b. When completed, the device will indicate "Test Passed" or "Test Failed".
i
ii
16. If the leak test has passed, then clear the device sieve hour counter.
a. Turn on your Zen-O lite portable oxygen concentrator.
b. Simultaneously press the
c. Press the
EN
played, where XXXX = number of hours for the old sieve was utilised.
d. Press the alarm mute button
e. Press the battery button
f.
The device is now ready to be used.
Important: Contact your home oxygen provider or distributor if the device does not pass after mul-
tiple attempts.
12.3. Disposal
• Contact your distributor regarding disposal of the device.
• Dispose of battery according to local regulations or contact your distributor.
Warranty
13 .
The device warranty is limited to three (3) years from date of manufacture, this excludes the
sieve modules which have only one year warranty. All accessories including batteries are lim-
ited to one (1) year warranty.
The standard warranty is only valid for products handled as stated in the user manual and in
accordance with general industry good practice and standards.
Trademarks and Disclaimer
14 .
14.1. Trademark
All trademarks are the property of their respective owners.
14.2. Disclaimer
The information in this document has been carefully examined and is believed to be reliable.
Furthermore, the manufacturer reserves the right to make changes to any products herein
to improve readability, function, or design. The manufacturer does not assume any liability
arising out of the application or use of any product or circuit described herein; neither does it
cover any license under its patent rights nor the rights of others.
www.oxygenconcentratorstore.com
Test Passed - Press battery button
Test Failed - Press battery button
or
button to navigate he diagnostic menu until "XXXH Sieve" is dis-
to exit, continue to step 16.
to exit, then repeat steps 4-15.
and
buttons until "Diagnostics" is displayed.
until the sieve hour counter shows "0H Sieve".
to exit the diagnostic mode.
28/292
877-774-9271