3)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4)
Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
5)
Rinse the immersible part of the probe with running tap water for
1 minute. (alternatively: immerse the immersible part of the probe
in a tray filled with deionized water/tap water (see Fig.10) for
5 min.)
6)
Visually check the outer surface of the probe for cleanness. If
necessary, use magnifying glass for visually check. If there is
still soil visible, repeat all above steps.
B) Magnetic sensor attachment
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Immerse the magnetic sensor attachment into the diluted detergent
solution. Wipe it under the surface of the detergent solution with
a soft cloth to remove all visible soil.
3)
The Magnetic sensor attachment should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4)
Rinse the Magnetic sensor attachment with running tap water for
1 minute. (alternatively: immerse it in a tray filled with deionized
water/tap water (see Fig.10) for 5 min.)
5)
Visually check the outer surface of the Magnetic sensor attachment
for cleanness. If necessary, use magnifying glass for visually check.
If there is still soil visible, repeat all above steps.
- 20 -
Q1E-EP1451