Delay times between two exposures

Delay times listed below must be followed in order to prevent the tube from overheating.
Exposure mAs value
max 100
max 200
max 300
max 400
max 500
M1000/3000 Nova
Installation and Start-Up
Delay time between two exposures (seconds)
min 15
min 30
min 45
min 60
min 75
Register 3
SPB7-230.033.11
Rev. 06
Protective measures
Page 2 of 2
05.06