5 - 4
Connecting the stand cable-harness
High-voltage connector
Fig. 4 High-voltage connector
NOTICE
M1000/3000 Nova
Installation and Start-Up
3
1
2
1
6
1
0
0
M
A
M
The high-voltage connector (1) is factory-adjusted for correct fit.
The high-voltage connector shall be inserted with silicone oil, sili-
cone-rubber ring (2) and silicone-rubber disc (3). Lubricate the sil-
icone-rubber disc and the pins with silicon oil before mounting
the disc. Make sure that there aren't any air bubbles between the
disc and the connector.
Register 3
SPB7-230.033.11
Rev. 06
Cable connections
Page 4 of 14
05.06
CS PS SP
5
5
Siemens AG
Medical Solutions