Manual EE371 Dew Point Transmitter / Switch down to -60 °C Td BA_EE371_e // V2.0 // technical data are subject to change...
Page 2
E+E Elektronik Ges.m.b.H. doesn't accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products. The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis. These changes will be implemented in later versions. The described products can be improved and changed at any time without prior notice.
The operation manual may not be used for the purposes of competition without the written consent of E+E Elektronik® and may not be forwarded to third parties. Copies may be made for internal purposes. All information, technical data and diagrams included in these instructions are based on the information available at the time of writing.
1.2.3 Intended Use The EE371 is intended for the measurement of dew point temperature (Td), frost point temperature (Tf) and volume concentration (Wv) in air (compressed air) and other non-corrosive and non-flammable gasses, depending on the model up to 20 bar (290 psi) respectively 100 bar (1450 psi). The use of the EE371 other than described in this manual is not allowed.
The EE371 transmitter features two freely selectable and scalable outputs for Td, Tf or Wv. Modell S: The EE371 switch has two relay outputs for control and alarm purposes. The status of the two relays (configured for instance for early warning and main alarm) is indicated by LEDs. The analogue outputs as well as the Td/Tf switch threshold and hysteresis can be set with the EE-PCS Product Configuration Software, free download from www.epluse.com/configurator.
Accuracy ≤ ±2 °C Td Accuracy not specified Air (gas) temperature Td [°C] Output signal limitation Fig. 1 Dew point measurement range and specified accuracy The maximum scaling of the analogue output is -100...80 °C (-148...176 °F) dew point. Manual EE371 Dew point transmitter...
Remove protection cap For accurate measurement and short response time, the sensing head shall be immersed in the air or gas to be monitored. The mounting of EE371 shall be tight; moisture ingress from the surroundings falsify the measurement results.
Please note: • Pressure gradients between the process and the sampling chamber lead to significant measuring errors. • The mounting of EE371 with sampling cell shall be tight; moisture ingress from the surroundings falsify the measurement results. Do not employ any hygroscopic materials for mounting EE371 or the sampling cell. • Keep the sampling line as short as possible.
4. Diagnostic LED 5. Display 1. Serial interface: Connector for serial interface cable HA010604 or EE-PCA Product Configuration Adapter with HA011063. See data sheets EE371, Accessories and EE-PCA at www.epluse.com/ee371. 2. Status lEDs: Provide information on the status of the EE371. See chapter „“5.2 Status LEDs”. 3. Current-/ voltage output: Jumpers for selecting the analogue output signal. The change from voltage to current or vice versa shall be performed both via hardware (jumpers) and software (with the EE-PCS Product Configuration Software).
An adjustment can be followed by a calibration which documents the accuracy of the adjusted specimen. The dew point temperature adjustment of EE371 can be performed with the free EE-PCS Product Con- figuration Software, free download at www.epluse.com/configurator...
EE371 is correctly powered / the microprocessor is active • constantly lit => Electronics defect => please contact the producer lED D1 (blue): • constantly lit => The sensing element is damaged • flashing => The sensing element is wet (condensation) Self-diagnostic via display (option): • Error 1 => The humidity sensing element is damaged • Error 2 => The sensing element is wet (condensation) • Error 3 => The temperature sensing element is damaged • Error 4 => Short circuit on the temperature sensing element Manual EE371 Dew point transmitter...
LC display: -20...50 °C (-4...122 °F) Storage temperature range -40...60 °C (-40...140 °F) Electromagnetic compatibility according to EN 61326-1 EN61326-2-3 ICES-003 ClassB Industrial Environment FCC Part15 ClassB 1) minimum supply voltage 15 V DC Manual EE371 Dew point transmitter...