Column stand
3-
3
Column stand
3.1

Checks

SIM
Steel ropes
Do the steel ropes (balancing and flap), the drive belt, the suspension, or the pulleys
have any damage (e.g. fraying or splitting) or show signs of wear?
PMA
Oil and grease
If they are OK, lightly oil the wire cable, grease the rails, and then remove the protective
strips.
SIE
Limit switches
System ON. Move the lifting carriage up and down several times, the lifting carriage
should be in an upright (0°) position.
NOTE
Do the limit switches stop carriage travel at the top and at the bottom?
Activate safety switch S882. System movements must be blocked.
PMF
Vertical travel
Is the vertical travel smooth and noiseless?
Switch OFF the system and attach the protective strips once again.
SIM
Safety catch and rotation safety catch
Do the safety catch
visible damage? Are the springs in good condition?
Fig. 1:
Siemens AG
Medical Solutions
It is normal to hear some noise from the rotation of the radiation
filter in the collimator when switching the system ON.
(1/Fig. 1 / p. 17)
Safety catch
SPB7-230.101.02.05.02
02.06
and rotation safety catch
Page 17 of 34
CS PS SP
(2/Fig. 1 / p. 17)
have any
MAMMOMAT 1000/3000 Nova
17