3. Operation Procedure
1) Connect the probe, operate the Main unit, and adjust the image
according to the instructions given in the operation manual for the
main unit.
2) Confirm that the probe is disinfected or sterilized.
3) Confirm that the biopsy attachment is sterilized.
4) Confirm that the Magnetic sensor attachment and the Spacer for
EZU-RV2S are disinfected or sterilized.
5) Prepare the following probe cover.
CIV-Flex™ Transducer Cover (610-542) (Option)
Sterile 14 x 91.5cm telescopically-folded cover
6) To use Real-time Virtual Sonography (RVS), attach the Magnetic sensor
to the probe as shown in "4. Attaching the Magnetic Sensor". Place
the probe cover to the probe, after attaching of the Magnetic sensor.
Do not attach the Magnetic sensor on top of the probe cover.
7) Place an appropriate amount of sterilized gel on the Probe head.
(See Fig.6)
CIV-Flex cover
Sterilized gel
Probe head
Fig.6
Placing the sterilized gel
-8-
Q1E-EP1245