Installing the generator and the stand
Installations with separate control console
Lay the connecting cable from the control console in additional cable ducts, alternatively
underfloor, to the free cable entry of the generator, see further section on page 5 - 8.
The separate control console can be wall-mounted – horizontally or vertically as shown in
Fig. 5 below – or be mounted on a table, either in the examination room or in an adjacent
room with radiation-proof window. It is important that the control console be placed so that
the operator has good view of the patient from the control console.
Ensure that the control console is attached to the wall with adequate safety margins. Dow-
els and screws are not supplied. Please obtain these locally and ensure that they are suit-
able for the material used in the wall. The separate control console has a weight of 4 kg.
There are two cable outlets on the separate control console; one at the back and one on
the underside. On delivery, the cable is led through the outlet at the back of the control
console. For wall-mounted control console, the cable shall be led through the outlet on the
underside.
If the control console is to be mounted on a table, the minimum/maximum length of the
screws penetrating into the control console must be observed, see Fig. 7 below.
The control console can also be mounted on the free-standing radiation shield (option) as
shown in Fig. 6 below, see separate instructions enclosed with the radiation shield.
8
6
1
0
0
M
A
M
Horizontal mounting
Fig. 5 Wall-mounted control console
Siemens AG
Medical Solutions
∅
Plastic cap
Wall-mounted cable duct
is recommended
Register 3
SPB7-230.033.11
Rev. 06
6.5 mm
Vertical mounting
05.06
Page 5 of 6
M1000/3000 Nova
CS PS SP
Installation and Start-Up
4 - 5
4
9
6
1
0
0
M
A
M
6.5 mm
∅