System status
2-
2
System status
2.1
Checks
Check the system status and fill out the Maintenance Protocol.
•
Check the serial no. and the part no. for the workstation, the CCD camera, and the biop-
sy controller.
•
Check the license key (DICOM option).
•
For instructions on how to check the hostname/hostid, see the Service Instructions
for Opdima®, chapter "Instructions for obtaining the system host ID and host name".
•
Document country settings in the Maintenance Protocol
•
If applicable, document the network and printer parameters in the Maintenance Proto-
col.
Siemens AG
Medical Solutions
SPB7-230.101.03.05.02
02.06
Page 9 of 18
CS PS SP
MAMMOMAT Nova - Opdima
9