Screen 6: Liquid level sensor
The Q-Series ultrasonic bath is fitted with a liquid level sensor to ensure it is
not under-filled prior to or during the cycle. This ensures that the optimum
cleaning process can occur and potential damage to the tank is minimised.
The level sensor is a resistive device, so will not detect liquids with very low
conductivity such as pure deionised (DI) water.
The level sensor can be disabled for specialist cleaning applications where
deionised (DI) water is used.
Use the ON and OFF keys to set the Low-Level function. When the desired
option is selected, use the MENU key to move to the next screen.
When the OFF button is pressed and the  key is selected to continue, a
message will appear to check that the operator has filled the tank up to the
level indicated (see page 7).
When the level sensor is disabled the liquid set temperature is automatically
reduced to 5°C. The set temperature may be increased by entering the
MENU system – see page 11.
If 'NO' is selected an instruction will appear requesting the tank be filled to
level before use.
Q-Series Operating Manual
21