Checking the compression device and OPCOMP
Checking the OPCOMP
1. Decompress.
2. Set the compression force to 20 kg by using the presetting knob (5/ Fig. 1) and
Adapting OPCOMP
If the factory-set OPCOMP is not to the customer's satisfaction, an adjustment can be
done with the tautness factors in the "Best compression table", see Service Instructions.
A lower tautness value will increase the compression force and vice versa.
•
Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this
setting for at least two weeks, before further changes are made.
NOTE
Siemens AG
Medical Solutions
compress.
The compression movement must be interrupted by OPCOMP before 20 kg is
reached, i.e. at 5-15 kg depending on the material (4/ Fig. 1) used.
OPCOMP has been thoroughly tested by Siemens. If any factors
are changed, OPCOMP will no longer be an optimization of com-
pression force and image quality according to Siemens' clinical
test.
Register 3
SPB7-230.033.11
Rev. 06
05.06
Page 3 of 4
M1000/3000 Nova
CS PS SP
Installation and Start-Up
17 - 3
17
17