Customer-Supplied Insulation; Transmitter Installation; Rotating And Replacing The Display - Siemens SITRANS F FC520 Operating Instructions Manual

Coriolis flowmeters
Table of Contents

Advertisement

FC5x0
Installation
Risk of overheating the transmitter due to increased ambient temperature
WARNING
Failure of measuring electronics
▶ Observe the maximum allowable ambient temperature for the transmitter.
▶ Install the transmitter at a sufficient distance from heat sources.

12.5.2 Customer-supplied insulation

For insulation provided by the customer it is important to select a sensor with the
appropriate design type (remote type, sensor with long neck). The space between upper
insulation edge and lower edge of the sensor's terminal box must be at least 40 mm.
Recommended insulation thickness is 80 mm and recommended heat transfer coefficient
0.4 W/m² K.
Fig. 65: Customer-supplied insulation
1
When installing in hazardous areas, the applicable Explosion Proof Type Manual must be
DANGER
considered.

12.6 Transmitter installation

12.6.1 Rotating and replacing the display

The transmitter display can be oriented in line with the flow meter installation position.
The following instruction must only be performed at the following ambient conditions:
NOTICE
▶ at temperatures up to 31 °C: relative humidity maximum 80 %
▶ at temperatures between 31 °C and 40 °C: from 80 % linearly decreasing to 50 % of
1. Switch off power supply.
94 / 178
min. 40 mm
Insulation box
maximum relative humidity
A5E52748362-AA, 2023/08
Transmitter installation