Caution destruction of the equipment. The TRIVAC NT 5/10 pumps from Oerlikon Leybold Vacuum have been designed for safe and efficient operation when used properly and in accordance with this manual. It is the responsibility of the user to carefully read and strictly observe all safety precautions described in this chapter and throughout the manual.
Moreover, these pumps are not suited for pumping of hazardous gases or extremely aggressive or corro- sive media. If in doubt please consult Oerlikon Leybold Vacuum. The drive motor of the TRIVAC NT is screwed directly to the bearing piece.
12 Diaphragm valve 13 Exhaust valve 14 Exhaust port Fig. 2 Operating principle of a TRIVAC NT 5 (other TRIVAC NT types operate in a similar way) Principle of Operation The rotor (2/6) which is eccentrically arranged in the pump housing (pump chamber) has two radially sliding vanes (2/7 and 2/9) which divide the pump chamber of the pump into several chambers.
177 01 177 02 20 l 177 03 Use only types of oil specified by Oerlikon Leybold Vacuum. Note Alternative oils are stated upon request. The necessary connection components for connecting the exhaust filter and the condensate separator in each case are stated in the following:...
If this is not the case you must interchange two of the mains phases. Note the direction indicating arrow on the fan cowl. 3.4.2.1 Current Drawn by the Three-Phase Motors of the TRIVAC NT 5 and NT 10 Voltage ranges for Max. current consumption Min.
Service at Oerlikon Leybold Vacuum Contamination If you send equipment to Oerlikon Leybold Vacuum indicate whether the pump is free of substances damaging to health or whether it is contamina- ted. If it is contaminated also indicate the nature of hazard. To do so, you must use a preprinted form which we shall send to you upon request.
■ Replacement of all gaskets = Annual maintenance ■ Functional check. n-a = Maintenance every n years This check should be performed by the Oerlikon- * At high gas throughputs and higher temperatures within a Leybold-Vacuum-Service. pressure range >100 mbar shorter maintenance intervals apply.
Disassemble clean equipment, separate and dispose of the materials accor- ding to their type. When returning equipment back to us, please note the information provided in Chapter “5.1 Service at Oerlikon Leybold Vacuum”. Disposal of Waste Oil Owners of waste oil are entirely self-responsible for proper disposal of this waste.