System
3.1

Checks

PMP
MO storage media and MO unit
The MO disks must be handled with care. They should never be forced into or out of the
drive. They should not be exposed to heat or fluids. Never place heavy objects on an MO
disk. The MO drive unit is relatively sensitive to dust.
Be sure that the unit is installed in a clean environment and is not exposed to dust.
Check that the correct MO media is used. Only Sony standard media
(Siemens part no. 66 33 734, Sony RMO-S561, 9.1 GB rewritable)
can be used as MO media.
Check that the MO disks are stored in a clean and safe place.
Check that the local disks and backup disks are properly labeled.
Check that backup disks are stored separately.
The MO disks used in the Opdima® system are equipped with a mechanical write pro-
tection lock. This lock should NEVER be used since the Opdima® system always at-
tempts to mount the MO disk as a read/write device.
Fig. 1:
PMP
SCSI cables, CD-ROM
It is important to check that the SCSI cables are properly connected and the hardware is
terminated correctly. If not, this might result in a number of problems, e.g. failure to write
images to MO disk, corrupted MO disk, corrupted hard disk.
Make sure the SCSI cables are properly connected and attached to the units and that
the cable is not bent too much just behind the MO.
The MO drive is to be terminated with the external termination plug.
If the system is not able to store images on the MO disk due to the above problem, it will
take longer to save images when the system can save.
MAMMOMAT Nova - Opdima SPB7-230.101.03.05.02
Write protection lock not activated.
02.06
Page 10 of 18
CS PS SP
System
Siemens AG
Medical Solutions