"device". It does not apply to earlier product versions. Due to continuing product innovation, specifications in this document are subject to change without notice. 12SL, CASE, CardioSoft, InSite ExC, MAC, MACCRA, MARS, MUSE, Marquette, MobileLink, and MULTI-LINK are trademarks owned by GE Medical Systems Information Technologies, Inc., a General Electric Company going to market as GE Healthcare. All other trademarks contained herein are the property of their respective owners.
Page 3
(https://sourceforge.net/projects/sshpass/) License details of the software used in the product can be viewed in the online help in the Open Source Licenses section. Contact GE Service to obtain the source code of the open-source software used in the product, if required.
Table of Contents Publication Information................... 2 1: Product Overview....................8 Front and Rear View............................8 Keyboard..................................10 Acquisition Screen Overview......................... 11 User Menu Options Description........................14 Battery Status............................... 16 Show Battery Status............................17 Show Network Connection Status......................18 Changing the Brightness of the Screen....................19 2: Login and Security.....................
Page 5
Automatically Print an ECG Patient Report................56 Record a Rhythm..............................57 Review a Digital Rhythm Report......................59 Transmit a Patient Report to a Configured Destination..............60 Print a Patient Report............................63 Edit Patient Information in a Patient Report..................65 Delete a Patient Report............................66 View the Patient Report........................... 66 Close a Patient Report............................67 Full Disclosure Overview..........................67 Record a Full Disclosure ECG......................
Page 6
11: Cleaning and Disinfection................113 Clean and Disinfect Guidelines........................113 Inspect the Device............................114 Care at the Point of Use..........................114 Preparation for Cleaning..........................114 Visual Inspection, Cleaning and Disinfection Frequency...............114 Cleaning and Disinfecting the Acquisition Module................115 Pre-Cleaning Inspection and Functional Testing..............115 Cleaning the Acquisition Module....................
Product Overview Product Overview ™ The MAC VU360 Resting ECG Analysis System, (referred to as "the device" throughout this document), supplies 12-lead or 15-lead ECG measurement and interpretative analysis, prints 12-leads or 15-leads of ECG, and transmits ECG data to and from a central ECG cardiovascular information system.
Page 9
Used to change the position of the rear bin. Rear Wheels and Used to mobilize and lock the trolley in position. The rear wheels are Lock swivel wheels and conductive. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 9...
Capitalizes the letters in a text. Moves to the next field in a screen. Shift + Tab key moves to the previous field in a screen. Closes a window on a screen. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Deletes characters or text. Acquisition Screen Overview The Acquisition screen is the main screen that displays when you first log on to the device. You can acquire and ECG from the Acquisition screen. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 11...
Page 12
Resting ECG Analysis System Setup and Configuration Manual. Patient Information Banner Displays Patient Information such as the patient first name, surname, and gender. Select anywhere on the banner to add or edit patient information. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 13
Lead Quality Indicator is consistently green for some seconds. Hookup Advisor Lead Displays the lead quality status indicator in three circles that Quality Status Indicator change to yellow, red, or green, based on the lead quality. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 13...
For a new patient, the values are reset to the default settings. User Menu Options Description The User Menu is located at the top right corner of the Acquisition screen. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 15
Service The user can get a service snapshot without the Service privileges. Snapshot Complete the snapshot to help identify a problem on the device. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 15...
AC Mains power. When the device is connected to the AC Mains power and the battery icon is Disconnected solid green, shows that the battery is fully charged. from AC Mains MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
To show the status of the two batteries, use the following steps. Select the battery icon on the Status Bar of the Acquisition screen. An image opens, showing the battery life. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 17...
WLAN Active The device is connected to a WLAN and has a valid IP address. The icon shows a number of wireless bars to indicate the strength of the wireless signal. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Power. battery. The changes you make are automatically saved to your device and will not change when you turn the device on or off. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 19...
