3.2 Specifications
3.2 Specifications
(1)
Power supply:
5 V (supplied from e-Line)
(2)
Communication function:
e-Line (equipped with two e-Line connectors)
(3)
Analog signal input:
One channel
(4)
Input voltage:
-0.1 to 1.1 V
(5)
Input impedance:
10 M or higher
(6)
Accuracy:
0.1% (from -10 mV to 1.0 V, after zero offset)
(7)
Noise level:
1 10
-6
V or less
(after warmup; SP = 400 ms; with the input shorted)
(8)
Drift:
1 10
-5
V/hr or less (after warmup; with the input shorted)
(9)
Data acquisition cycle:
50, 100, 200, 400, 800, 1600, 3200 ms
(10) Ambient (environmental) temperature range:
4 to 35 C
(11) Ambient (environmental) humidity range:
25 to 85% RH
(12) Outer dimensions:
150 W 90 D 33 H mm
(13) Other functions:
Automatic zero-span adjusting function
(adjustment date is recorded as log)
3 - 2