DEFIBRILLATION ELECTRODES (PADS)
An audible and visual alert will indicate after the self-test if the pads are missing, unplugged or damaged.
CAUTION: Damaged or Expired Pads
Using pads that are damaged or expired may result in improper AED Pro performance.
PAD INSTALLATION
CAUTION: Use only Approved Equipment
Using batteries, defibrillation pads, cables, or optional equipment other than those approved by
General Electric may cause the AED Pro to function improperly during a rescue.
CAUTION: Possible Improper AED Pro Performance
Using defibrillation pads that are damaged or expired may result in improper AED Pro performance.
CAUTION: Both polarized and non-polarized defibrillation pads are available for Responder AED Pro. If using
polarized pads, always place sternum and apex pads as shown on the packaging. Non-polarized
pads may be placed in either position for a rescue; however, the ECG waveform will be correctly displayed in only
one position. To correctly display an inverted ECG, simply reverse the defibrillation pads.
DIRECTIONS FOR USE:
1.
Do NOT open until ready to use, short term use only.
2.
Ensure the skin site is clean and dry.
3.
Separate one pad from liner.
4.
Place one pad on skin.
5.
Peel and place remaining pad.
Page 16 of 53
The defibrillation pads come in a ready-to-use, sealed package containing one pair of
self-adhesive pads with an attached cable and connector. The pads are disposable
and should be discarded after each rescue. The pads have a limited shelf life and
shall not be used beyond the expiration date. Keep a fresh, unopened pair of pads
plugged into the AED Pro at all times. Refer to the pads package label for operation
temperatures.
1.
Open the lid of the AED Pro.
2.
Place the pad package into the lid so that the expiration label is visible
through the clear window on the lid. The expiration date of the pads will
then be readable without opening the lid of the AED Pro.
3.
Match the color of the connectors (red to red), slightly lift the tab of the
pad socket then plug the pad connector into the AED Pro case as shown
in the drawing.
4.
Tuck the excess cable length in the bottom holder as shown in the
drawing. With the pad package completely secured to the AED Pro lid,
close the lid.
5.
Make sure the expiration date is visible through the clear window of the
lid.
6.
Make sure that the STATUS INDICATOR is GREEN.