The Power off window opens and displays a message. Select Power Off. The device is off. The Power on LED on the keyboard is off. Press the Power button on the keyboard for a few seconds: MAC VU360 ™ Resting ECG Analysis System...
LDAP users (If LDAP-based user authentication is configured) • Local users (If user profiles are locally managed on the device) • STAT User if STAT button is selected. The user can configure the text for this button. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 21...
• If the device is shut down, power on the device. See Power On the ECG Device on page • If the device is locked, unlock the device. See Unlock the Device on page MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 23
LDAP group authorized for this device and log on to the device again. • Your current password has • Contact your LDAP expired. administrator to change your password. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 23...
• From the User Menu on the Acquisition screen, select Log Out to log off the device. If you log off before a task is completed, a message displays that you will lose incomplete data. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Login and Security The GE logo is displayed on the center of the screen with a black background, and a message displays indicating screen privacy is turned on. • To deactivate privacy mode, press any key on the keyboard or tap anywhere on the screen.
Use the default admin password to log on to the device. It keeps the previously enabled option codes, serial number, keyboard language, MAC address, and Wireless Country of Operation configuration. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 27...
Page 28
The device reboots. You can access the device as the Default user without login credentials. To reconfigure the device, access the Settings screen from the user menu. A login screen opens. Log on as Admin user with the default password admin123. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Make sure that you enter patient data for the correct patient. On the Acquisition screen, the Patient Information bar above the waveform and shows minimum information about the patient. The text boxes shown in the bar are configurable. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 29...
Page 30
Shows the unique identification number of the patient (patient ID). Select the Patient Information bar to expand it into a full screen view. Fields configured by your administrator are shown in the Patient Information screen. MAC VU360 ™ Resting ECG Analysis System...
If there are pending print tasks in the queue, a message opens that tells pending print tasks will be deleted from the queue. • Select Continue to delete pending print tasks and start the test for the new patient. The Patient Information screen expands. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 31...
• The barcode reader is connected to the device, and the device is correctly ™ configured to use the peripheral. For more information, see the MAC VU360 Resting ECG Analysis System Setup and Configuration Manual. • You are in the Live or Full Disclosure ECG screen, or when patient demographics banner is open.
• Use orders of ADT data as configured on the device to do remote patient query. • A barcode reader is connected to the device. Start a test for a new patient. For more information, see Start a Test for a New Patient on page 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 33...
Page 34
No local orders are found on the device A message displays in the notification area that no matching local orders are found. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 35
Remote ADT data query failed on the MUSE A message opens in the notification area that ADT system query has failed. When local or remote orders are found, and you try to attach the order: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 35...
Page 36
No local orders are found on the device A message displays in the notification area that no matching local orders are found. The device then searches for remote orders in the MUSE system with Patient ID. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 37
Remote order query failed on the MUSE system A message displays in the notification area that remote order query has failed. The device then searches for ADT data in the MUSE system with Visit Number. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 37...
Page 38
Patient Information screen screen: Are you sure you want to attach this order to the current test? Select Yes to attach the order, No to cancel the action. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
• If you enter incorrect data in a field, the field border changes to red. • An asterisk (*) displays adjacent to mandatory fields in the Patient Information. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 39...
Page 40
ECG report will not be accepted, transmitted, or printed until you enter the patient demographic data for any of the mandatory fields. You need to complete the data for the mandatory fields. MAC VU360 ™ Resting ECG Analysis System...
When you connect the RA/R leadwire and another leadwire to the patient, an electrode placement image expands in the Hookup Advisor panel. If all leadwires are disconnected from the patient, the electrode placement image collapses after a few seconds. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 41...
Page 42
Table 13: Lead Quality Indicators on the Electrode Placement Image Lead Quality Description Indicator Green The leadwire connection is good. The acquisition module is sending a good signal to the device. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 43
ELECTRICAL INTERFERENCE - Electrostatic discharges may interfere with the acquisition of ECG recordings. The acquisition module could be temporarily disconnected with an error message displayed due to an ESD event. The device 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 43...
