Important Notes
Delay times between two exposures
The delay times listed below must be followed to prevent the tube from overheating.
Tab. 4
Delay times between two exposures
Exposure mAs value
max. 100
max. 200
max. 300
max. 400
max. 500
Siemens AG
Medical Solutions
Delay time between two exposures (seconds)
min. 15
min. 30
min. 45
min. 60
min. 75
SPB7-250.815.06.02.02
12.05
MAMMOMAT Novation DR
Page 15 of 116
CS SD 24
15
0