® reset SENSORPROM unit with MAG ® CT transmitter Defective SENSORPROM Replace SENSORPROM unit ® ® unit Wrong type of Replace SENSORPROM unit ® SENSORPROM unit ® SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 64
Change pulse width Measuring approx. Missing one electrode Check cables connection Loss of totalizer data Initializing error Reset totalizer manually ##### Signs in Totalizer roll over Reset totalizer or increase totalizer display unit SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
0 ... 9.9% of maximum flow Galvanic isolation All inputs and outputs are galvanically isolated Max. measuring error MAG 5000 0.4% ± 1 mm/s (for v > 0.1 m/s) (incl. sensor and zero point) SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 66
Prepared for client mounted add-on modules: HART, MODBUS RTU/RS485, FOUNDATION Fieldbus H1, DeviceNet, PROFIBUS PA, PROFIBUS DP as add-on modules MAG 5000 CT / MAG 6000 CT No communication modules approved SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 67
Technical data 11.1 Technical specifications Transmitter IP67/NEMA 4X/6 compact polyamide Weight: MAG 5000/6000: 0.75 kg (1.65 lbs) Transmitter IP67/NEMA 4X/6 wall-mounted polyamide Weight(transmitter and wall mounting bracket): 1.65 kg (3.64 lbs) SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
For accuracy reference conditions, please see below. Figure 11-1 MAG 5000 with MAG 1100, MAG 1100 F, MAG 5100 W, MAG 3100 and MAG 3100 P and MAG 6000 with MAG 1100 (PFA), MAG 1100 F (PFA) SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 69
10 x DN (DN ≤ 1200/48") 5 x DN (DN > 1200/48") Outlet section 5 x DN (DN ≤ 1200/48") 3 x DN (DN > 1200/48") Flow conditions Developed flow profile Reference conditions for sensor calibration SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
≥ 10 μS/cm Additions in the event of deviations from reference conditions 11.3 Output characteristics Current output Output characteristics Bidirectional mode Unidirectional mode 0...20 mA 4...20 mA Frequency output Bidirectional mode Unidirectional mode SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 71
Active Error relay output Bidirectional mode Unidirectional mode No error Error Limit switch (can be used as direction switch) 1 set point 2 set points Passive Digital output Passive Digital output SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 72
11.3 Output characteristics 1 set point 2 set points Note Active digital output is not avaiable with MAG 6000 I. Batch on digital output Unidirectional mode (forward flow only) Totalizer lap SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
For Ex zone 1 installations only certified cable glands with protection type "e" can be used for the power supply and the coil cable. The cable glands must be approved for the actual temperature and cable dimension. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
In order to ensure that the ordering data you are using is not outdated, the latest ordering data is always available on the Internet: Catalog process instrumentation (http://www.siemens.com/processinstrumentation/catalogs) 12.2 Accessories Description Wall mounting unit Display protection lid Communication modules for MAG 6000 SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Spare parts/Accessories 12.3 Spare parts 12.3 Spare parts Description Connection plate SENSORPROM memory unit ® Display unit Communication modules for MAG 6000 12.4 Sun shield Description Sun shield SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
The menu diagrams shown on the following pages apply to MAG 5000/6000 as well as MAG 6000 I. Not available in MAG 5000 Add-on module Factory-set password: 1000 Not available when batch Only available when batch SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
The flow rate unit can be "?/?" or show customer unit texts as for example "AcrFt/y" (Acre feet/year). When batch is selected on digital output or relay, Totalizer 2 is not shown because it is controlled by the batch function. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Do not combine differentiation and low alarm level together with passive output. The output will try to pull down the level to 1.3 mA at fatal errors which is not possible for passive output. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Menu diagrams A.4 Digital output - pulse Digital output - pulse Digital output - frequency Error level SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
If digital output is set up for pulse, frequency or batch, then the relay output will be turned off if it has been set up for batch. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
(The display is locked during this time). The time cycle can be set at 1 to 240 hours. If the cycle is set at for example three hours, the transmitter will be active every three hours. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
In this case the power must be switched off/on to suppress the P40 error on the output. The frequency can be set to 44 Hz in the MAG 6000 SV transmitter only. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
9 is the best signal measured. If digital output is set to pulse (standard). If digital output is set to frequency. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
2 and flow velocity is chosen in line 3, the text "Flow velocity " is shown in line2 and the measured flow velocity is shown in line 3. SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Burst mode is not available with HART communication A.18 Cleaning Note Relay outputs If cleaning unit is installed, relay outputs must always be used to operate cleaning. Relay outputs cannot be used for other purposes SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Menu diagrams A.20 Change password Note Sealing Menus marked with gray are locked when transmitter is sealed. A.20 Change password SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Current output Off for On/off, Unidirectional/bidirectional, 0...20 Output MAG5000/ mA/4...20 mA/4...20 mA + Alarm settings (Page 47) 6000 On for MAG6000I Namur High/Low Alarm level Yes/No Alarm diff. 0.1...30 s Time constant SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 94
0.1 s 0.1 ... 30 s Time constant External input External input Batch, reset totalizer, freeze output, forced output, External input (Page 80) Start Start, hold/continue, stop, Qmax 2 Batch SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
4778.363 191134.4 m3/h 280 (114) 65000 5541.77 221670.7 m3/h 3000 (120) 70000 6361.726 254469 m3/h * The min. and max. amount values show mathematical values and do not indicate measurement accuracy SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 97
US GPM 1200 (48) 66044 4481.574 179262.9 4481.574 179262.9 US GPM 1400 (54) 110072 6099.92 243996.7 US GPM 1500 (60) 1320867 7002.459 280098.3 US GPM 1600 (66) 154101 7967.242 318689.6 US GPM SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
171.5728 m 125 (5) 9.265244 ml 242.8828 m 10.22654 ml 268.0825 m 150 (6) 14.57896 ml 382.1785 m 14.72622 ml 386.0389 m 200 (8) 23.09071 ml 605.309 m 26.17994 ml 686.2913 m SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
5.890487 l 154415.5 m * The min. and max. amount values show mathematical values and do not indicate measurement accuracy. ** For CT devices the totalizer 1 unit is in m SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 101
US G max. US G min. US G max. 2000 (78) 0.6916009 18129900 2200 (90) 0.836837 21937170 2400 (96) 0.995906 26107050 2600 (102) 1.168806 30639530 2800 (114) 1.355538 35534600 3000 (120) 1.556102 40792270 SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Approvals/Certificates All certificates are posted on the Internet. Additionally, the CE Declaration of Conformity as well as ATEX approvals are available on the SITRANS F literature CD-ROM. Certificates (http://support.automation.siemens.com/WW/view/en/10806951/134200) SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...
Page 104
Approvals/Certificates SITRANS F M MAG 5000/6000 Operating Instructions, 12/2013, A5E02338368-002...