The Acquisition screen displays the waveform based on the configured lead format, lead set, speed, gain, and filter. After you start a new patient, you can change the lead set or lead format on the Acquisition screen. MAC VU360 ™ Resting ECG Analysis System...
ACS must be enabled on a per-patient basis. At the bottom, left-side of the Acquisition screen, select anywhere to the left of the ellipsis icon (1). 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 45...
To adjust the speed, gain, and filter of the waveform as desired, select a different value from the list. The selected values are applied to the waveform. Select anywhere outside the menu to collapse it. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Accept or Reject an ECG Patient Report on page Automatic ECG acquisition is triggered only one time for the current patient test. Start any new ECG tests for that same patient manually. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 47...
Start recording the patient ECG by performing one of the following: • Select the Start ECG icon at the bottom of the Acquisition screen: • Press the ECG key on the keyboard: • Press the ECG button on the Acquisition Module: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 49
Accept or Reject an ECG Patient Report on page as Yellow and the Hookup Advisor Status is Yellow or Red Print preview mode is configured as Red and the Hookup Advisor Status is Red 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 49...
CAUTION: DELAY IN TREATMENT - Unaccepted ECGs in the preview screen will be automatically rejected and deleted when all patient leads are disconnected and the MAC VU360 device is inactive for 2 minutes. MAC VU360 ™ Resting ECG Analysis System...
Page 51
Information banner to edit patient information for the patient report. Modify the information by using the keyboard, attaching an order, scanning the patient barcode, selecting a patient from the Patients tab, or performing an ADT query. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 51...
Page 52
Select one of the options as follows: • If you select Continue, the preview will be lost. • If you select Cancel, you can proceed to accept or reject the ECG preview. MAC VU360 ™ Resting ECG Analysis System...
Page 53
An error message Unable to accept. Incomplete patient data. displays on the Acquisition screen. You need to complete the data for the mandatory fields to accept the ECG report. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 53...
Acquisition screen. Review an ECG Patient Report When you disconnect patient leads and there is no activity with the MAC VU360 device for 2 minutes, all ECG reports will automatically close. After the 10 seconds ECG is acquired and the ECG preview is accepted, the patient report is displayed in the ECG tab in the configured report format for review.
Page 55
Deletes the patient report. For more information, see Delete a Patient Report on page Edit icon Edits patient information for the patient report. For more information, see Edit Patient Information in a Patient Report on page 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 55...
After you select Done on the Review screen of an ECG patient report, disconnect patient leads before you select Start New Patient on the MAC VU360 device. To start an ECG for a new patient, select Start New Patient. For more information,...
Start the rhythm for the patient in one of the following ways: • Select the Start Rhythm icon on the Acquisition screen: • Press the Rhythm key on the keyboard: • Press the Rhythm button on the Acquisition Module: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 57...
Page 58
Select the Stop Rhythm icon on the Acquisition screen to stop both digital rhythm recording and printing: To stop printing the rhythm report while the digital acquisition of rhythm continues, perform one of the following: • Press the Stop key on the keyboard: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
DICOM Gateway Pro SP1 or higher) Resting ECG USB R/W flash drive PDF, Hilltop and Sapphire XML (Sapphire XML format is available if you enable the XML format output in the Option Manager) formats. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 61
Review a Digital Rhythm Report on page To review a Full Disclosure report, see Review a Full Disclosure Report on page To transmit the report to the required destination, perform one of the steps below: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 61...
Page 62
Select Continue to re-transmit the patient message displays in the notification asking report. you to confirm if you want to re-transmit the already transmitted report. • Select Cancel to cancel the report transmission. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
• The patient information in the patient report is correct. • The ECG or Rhythm or Full Disclosure ECG is acquired with the desired gain and filter. Perform one of the following steps: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 63...
Page 64
The barcode can be used to perform a query by patient ID in the MUSE system. To stop printing a patient report, do one of the following: MAC VU360 ™ Resting ECG Analysis System...
