MN1-5823 Rev.6
4-2. How to attach/release Magnetic Sensor Attachment and Magnetic Position Sensor
Follow the instructions below to attach/release the Magnetic Position Sensor and the Magnetic Position Sensor
Attachment. All following procedure must be performed out of operative field.
4-2-1. How to attach Magnetic Position Sensor
(1) Align the direction mark of the Magnetic Position Sensor and the hole of the Magnetic Position Sensor
Attachment, and place the sensor into the attachment (Figure 3).
The Magnetic Position
Sensor Attachment
Align the direction mark
The Magnetic Position
Sensor
(2) Attach the attachment to the probe as shown in Figure 4.
Caution
Attach the Magnetic Position Sensor Attachment to the probe correctly.
If the sensor is attached in the wrong direction, this causes false diagnosis.
Figure 3 How to attach Magnetic Position Sensor
Figure 4 How to attach Magnetic Position Sensor Attachment
-10-
Align the direction mark