hit counter script

Checking And Recording For The Area Of Application Of The X-Ray Decree (§16 Germany) - Siemens MAMMOMAT 1000 Nova Installation And Start-Up Instructions Manual

Hide thumbs Also See for MAMMOMAT 1000 Nova:
Table of Contents

Advertisement

1 - 4
Important notes on start-up
The
MAMMOMAT
tion to the on-site mains voltage, adjustment of the AEC and the functional tests to be per-
formed.
When the measurements to be made (kV, mAs, etc.) are within the tolerances stated in
these instructions, this confirms that the settings made in the factory have not changed
and the equipment is fully serviceable.
The measured values marked with
entered in the "start-up" column in the test certificate provided.
The service PC is required only for programming according to:
• Chapter 10 "Checks with high voltage
• Chapter 11 "Calibrating and adjusting the AEC
• Chapter 12 "Dose Calculation System
• Chapter 13 "Testing"
• Chapter 14 "Further programming"
• Chapter 20 "Final procedures
Description of the service PC syntax is found in Chapter 22 "Appendix".
NOTE
Checking and recording for the area of application of the X-ray
decree (§16 Germany)
In the area of application of the X-ray decree, an acceptance test certificate is supplied,
with most of the data filled in by the factory.
Only:
the cut-off dose
the resolution of the film/screen systems used
the output values of the constancy test
must be determined by the owner of the equipment and recorded in the acceptance test
certificate. Furthermore, the front page of the acceptance test certificate must also be
completed with the operator´s data.
M1000/3000 Nova
Installation and Start-Up
 is adjusted, programmed and tested in the factory, leaving the adapta-
"
If the generator is switched off with the service PC connected,
wait approximately 5 s before switching it on again.
Register 3
SPB7-230.033.11
Rev. 06
in Chapter 10 "Checks with high voltage" shall be
"
"
"
Page 4 of 6