Select Save to save your changes for this patient and collapse the screen. If you select any other icons at the bottom of the tab prior to saving, the Patient Information screen collapses and the edited patient information is saved. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 65...
Select this icon or double-tap the patient report to maximize the view of the patient report. Minimize View Select this icon or double-tap the maximized patient report to minimize the view of the patient report. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Options Manager. Automatic acquisition of ECG does not start while in the Full Disclosure tab. Automatic ECG acquisition stops when you change from the Live tab to the Full Disclosure tab. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 67...
To restart the Full Disclosure ECG, click Restart. Deletes all the current waveform data. The message unsaved data will be lost. Confirm to proceed displays. To record a 10 seconds ECG when in the Full Disclosure screen, do the steps that follow: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
To review the Full Disclosure report, see Review a Full Disclosure Report on page Review a Full Disclosure Report The Full Disclosure report displays in the FD Report tab for your review. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 69...
Page 70
Only a FD Report or Rhythm Report can have a single tab. ECG will always have a tab, but if you enable a Rhythm tab and select a FD Report, the Rhythm tab will be replaced by the FD tab. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 71
To start a new Full Disclosure report for the same patient, click the Full Disclosure tab to navigate to the Full Disclosure waveform and application, and restart the Full Disclosure ECG. For more information, see Record a Full Disclosure ECG on page 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 71...
(such as a MUSE system) that is connected to your network. When the orders list is updated, either automatically or manually, new orders are populated in the list. The figure illustrates the Orders collapsed list: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 73
Filter Location Select anywhere on the Filter Location field. From the drop-down menu, list select the location filter you want to apply to the orders list. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 73...
Page 74
• Bed Number • Test Indication • Priority • Comments • Technician • Patient History • Location • <Question> • Attending MD ID • Attending MD First Name • Attending MD Last Name MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
The Orders list is filtered by Show All Locations, Current Patient Location, or 1 out of 10 pre-configured filter groups. All of the order lists display based on the location filter you apply. From the Acquisition screen, select the Orders tab. The Orders collapsed list opens. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 77
The orders list displays only those locations in the selected filter group. The Current Patient Location The orders list displays only the current location of the device configured in the device settings. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 77...
Page 78
To display an expanded list of Orders, select the Expand icon. The Orders expanded list opens. By default, the applied filter is Current Patient Location , and displays all orders from the current patient location. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Start a test for the new patient. For more information, see Start a Test for a New Patient on page Double-tap the order in the Orders list to attach it to the current patient test. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 79...
Patient Information screen after starting the displays asking you to confirm that the test selected order can be attached to the patient test, overwriting existing patient data. Go to step 3. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Select OK. If the patient test is not A message displays notifying you that the order will be transmitted detached from the patient test. Go to step 4. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 81...
The patient report displays in the ECG tab. Expand the Patient Information screen and scroll down to the Order Number field. Select Detach next to the Order Number field to clear the field. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
When the order list is updated, the orders attached to completed ECG patient reports transmitted to the MUSE server are cleared from the Orders list, and new orders are downloaded from the MUSE server. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 83...
Page 84
Work with Orders NOTE: The Attached order status is not included in the transmitted patient reports. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Opens the Files collapsed list that stores the saved patient reports. Expand icon Opens the Files expanded list. Files collpased Displays columns that provide information about the stored patient list columns reports. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 85...
Page 86
Navigates to the previous and next pages in the Files list. arrows Notification Shows the progress, error, or successful messages. status Delete icon Deletes the selected patient reports. Edit icon Edits patient information for the selected patient report. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
A Full Disclosure report opens in a FD Report tab. Review and make necessary changes to the patient report, before printing a copy of the report or transmitting the report to a configured destination. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 87...
Page 88
Select the Full Disclosure tab to close the opened stored patient Disclosure tab report tab and navigate to the Full Disclosure waveform and application. The Patient Information banner displays the demographics of the active patient. Select to close the tab. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
SFTP server destination with remote PDF format directory path Full Disclosure Shared directory destination with PDF format folder path To transmit a patient report to the default or configured destination, perform the steps below: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 89...
