Calibrating and adjusting the AEC

Sensitivity

Adjustment of O.D. for the whole AEC is performed under <Mainmenu>, <Configuration>,
<AEC> and <Sensitivity>.
<ESC> to exit, <TAB> move to next entry field
1 Help
2
Save
NOTE
Siemens AG
Medical Solutions
H
Sensitivity
0
valid entries from -64 to 64 in 1/8 E.P. (+/-8 E.P.)
3
5
The O.D. preferred by the customer can be set with this
function. The customer's own phantom(s) can also be
used.
Register 3
D
0
6
7
SPB7-230.033.11
Rev. 06
05.06
8
9
10 Quit