Further programming

Setting the real time clock

Prerequisite

The correct date and time must be set in the service PC. If not set date at C:\> by typing
ENTER
<

Setting the time in the MAMMOMAT

Main menu:

1. Select <Configuration>.
2. Select <Clock>.
3. Press <F2>.

Reducing the generator power

NOTE
To achieve full power, the measured resistance must not exceed the following values:
0.25
O
0.45
O
0.50
O
0.60
O
0.65
O
0.85
O
If the above values are exceeded, reduce the generator power as follows:
Main menu:
1. Select <Configuration> <Miscellaneous >.
2. Select <Red. of power>.
3. Select nominal line voltage ____ V with cursor keys <
4. Type in line impedance ____
5. Save the programmed values with <F2>.
Siemens AG
Medical Solutions
and time at C:\> by typing
>
With <F2> the date and time set in the service PC are taken over by the
MAMMOMAT.
If the power is reduced after calibrating and adjusting the AEC, a
new calibration will be necessary!
at 110 V (1-phase)
at 208 V
at 230 V
at 240 V
at 277 V
at 400 V (2-phase)
Reduction of power depending on line quality.
Tube is P 40 MoW.
Calculated power is ___ kW. Max. possible tube power is displayed.
Register 3
ENTER
time <
>.
. Enter the measured line impedance.
SPB7-230.033.11
Rev. 06
05.06
14
> <
> and < ENTER>.
Page 1 of 2
M1000/3000 Nova
CS PS SP
Installation and Start-Up
14 - 1
14
14
date
14
14
14