Page 90
If you select patient reports for transmission from the expanded Files list, the message displays on the notification area in the lower, right-side of the screen: <Count> reports added to the queue, where <Count> is the number of selected reports. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
From the Acquisition screen, select the Files list. The Files collapsed list opens. You can also select the Expand icon to open the Files expanded list: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 91...
Page 92
ECG or FD Report or Rhythm tab A job to print one copy of the patient report in the displayed report format is sent to the printer. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
• Press the Stop key on the keyboard: • Press the Stop button on the Acquisition Module: Print a List of Stored Records You can print all the stored records that display in the Files Manager. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 93...
Data assigned to the wrong patient causes erroneous patient data that can affect diagnosis and treatment. Be sure to check the patient information screen for each patient. Make sure that you enter patient data for the correct patient. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
If you do not have the privilege to view patient reports, but you have the privilege to delete patient reports, you can only view and delete patient reports you created in the current session. From the Acquisition screen, select Files. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 95...
Page 96
Acquisition screen, deleting the patient report closes the tab, and returns you to the Live tab. • Select Cancel to cancel the deletion. The selected patient reports are not deleted from the Files list. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Queue list. The successfully transmitted digital rhythm, FD report and ECG patient reports are immediately deleted from the Queue list. The figure illustrates the Queue collapsed list: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 97...
Page 98
Navigates to the previous and next pages in the Queue list. arrows Delete icon Deletes the selected transmission job from the Queue list. Retry icon Retries the transmission of an unsuccessful job. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
From the Acquisition screen, select Queue. The Queue collapsed list opens. To open the Queue expanded list, select the Expand icon: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 99...
The system automatically tries to transmit a patient report. If you need to re-transmit the patient report before the next automatic attempt, you can use this procedure to transmit the patient report immediately. From the Acquisition screen, select Queue. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 101
Job Failure Reason field. You can try to re- transmit the report again. Select the Collapse icon to close the Queue expanded list and return to the Acquisition screen: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 101...
The list of recent patients is stored in the Patients list, and you can view up to 500 recent patient records. The following figure illustrates the Patients collapsed list: The following figure illustrates the Patients expanded list: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 103
Select the same column header again to sort in descending order. Navigation Navigates to the previous and next pages in the Patients list. arrows 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 103...
The patient data from the selected patient record is populated in the Patient Information banner and screen, and the screen expands. Edit patient information in the fields. For more information, see Enter or Edit Patient Information Using the Keyboard on page MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
To lock the device while working on remote applications, select ALT + L on the keyboard. To unlock the device that was locked while working on remote applications, perform the steps in the procedure Unlock the Device on page 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 105...
The browser session of the remote application does not end but remains inactive. The browser session resumes when you go back to the open remote application. Close the browser session and return to the Acquisition screen. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Press the tab on the right side of the paper tray with a flat-head screwdriver to release the paper tray from the printer assembly and pull it out completely. Put the spacer in the paper tray. The spacer placement depends on the paper size. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 107...
Store Thermal Paper When imaged and stored properly, ECG tracings resist fading for several years. If your retention requirements exceed five years, consider using GE Archivist paper. To ensure the tracing is imaged properly, the device must be maintained in accordance with its service manuals and technical memoranda.
Clean the Printhead If the printer does not function, you may need to clean dust and foreign particles from the printhead. Use below procedure to clean the printhead: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 109...
Re-insert the paper and close the printer door when the heating element is completely dry. NOTE: Use only original GE Healthcare writer paper. This paper has a special coating that prevents electrostatic buildup and contamination and debris collection on the printhead. Using other paper may result in recordings of poor quality.
Press the lever of the battery slot from which you want to remove the battery in the upward direction and pull the battery handle to remove the battery. Repeat the action for the second battery. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 111...
