10
1.6

Safety information and preventive measures

NOTE
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02
When completing repair work and tests, please note:
- the product-specific safety information in the document,
- the safety information in TD00-000.860.01... ,
When the system is switched off via the operating panel or
S2/D711, line voltage is still present at the line voltage connec-
tion for the generator, at line voltage filter Z1, at transformer T1,
at transformer T10, and on board D711 (refer to circuit dia-
gram).
After switching off the system, approximately 380 V DC may be
present in the intermediate circuit as indicated by the V24 LED
on board D710. Within 3 minutes, the voltage will drop to less
than 30 V, and the LED will go out.
Tests or adjustments that must be made with the radiation
switched on are identified by the radiation warning symbol.
During these types of adjustments, radiation protective cloth-
ing must be worn.
02.06
General information
Page 10 of 34
CS PS SP
Siemens AG
Medical Solutions