Removing The Swivel-Arm System Transport Safeguard - Siemens MAMMOMAT 1000 Nova Installation And Start-Up Instructions Manual

Hide thumbs Also See for MAMMOMAT 1000 Nova:
Table of Contents

Advertisement

Preparatory work
Removing the transport safeguards

Removing the swivel-arm system transport safeguard

Fig. 2 Swivel-arm system transport safeguard
1. Loosen the transport safeguard nuts (1/  F ig. 2) and remove the tube head screws
2. Push the red transport safeguard (3/  F ig. 2) inwards, away from the tube head.
NOTE
Transport safeguard for the lifting carriage
Fig. 3 Transport safeguard for the lifting carriage
1. Loosen the two screws (1/  F ig. 3) and remove the red transport safeguard (2/ Fig.
Siemens AG
Medical Solutions
1
(2/  F ig. 2).
Removal of the swivel-arm system transport safeguard is done
after the equipment has been powered up, see further section
"Attaching the swivel-arm covers" on page 7 - 1.
3). Reinsert screws (1/  F ig. 3), they might be needed for service purposes.