Page 112
Slide the two batteries into the battery compartment slots in the correct orientation and direction indicated by the arrow. Lift the battery compartment door in the upward direction and close it. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Clean and Disinfect Guidelines Observe the guidelines while cleaning and disinfecting the device. • Follow cleaning instructions and observe hazards exactly as issued by GE Healthcare or other suppliers listed. • Avoid exposure to hypochlorite solutions and solutions containing iodine or high chlorine content, as these will promote corrosion.
Preparation for Cleaning For multi-piece or complex instruments, please refer to their disassembly instructions. ™ The disassembly instructions are available in the MAC VU360 Resting ECG Analysis System Service Manual. Contact your local GE Healthcare service representative for further information.
• Check the pins that the leadwires plug into. They should not be bent or loose. Contact GE Healthcare Customer Support for any repairs needed. • Check the Acquisition Module plastic case for any damage. Contact GE Healthcare Customer Support for any repairs needed.
Discard wipes to clinical waste. Do not reuse wipes. Post-Clean Inspection GE Healthcare devices should be visually inspected and functionally tested after cleaning and prior to disinfection for below items: • Cleanliness. • Damage, including but not limited to corrosion (rust, pitting), discoloration, excessive scratches, flaking, cracks, and wear.
If soil is still present, re-clean the equipment by repeating step 2. Allow the device to air dry. Discard wipes to clinical waste. Do not reuse wipes. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 117...
Cleaning and Disinfection Post-Clean Inspection GE Healthcare devices should be visually inspected and functionally tested after cleaning and prior to disinfection for below items: • Cleanliness. • Damage, including but not limited to corrosion (rust, pitting), discoloration, excessive scratches, flaking, cracks, and wear.
• The cleaning and disinfection information is provided in accordance with ANSI/ AAMIST81, ISO 17664. The recommendations provided above have been validated ™ as capable of preparing non-sterile GE Healthcare MAC VU360 medical devices. It remains the responsibility of the user to make sure that the cleaning and disinfection are performed using appropriate equipment, materials, and personnel and achieves the desired result.
Battery error. Attach power • Power sensor failure Replace the battery. cord. Contact Service. • Battery capacity sensor Contact your GE Healthcare failure Service Support representative if the error persists. Battery unknown error An unexpected battery error Replace the battery.
The keyboard is not working. Contact your GE Healthcare Service Support representative if the error persists. Touchscreen failure The touchscreen is not working. Contact your GE Healthcare Service Support representative if the error persists. Cannot perform action while You have tried to perform Perform only allowed actions.
The paper tray is empty. Insert sufficient paper in the paper tray. Printing will restart automatically. Printer error. Door is open. The printer door is open. Close the printer door. Printing will restart automatically. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 123...
Page 124
Printer error. Door is open. The printer door is open. Close the printer door. Select Print List from the Files Manager to restart printing of the stored records list. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 125
High printer temperature Restart the system to check Printer disabled. Contact if the error is resolved. If the Service. error persists, contact your GE Healthcare Service Support representative. Printer error. Restart system. Unknown error or hardware Restart the system to check Contact Service.
Destination unknown The destination is Reconfigure the destination. See not found. ™ MAC VU360 Resting ECG Analysis System Setup and Configuration Manual. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 127
USB port. the USB flash drive. Hilltop generation unsuccessful The file generation Retry transmission. is unsuccessful. Contact your GE Healthcare Service PDF generation unsuccessful Support representative if the error persists. Sapphire generation unsuccessful Unknown error...
How to troubleshoot: CA Certificate has The CA certificate has Obtain a PEM-encoded CA certificate. expired expired. Client Certificate has The client certificate has Obtain a PEM-encoded client certificate. expired expired. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 129
FIPS module is not FIPS module is not Contact your GE Healthcare Service detected installed. representative to install the FIPS module. FIPS unrecoverable error Incompatible firmware Contact your GE Healthcare Service version of FIPS module.
This results in a two page report. 4 by 2.5s Shows four columns of three rows of waveforms. Each One page column is 2.5 seconds wide and printed at 25 mm/s. report MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 131
The lower part of the report is text, such as the measurements or interpretation. Each column (page) is from the first 5 seconds of data. This results in a two page report. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 131...
Page 132
When printing 12 rows, this report does not have space for any additional text, such as measurements or interpretation. • When printing three or six rows, this report contains additional text, such as measurements or interpretation. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 133
The medians can have vertical tic marks added at the wave onsets and offsets (pon, poff, qon, qoff, and toff). There is no space in this report to print the interpretation. This results in a one page report. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 133...
Page 134
The scale shows how many mV are in 20 mm along the axes of each chart. The figure shows the standard report layout: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 135
• Location Vital Signs Displays information about the patient's vital signs, such as: • Heart rate • PR interval • QRS duration • QT/QTc • P-R-T axes • Blood pressure 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 135...
Page 136
Phase Distortion (ZPD for High Pass filter). Gain Setting Displays the gain of the ECG waveform (measured in mm/mV). Speed Setting Displays the speed of the ECG waveform (measured in mm/s). MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
After generating a rhythm report, it is a recommended best practice to review the report before allowing the patient to leave. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 137...
Page 138
When 2 pacemaker pulses are sensed so near each other that they cannot be uniquely shown on the report, a one pace spike with two flag tips is printed to show this condition: MAC VU360 ™ Resting ECG Analysis System...
Page 139
Each time the rhythm printing starts after a change in speed, a calibration pulse is added for each lead that tells the speed on the is printed rhythm. The digital rhythm report is recorded at the configured rhythm speed. The Leads. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 139...
This section describes the information contained in a FD report and explains where in the report that the information is located. After generating a FD report, it is a recommended best practice to review the report before allowing the patient to leave. MAC VU360 ™ Resting ECG Analysis System...
Page 141
For the FD report, the page number is shown as Page <1> of <3>. Prints the total number of seconds in the rhythm report on each page. Prints the selected single Lead information. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 141...
Page 142
The standard grid paper is divided into small squares of 1 mm x 1 mm and large squares of 5 mm x 5 mm. When recording 25 mm/s, 10 rows of data, 10 seconds of data per row is shown in each page on the FD report. MAC VU360 ™ Resting ECG Analysis System...
Check the Hookup Advisor for any indication of lead problems. NOTE: Use only electrodes and contact agents recommended by GE Healthcare. The signal quality on the Hookup Advisor is not indicated until the RA/R electrode is applied. If RA/R becomes disconnected, the system reports that all electrodes are off the patient.
Standard 12–Lead Electrode Placement To acquire a standard 12–lead ECG, use the electrode placement shown in the following diagram. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
RA white R red Right deltoid. Standard 15–Lead Electrode Placement To acquire a standard 15–lead ECG, use the electrode placement shown in the following diagram. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 145...
V8 gray C8 gray Under left mid-scapular line. V9 gray C9 gray Left paraspinal border. Pediatric Electrode Placement To acquire a pediatric ECG, use the electrode placement shown in the following diagram. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Same horizontal level of 4 in the left posterior axillary line. NEHB Electrode Placement To acquire a NEHB ECG, use the standard 12–lead electrode placement and items 1, 2 and 3 as shown in the following diagram. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 147...
Mid-clavicular line in the fifth intercostal space. V6 violet C6 violet Mid-axillary line on the same horizontal level as 3. LL red F green Left leg, in the lower abdominal quadrant. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 149
H light blue Back of the neck—avoid the carotid artery and jugular vein. M orange M light blue Center of the spine on the same horizontal level as 3 and 4. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 149...
Patient Information Text Box Names When an order is attached to a patient test, all Patient Information text boxes are read-only. The text box names in the table with an asterisk (*) are not. MAC VU360 ™ Resting ECG Analysis System...
Page 151
2000 to existent year. • xx = patient place of birth • g = patient gender • male = odd number • female = even number Swedish Patient ID: 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 151...
Page 152
100 years. • xx = patient place of birth • g = patient gender • male = odd number • female = even number • c = checksum digit Norwegian PID: MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 153
Based on the Mandatory fields apply for Acquisitionsettings, the ECG report is not accepted, transmitted, or printed until you enter the patient demographic data for the mandatory fields. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 153...
Page 154
• The date must be less than 127 years from the current date. The date of birth (DOB) also shows in the Patient Information bar. The calculated age shows near it. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 155
Secondary An alternative identification 1 to 17 • A to Z method. characters • a to z • 0 to 9 • All characters are supported. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 155...
Page 156
All characters are supported. Ordering Physician last name who ordered Maximum • A to Z MD Last the ECG. • a to z Name characters • 0 to 9 • All characters are supported. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 157
You can edit the Test The ECG is done because of this. Maximum • A to Z Indication* • a to z characters • 0 to 9 • All characters are supported. 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 157...
Page 158
Applicable • • Unknown Attending Attending physician last name Maximum • A to Z MD Last • a to z Name* characters • 0 to 9 • All characters are supported. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 159
A to Z MD First • a to z Name* characters • 0 to 9 • All characters are supported. Attending Attending physician ID Maximum 5 0 to 65534 MD ID* characters 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 159...
75.3 lbs max with adjustable height premium trolley and 1 battery USB Port 3 USB 2.0 ports supplying 0.5 Amps of current each Mechanical design Housing with built-in tiltable graphics display Software on mainboard MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 161
5, 12.5 mm/s at ±5% 25, 50 mm/s at ±2% Amplitude accuracy ±5% Horizontal resolution 40 dots/mm at 25 mm/s Vertical resolution 8 dots/mm Paper type Z-fold thermal with pre-printed grid and perforation 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 161...
Page 162
5, and Data Matrix symbology for characters A-Z (upper case), a-z (lower case), and 0-9 for all supported languages. Acquisition Module Real-time Hookup Advisor with LED lead quality indicators Ingress protection level: IPx4 Optional integrated barcode reader MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
30 to 300 beats per minute (BPM) with an accuracy of ±10% or 5 BPM, whichever is greater. Heart rates outside this range will not display. ECG data formats GE Hi-Fidelity ECG, XML Pre-acquisition Provides 10 seconds of instantaneous ECG acquisition Digital rhythm...
50kPa to 106kPa Safety Specifications Table 62: Safety Specifications Item Description Certification marks The medical device has a lifetime of 7 years with respect to the Council Directive 93/42/EEC essential requirement #4. MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
DSSS, CCK, OFDM, BPSK, QPSK, QAM Supported certificate key lengths 1024/2048/4096-bit encryption Supported certificate digest SHA1, SHA2 algorithms Wireless specifications Wireless Standards 802.11a/b/g/n WLAN interfaces Configured manually or via DHCP 2088531-370-2 MAC VU360 ™ Resting ECG Analysis System 165...
Page 166
Authentication Methods (FIPS) • • PEAP-GTC • EAP-TLS Wired Specifications Ethernet interface 802.3 Ethernet interface via RJ45 connector Wired Standards 10Base-T, 100Base-T, and 1000Base-T LAN x 1 port Configured manually or via DHCP MAC VU360 ™ Resting ECG Analysis System 2088531-370-2...
Page 167
Glossary Acute Coronary Syndrome Admission, Discharge, Transfer A filter sets the upper frequency limit for the ECG waveform displayed on the Filter Acquisition screen and the printout. Selecting a filter eliminates signals that exceed that frequency. The smaller the filter selected, the more signal is filtered out. For example, a filter of 40 Hz displays only signals less than 40 Hz;...
Page 168
Munzinger Straße 5 Milwaukee, WI 53223 USA 79111 Freiburg Germany +1 414 355 5000 +49 761 45 43-0 +1 800 558 7044 (US Only) GE Medical Systems Information Technologies, Inc., a General Electric Company, going to market as GE Healthcare. www.gehealthcare.com...